Hilde Kjernlie Andersen1, David Völgyes2, Anne Catrine Trægde Martinsen1,3.
Abstract
BACKGROUND: Iterative reconstruction techniques for reducing radiation dose and improving image quality in CT have proved to work differently for different patient sizes, dose levels, and anatomical areas.Entities:
Keywords: CNR, contrast-to-noise ratio; CT; CT, computed tomography; FBP, filtered back-projection; Iterative reconstruction; Lung; MTF, modulation transfer function; NPS, noise power spectrum; Physics
Year: 2018 PMID: 29719856 PMCID: PMC5926271 DOI: 10.1016/j.ejro.2018.02.002
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
Scan technique for all scanners and both phantoms.
| GE | Siemens | Philips | Toshiba | |
|---|---|---|---|---|
| Detector width [mm] | 40 | 38,4 | 40 | 40 |
| Detector Collimation | 64 × 0.625 | 64 × 0.6 | 64 × 0.625 | 80 × 0.5 |
| rotation time [s/rotation] | 0.5 | 0.5 | 0.5 | 0.5 |
| pitch | 0.984 | 1.2 | 1.015 | 1.1 |
| mA | 65/125/255 | 89/178/360 | 77/154/315 | 70/130/260 |
| CTDIvol [mGy] | 2.6/4.9/10 | 2.5/5/10.1 | 2.5/5/10.1 | 2.7/5/9.9 |
| kVp | 120 | 120 | 120 | 120 |
| Reconstructed slice thickness [mm] | 2 | 2 | 2 | 2 |
| Displayed field-of-view [cm] for lung phantom | 32 | 32 | 32 | 32 |
| Displayed field-of-view [cm] for Catphan (MTF) | 21 | 21 | 21 | 21 |
Reconstruction techniques for all image series.
| Scanner | Reconstruction | |
|---|---|---|
| GE | VEO | VEO |
| Lung | ASIR20 | |
| FBP | ||
| Siemens | 70f | SAFIRE3 |
| FBP | ||
| Philips | IMR1 | Routine |
| Sharp+ | ||
| YB | iDose2 | |
| FBP | ||
| Toshiba | FC56 | AIDR3DSTD |
| FBP(QDS) | ||
Fig. 1Measurement of CNR in low-density insert simulating nodule.
Fig. 2Measurement of CNR in high-density insert simulating nodule.
Fig. 3Measurement of NPS in liver area of chest phantom.
CNR, NPS peak frequency and MTF for all reconstructed series.
| Dose level | Vendor | Algorithm/Kernel | Technique | CNR | NPS peak frequency | MTF critical frequency | |||
|---|---|---|---|---|---|---|---|---|---|
| Large | Small | Large | Small | 10% contrast | 2% contrast | ||||
| 10 mGy | GE | VEO | VEO | 20272 | 26195 | 0.38 | 0.43 | 10.2 | 12.2 |
| Lung | ASIR20 | 360 | 1109 | 0.55 | 0.58 | 10.1 | 12.2 | ||
| FBP | 281 | 820 | 0.60 | 0.58 | 10.9 | 12.2 | |||
| Siemens | 70f | SAFIRE3 | 441 | 1819 | 0.65 | 0.63 | 9.8 | 10.6 | |
| FBP | 141 | 569 | 0.65 | 0.63 | 10.6 | 12.2 | |||
| Philips | IMR1 | Routine | 19555 | 44991 | 0.50 | 0.50 | 8.1 | 9.6 | |
| Sharp+ | 6774 | 11063 | 0.58 | 0.60 | 10.2 | 11.4 | |||
| YB | iDose2 | 360 | 1567 | 0.65 | 0.63 | 10.1 | 11.3 | ||
| FBP | 248 | 1084 | 0.65 | 0.63 | 10.2 | 11.4 | |||
| Toshiba | FC56 | AIDR3DSTD | 963 | 2180 | 0.43 | 0.45 | 8.8 | 10.5 | |
| FBP | 326 | 1233 | 0.50 | 0.50 | 8.9 | 10.1 | |||
| 5 mGy | GE | VEO | VEO | 9750 | 17291 | 0.45 | 0.43 | ||
| Lung | ASIR20 | 121 | 630 | 0.58 | 0.55 | ||||
| FBP | 104 | 475 | 0.58 | 0.55 | |||||
| Siemens | 70f | SAFIRE3 | 380 | 754 | 0.58 | 0.60 | |||
| FBP | 117 | 237 | 0.60 | 0.60 | |||||
| Philips | IMR1 | Routine | 8496 | 32425 | 0.45 | 0.50 | |||
| Sharp+ | 3222 | 7652 | 0.60 | 0.63 | |||||
| YB | iDose2 | 94 | 569 | 0.60 | 0.63 | ||||
| FBP | 65 | 393 | 0.60 | 0.63 | |||||
| Toshiba | FC56 | AIDR3DSTD | 636 | 1130 | 0.40 | 0.43 | |||
| FBP | 162 | 610 | 0.50 | 0.50 | |||||
| 2.5 mGy | GE | VEO | VEO | 9480 | 10317 | 0.40 | 0.45 | ||
| Lung | ASIR20 | 82 | 204 | 0.58 | 0.58 | ||||
| FBP | 62 | 168 | 0.58 | 0.58 | |||||
| Siemens | 70f | SAFIRE3 | 208 | 524 | 0.63 | 0.60 | |||
| FBP | 67 | 171 | 0.63 | 0.65 | |||||
| Philips | IMR1 | Routine | 5888 | 10257 | 0.33 | 0.35 | |||
| Sharp+ | 2546 | 4127 | 0.60 | 0.58 | |||||
| YB | iDose2 | 66 | 194 | 0.60 | 0.60 | ||||
| FBP | 43 | 134 | 0.65 | 0.60 | |||||
| Toshiba | FC56 | AIDR3DSTD | 316 | 887 | 0.33 | 0.40 | |||
| FBP | 69 | 337 | 0.48 | 0.48 | |||||