Comparing five different iterative reconstruction algorithms for computed tomography in an ROC study.

Kristin Jensen1, Anne Catrine T Martinsen, Anders Tingberg, Trond Mogens Aaløkken, Erik Fosse.   

Abstract

OBJECTIVES: The purpose of this study was to evaluate lesion conspicuity achieved with five different iterative reconstruction techniques from four CT vendors at three different dose levels. Comparisons were made of iterative algorithm and filtered back projection (FBP) among and within systems.
METHODS: An anthropomorphic liver phantom was examined with four CT systems, each from a different vendor. CTDIvol levels of 5 mGy, 10 mGy and 15 mGy were chosen. Images were reconstructed with FBP and the iterative algorithm on the system. Images were interpreted independently by four observers, and the areas under the ROC curve (AUCs) were calculated. Noise and contrast-to-noise ratios (CNR) were measured.
RESULTS: One iterative algorithm increased AUC (0.79, 0.95, and 0.97) compared to FBP (0.70, 0.86, and 0.93) at all dose levels (p < 0.001 and p = 0.047). Another algorithm increased AUC from 0.78 with FBP to 0.84 (p = 0.007) at 5 mGy. Differences at 10 and 15 mGy were not significant (p-values: 0.084-0.883). Three algorithms showed no difference in AUC compared to FBP (p-values: 0.008-1.000). All of the algorithms decreased noise (10-71%) and improved CNR.
CONCLUSIONS: Only two algorithms improved lesion detection, even though noise reduction was shown with all algorithms. KEY POINTS: Iterative reconstruction algorithms affected lesion detection differently at different dose levels. One iterative algorithm improved lesion detectability compared to filtered back projection. Three algorithms did not significantly improve lesion detectability. One algorithm improved lesion detectability at the lowest dose level.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25048191     DOI: 10.1007/s00330-014-3333-4

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  27 in total

1.  Upgrade to iterative image reconstruction (IR) in abdominal MDCT imaging: a clinical study for detailed parameter optimization beyond vendor recommendations using the adaptive statistical iterative reconstruction environment (ASIR).

Authors:  F G Mueck; M Körner; M K Scherr; L L Geyer; Z Deak; U Linsenmaier; M Reiser; S Wirth
Journal:  Rofo       Date:  2011-12-23

2.  Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT.

Authors:  Peter B Noël; Alexander A Fingerle; Bernhard Renger; Daniela Münzel; Ernst J Rummeny; Martin Dobritz
Journal:  AJR Am J Roentgenol       Date:  2011-12       Impact factor: 3.959

Review 3.  Iterative reconstruction methods in X-ray CT.

Authors:  Marcel Beister; Daniel Kolditz; Willi A Kalender
Journal:  Phys Med       Date:  2012-02-10       Impact factor: 2.685

4.  Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients.

Authors:  Yoshiko Sagara; Amy K Hara; William Pavlicek; Alvin C Silva; Robert G Paden; Qing Wu
Journal:  AJR Am J Roentgenol       Date:  2010-09       Impact factor: 3.959

5.  Iterative reconstruction reduces abdominal CT dose.

Authors:  Anne Catrine Trægde Martinsen; Hilde Kjernlie Sæther; Per Kristian Hol; Dag Rune Olsen; Per Skaane
Journal:  Eur J Radiol       Date:  2011-05-02       Impact factor: 3.528

Review 6.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm.

Authors:  Alvin C Silva; Holly J Lawder; Amy Hara; Jennifer Kujak; William Pavlicek
Journal:  AJR Am J Roentgenol       Date:  2010-01       Impact factor: 3.959

7.  Noise-reducing algorithms do not necessarily provide superior dose optimisation for hepatic lesion detection with multidetector CT.

Authors:  K L Dobeli; S J Lewis; S R Meikle; D L Thiele; P C Brennan
Journal:  Br J Radiol       Date:  2013-02-07       Impact factor: 3.039

8.  Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver.

Authors:  Mark E Baker; Frank Dong; Andrew Primak; Nancy A Obuchowski; David Einstein; Namita Gandhi; Brian R Herts; Andrei Purysko; Erick Remer; Neil Vachhani; Neil Vachani
Journal:  AJR Am J Roentgenol       Date:  2012-07       Impact factor: 3.959

9.  Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments--a phantom approach.

Authors:  Frédéric A Miéville; François Gudinchet; Francis Brunelle; François O Bochud; Francis R Verdun
Journal:  Phys Med       Date:  2012-01-02       Impact factor: 2.685

10.  Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Jiang Hsieh; Paul E Licato; Synho Do; Homer H Pien; Michael A Blake
Journal:  Radiology       Date:  2010-09-09       Impact factor: 11.105

View more
  14 in total

1.  Comparison of the image qualities of filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction for CT venography at 80 kVp.

Authors:  Jin Hyeok Kim; Ki Seok Choo; Tae Yong Moon; Jun Woo Lee; Ung Bae Jeon; Tae Un Kim; Jae Yeon Hwang; Myeong-Ja Yun; Dong Wook Jeong; Soo Jin Lim
Journal:  Eur Radiol       Date:  2015-10-20       Impact factor: 5.315

2.  Fine focal spot size improves image quality in computed tomography abdomen and pelvis.

Authors:  Yin P Goh; Kenneth K Lau; Keat Low; Kevin Buchan; Lawrence Chia Wei Oh; Ahilan Kuganesan; Minh Huynh
Journal:  Eur Radiol       Date:  2016-03-14       Impact factor: 5.315

3.  Reaching for better image quality and lower radiation dose in head and neck CT: advanced modeled and sinogram-affirmed iterative reconstruction in combination with tube voltage adaptation.

Authors:  Andrea I Schmid; Michael Uder; Michael M Lell
Journal:  Dentomaxillofac Radiol       Date:  2016-09-08       Impact factor: 2.419

4.  Diagnostic performance of reduced-dose CT with a hybrid iterative reconstruction algorithm for the detection of hypervascular liver lesions: a phantom study.

Authors:  Atsushi Nakamoto; Yoshikazu Tanaka; Hiroshi Juri; Go Nakai; Shushi Yoshikawa; Yoshifumi Narumi
Journal:  Eur Radiol       Date:  2016-12-12       Impact factor: 5.315

5.  Estimation of the Initial Image's Contributions to the Iterative Landweber Reconstruction.

Authors:  Gengsheng L Zeng
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2017-11-17

6.  Dose Reduction While Preserving Diagnostic Quality in Head CT: Advancing the Application of Iterative Reconstruction Using a Live Animal Model.

Authors:  F D Raslau; E J Escott; J Smiley; C Adams; D Feigal; H Ganesh; C Wang; J Zhang
Journal:  AJNR Am J Neuroradiol       Date:  2019-10-10       Impact factor: 3.825

7.  Comparison of full-iodine conventional CT and half-iodine virtual monochromatic imaging: advantages and disadvantages.

Authors:  Haruto Sugawara; Shigeru Suzuki; Yoshiaki Katada; Takuya Ishikawa; Rika Fukui; Yuzo Yamamoto; Osamu Abe
Journal:  Eur Radiol       Date:  2018-09-12       Impact factor: 5.315

8.  Application of 80-kVp scan and raw data-based iterative reconstruction for reduced iodine load abdominal-pelvic CT in patients at risk of contrast-induced nephropathy referred for oncological assessment: effects on radiation dose, image quality and renal function.

Authors:  Yasunori Nagayama; Shota Tanoue; Akinori Tsuji; Joji Urata; Mitsuhiro Furusawa; Seitaro Oda; Takeshi Nakaura; Daisuke Utsunomiya; Eri Yoshida; Morikatsu Yoshida; Masafumi Kidoh; Machiko Tateishi; Yasuyuki Yamashita
Journal:  Br J Radiol       Date:  2018-03-02       Impact factor: 3.039

9.  A qualitative and quantitative analysis of radiation dose and image quality of computed tomography images using adaptive statistical iterative reconstruction.

Authors:  Fahad Ahmed Hussain; Noor Mail; Abdulrahman M Shamy; Alghamdi Suliman; Abdelhamid Saoudi
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

10.  Image quality with iterative reconstruction techniques in CT of the lungs-A phantom study.

Authors:  Hilde Kjernlie Andersen; David Völgyes; Anne Catrine Trægde Martinsen
Journal:  Eur J Radiol Open       Date:  2018-03-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.