Image Quality in Oncologic Chest Computerized Tomography With Iterative Reconstruction: A Phantom Study.

Kristin Jensen1, Trond Mogens Aaløkken, Anders Tingberg, Erik Fosse, Anne Catrine T Martinsen.   

Abstract

OBJECTIVE: The purpose of this study was to validate iterative reconstruction technique in oncologic chest computed tomography (CT).
METHODS: An anthropomorphic thorax phantom with 4 simulated tumors was scanned on a 64-slice CT scanner with 2 different iterative reconstruction techniques: one model based (MBIR) and one hybrid (ASiR). Dose levels of 14.9, 11.1, 6.7, and 0.6 mGy were used, and all images were reconstructed with filtered back projection (FBP) and both iterative reconstruction algorithms. Hounsfield units (HU) and absolute noise were measured in the tumors, lung, heart, diaphragm, and muscle. Contrast-to-noise ratios (CNRs) and signal-to-noise ratios (SNRs) were calculated.
RESULTS: Model-based iterative reconstruction (MBIR) increased CNRs of the tumors (21.1-192.2) and SNRs in the lung (-49.0-165.6) and heart (3.1-8.5) at all dose levels compared with FBP (CNR, 1.1-23.0; SNR, -7.5-31.6 and 0.2-1.1) and with adaptive statistical iterative reconstruction (CNR, 1.2-33.2; SNR, -7.3-37.7 and 0.2-1.5). At the lowest dose level (0.6 mGy), MBIR reduced the cupping artifact (HU range: 17.0 HU compared with 31.4-32.2). An HU shift in the negative direction was seen with MBIR.
CONCLUSIONS: Quantitative image quality parameters in oncologic chest CT are improved with MBIR compared with FBP and simpler iterative reconstruction algorithms. Artifacts at low doses are reduced. A shift in HU values was shown; thus, absolute HU values should be used with care.

Entities:  

Mesh:

Year:  2016        PMID: 27192499     DOI: 10.1097/RCT.0000000000000364

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  5 in total

1.  Single-energy pediatric chest computed tomography with spectral filtration at 100 kVp: effects on radiation parameters and image quality.

Authors:  Boris Bodelle; Constanze Fischbach; Christian Booz; Ibrahim Yel; Claudia Frellesen; Moritz Kaup; Martin Beeres; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Pediatr Radiol       Date:  2017-03-28

2.  Task-Based Model Observer Assessment of A Partial Model-Based Iterative Reconstruction Algorithm in Thoracic Oncologic Multidetector CT.

Authors:  David C Rotzinger; Damien Racine; Catherine Beigelman-Aubry; Khalid M Alfudhili; Nathalie Keller; Pascal Monnin; Francis R Verdun; Fabio Becce
Journal:  Sci Rep       Date:  2018-12-07       Impact factor: 4.379

3.  Automatic current selection with iterative reconstruction reduces effective dose to less than 1 mSv in low-dose chest computed tomography in persons with normal BMI.

Authors:  Li-Guo Chen; Ping-An Wu; Min-Huei Sheu; Hsing-Yang Tu; Li-Chuan Huang
Journal:  Medicine (Baltimore)       Date:  2019-07       Impact factor: 1.817

4.  Iterative reconstruction improves image quality and reduces radiation dose in trauma protocols; A human cadaver study.

Authors:  Johannes Clemens Godt; Cathrine K Johansen; Anne Catrine T Martinsen; Anselm Schulz; Helga M Brøgger; Kristin Jensen; Arne Stray-Pedersen; Johann Baptist Dormagen
Journal:  Acta Radiol Open       Date:  2021-11-18

5.  Image quality with iterative reconstruction techniques in CT of the lungs-A phantom study.

Authors:  Hilde Kjernlie Andersen; David Völgyes; Anne Catrine Trægde Martinsen
Journal:  Eur J Radiol Open       Date:  2018-03-08
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.