Abstract
OBJECTIVES: Computed tomography (CT) allows reproducible assessment of left ventricular (LV) function, left ventricular outflow tract area (LVOTarea) and aortic valve area (AVA). We evaluated the influence of image reconstruction parameters on these measurements.Entities:
Mesh:
Year: 2021 PMID: 34233477 PMCID: PMC8523190 DOI: 10.1259/bjr.20201306
Source DB: PubMed Journal: Br J Radiol ISSN: 0007-1285 Impact factor: 3.039
Figure 1.(a) Automatic diastolic and systolic contour for LV functional assessment. (b–g) Assessment of left ventricular ejection fraction in a (b) standard reconstruction with 0.6-mm slice thickness, 20 phases of the cardiac cycle and 512 × 512 matrix size and modified reconstructions with (c) an iterative reconstruction algorithm, (D) a reduction in matrix size to 256 × 256, a reduction in the number of reconstructed phases of the cardiac cycles to (e) 10 (10% steps between 0 and 90%) and (f) 5 (20% steps between 0 and 80%) and (g) an increase of slice thickness to 8 mm.
Baseline demographics and acquisition parameters
| Value | |
|---|---|
| Patient characteristics | |
| Age (y) | 79 ± 6 |
| Male, n (%) | 28 (62) |
| BMI (kg/m2) | 26.7 ± 5.6 |
| Body weight (kg) | 75.9 ± 16.7 |
| Medication, n (%) | |
| Beta-blockers | 26 (58) |
| Ivabridine | 0 (0) |
| CT parameters | |
| Tube voltage, n (%) | |
| 90 kV | 1 (2) |
| 100 kV | 42 (94) |
| 110 kV | 2 (4) |
| Rhythm during acquisition, n (%) | |
| Sinus rhythm | 27 (60) |
| Atrial fibrillation | 18 (40) |
| Acquisition heart rate (beats/min) | 65 ± 11 (range 39–89) |
| Radiation exposure parameters | |
| Dose-length product (mGy*cm) | 1230 ± 222 (median 1190, range 898–1740) |
| Effective dose (mSv) | 1.7 ± 0.4 |
| Time to peak bolus (s) | 20.4 ± 1.5 (range 18–24) |
Assessment of global left ventricular ejection fraction using standard reconstruction and modified reconstructions
| Standard reconstruction | Modified reconstructions | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 0.6 mm | 0.6 mm | 0.6 mm | 0.6 mm | 0.6 mm | 1 mm | 2 mm | 5 mm | 8 mm | |
| Ejection fraction | |||||||||
| Mean ± SD (%) | 69 ± 12 | 68 ± 12 | 69 ± 12 | 67 ± 12 | 66 ± 12 | 68 ± 12 | 68 ± 12 | 68 ± 11 | 69 ± 10 |
| Median (%) | 73 | 72 | 74 | 72 | 72 | 72 | 73 | 71 | 70 |
|
| |||||||||
| Coefficient | 1.0 | 1.0 | 0.99 | 0.90 | 0.99 | 0.98 | 0.97 | 0.88 | |
| | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | |
|
| |||||||||
| 95% limits of agreement | 1.3; −1.8 | 1.8;−1.6 | 1.7;−4.1 | 7.0; −13.3 | 2.4;−3.6 | 3.7;−4.5 | 6.9;−7.4 | 12.5;−9.8 | |
| Mean bias | −0.30 | 0.1 | −1.2 | −3.1 | −0.6 | −0.4 | −0.2 | 1.3 | |
| | 0.022 | 0.594 | <0.001 | <0.001 | 0.011 | 0.332 | 0.577 | 0.436 | |
| Stroke Volume | |||||||||
| Mean ± SD (ml) | 87.6 ± 21.9 | 86.4 ± 22.1 | 88.1 ± 22.3 | 85.0 ± 20.8 | 82.3 ± 20.7 | 85.4 ± 20.3 | 86.0 ± 20.4 | 88.4 ± 21.0 | 94.5 ± 23.4 |
| Median (ml) | 82.2 | 80.8 | 82.1 | 78.6 | 78.2 | 80.5 | 80.3 | 83.7 | 94.8 |
|
| |||||||||
| Coefficient | 0.99 | 0.99 | 0.99 | 0.95 | 0.98 | 0.96 | 0.91 | 0.87 | |
| | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | |
|
| |||||||||
| 95% limits of agreement | 3.3; −5.7 | 5.0;−4.0 | 3.7;−8.9 | 8.7; −19.3 | 7.4;−11.8 | 10.6;−13.9 | 14.5;−21.2 | 31.8;−21.1 | |
| Mean bias | −1.2 | 0.5 | −2.6 | −5.3 | −2.2 | −1.6 | −0.8 | 5.3 | |
| | 0.003 | 0.218 | <0.001 | <0.001 | 0.021 | 0.307 | 0.939 | 0.015 | |
| End-diastolic volume | |||||||||
| Mean ± SD (ml) | 129.5 ± 31.1 | 128.2 ± 31.5 | 130.0 ± 31.1 | 127.9 ± 30.4 | 127.4 ± 30.4 | 127.6 ± 29.4 | 127.9 ± 29.4 | 132.9 ± 31.3 | 138.2 ± 35.3 |
| Median (ml) | 124.4 | 125.9 | 126.8 | 120.9 | 120.5 | 126.5 | 125.1 | 130.4 | 139.2 |
|
| |||||||||
| Coefficient | 1.0 | 1.0 | 1.0 | 1.0 | 0.99 | 0.97 | 0.96 | 0.94 | |
| | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | |
|
| |||||||||
| 95% limits of agreement | 4.7; −7.3 | 4.7;−3.5 | 3.4;−6.5 | 2.0; −4.1 | 8.9;−12.7 | 12.3;−15.4 | 20.5;−21.5 | 25.7;−16.2 | |
| Mean bias | −1.3 | 0.6 | −1.6 | −1.1 | −1.9 | −1.6 | −0.5 | 4.7 | |
| | 0.026 | 0.184 | <0.001 | <0.001 | 0.160 | 0.786 | 0.411 | 0.016 | |
| End-systolic volume | |||||||||
| Mean ± SD (ml) | 41.9 ± 21.8 | 41.9 ± 21.9 | 41.9 ± 21.9 | 42.9 ± 21.7 | 44.8 ± 21.2 | 42.2 ± 21.7 | 41.9 ± 21.1 | 44.5 ± 20.9 | 43.7 ± 20.9 |
| Median (ml) | 33.4 | 31.6 | 33.4 | 35.8 | 39.4 | 33.2 | 33.8 | 40.0 | 39.8 |
| Pearson Correlation compared to standard reconstruction | |||||||||
| Coefficient | 1.0 | 1.0 | 1.0 | 0.97 | 0.99 | 1.0 | 40.0 | 39.8 | |
| p-value | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | |
| Bland-Altman Analysis compared to standard reconstruction | |||||||||
| 95% limits of agreement | 1.8; −1.7 | 2.2; −2.1 | 4.3; −2.2 | 13.9; −8.0 | 3.7; −3.1 | 5.2; −5.2 | 10.6; −10.1 | 16.5; −17.6 | |
| Mean bias | 0 | 0 | 1.0 | 2.9 | 0.3 | 0 | 0.2 | −0.6 | |
| | 0.742 | 0.678 | <0.001 | <0.001 | 0.146 | 0.536 | 0.878 | 0.617 | |
| End-systolic phase for measurement | |||||||||
| Mean ± SD (%) | 38 ± 5 | 37 ± 5 | 37 ± 5 | 37 ± 6 | 39 ± 4 | 37 ± 5 | 38 ± 10 | 38 ± 5 | 38 ± 5 |
| | 0.414 | 0.102 | 0.162 | 0.032 | 0.655 | 0.782 | 0.102 | 0.331 | |
IR, iterative reconstruction; FBP, filtered back projection.
Figure 3.(a) Correlation graph for left ventricular ejection fraction and (b) Bland-Altman analysis for left ventricular ejection fraction, stroke volume, end-diastolic and end-systolic volume for modified reconstructions with an iterative reconstruction, a reduction in matrix size to 256 × 256, a reduction of the phases of the cardiac cycle to 5 (20% steps between 0–80%) and an increase in slice thickness of 8 mm in comparison with the standard reconstruction.
Assessment of left ventricular outflow tract area using standard reconstruction and modified reconstructions
| Standard reconstruction | Modified reconstructions | |||||||
|---|---|---|---|---|---|---|---|---|
| 0.6 mm | 0.6 mm | 0.6 mm | 0.6 mm | 0.6 mm | 1 mm | 2 mm | 5 mm | |
| Mean ± SD (mm²) | 482 ± 90 | 482 ± 88 | 481 ± 90 | 477 ± 89 | 472 ± 88 | 481 ± 89 | 483 ± 89 | 486 ± 89 |
| Median (mm²) | 482 | 495 | 476 | 469 | 460 | 470 | 480 | 480 |
|
| ||||||||
| Coefficient | 1.0 | 1.0 | 0.99 | 0.99 | 0.99 | 0.98 | 0.96 | |
| <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||
|
| ||||||||
| 95% limits of agreement | 13.2; −11.9 | 11.8;−12.4 | 11.5;−21.6 | 12.6; −32.4 | 22.8;−24.4 | 39.9;−38.3 | 54.6;−46.9 | |
| Mean bias | 0.6 | −0.3 | −5.0 | −9.9 | −0.8 | −0.4 | 3.9 | |
| 0.427 | 0.676 | <0.001 | <0.001 | 0.553 | 0.933 | 0.262 | ||
FBP, filtered back projection;IR, iterative reconstruction.
Assessment of aortic valve area using standard reconstruction and modified reconstructions
| Standard reconstruction | Modified reconstructions | ||||||
|---|---|---|---|---|---|---|---|
| 0.6 mm | 0.6 mm | 0.6 mm | 0.6 mm | 0.6 mm | 1 mm | 2 mm | |
| Aortic valve area | |||||||
| Mean ± SD (cm²) | 0.91 ± 0.21 | 0.90 ± 0.20 | 0.91 ± 0.21 | 0.88 ± 0.21 | 0.82 ± 0.20 | 0.90 ± 0.21 | 0.90 ± 0.20 |
| Median (cm²) | 0.95 | 0.94 | 0.96 | 0.89 | 0.84 | 0.94 | 0.96 |
|
| |||||||
| Coefficient | 0.97 | 0.96 | 0.97 | 0.88 | 0.96 | 0.93 | |
| | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | |
|
| |||||||
| 95% limits of agreement | 0.10; −0.12 | 0.12;−0.12 | 0.07;−0.14 | 0.11; −0.29 | 0.10; −0.12 | 0.14;−0.16 | |
| Mean bias | −0.01 | 0.0 | −0.04 | −0.09 | −0.01 | −0.01 | |
| | 0.354 | 0.938 | <0.001 | <0.001 | 0.120 | 0.341 | |
|
| |||||||
| End-systolic phase for measurement | |||||||
| Mean ± SD (%) | 14 ± 4 | 14 ± 4 | 14 ± 4 | 15 ± 5 | 20 ± 0 | 14 ± 5 | 14 ± 4 |
| | 0.290 | 0.617 | 0.269 | <0.001 | 0.796 | 0.617 | |
FBP, filtered back projection; IR, iterative reconstruction.