Literature DB >> 12184422

The influence of sodium hypochlorite irrigation on the accuracy of the Root ZX electronic apex locator.

W Anthony Meares1, H Robert Steiman.   

Abstract

The purpose of this in vitro investigation was to determine whether the presence of sodium hypochlorite influences the accuracy of the Root ZX electronic apex locator. Forty, extracted, human teeth were mounted in an experimental apparatus. After achieving ideal access, working length measurements were obtained using the Root ZX. The canals were flushed with 2.125% sodium hypochlorite and measurements were again made with the electronic apex locator. Before measuring a third time, 5.25% sodium hypochlorite was then administered to each canal. Finally, the tooth was removed from the apparatus and the actual canal length was determined by measuring a file brought to the apical constriction (verified by direct visualization). No significant differences were found between the experimental groups. Overall, Root ZX measurements were within 0.5 mm of the actual length 83% of the time. The results of this study indicate that the Root ZX is not adversely affected by the presence of sodium hypochlorite.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12184422     DOI: 10.1097/00004770-200208000-00008

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  14 in total

1.  Comparison of two electronic apex locators on human cadavers.

Authors:  Luca Marigo; Gian Luca Gervasi; Francesco Somma; Giuseppe Squeo; Raffaella Castagnola
Journal:  Clin Oral Investig       Date:  2015-11-04       Impact factor: 3.573

2.  Accuracy of two different apex locators in primary teeth with and without root resorption.

Authors:  Haluk Bodur; Mesut Odabaş; Ozlem Tulunoğlu; Ali Cemal Tinaz
Journal:  Clin Oral Investig       Date:  2007-11-20       Impact factor: 3.573

3.  Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.

Authors:  Franziska Haupt; M Hülsmann
Journal:  Clin Oral Investig       Date:  2018-03-07       Impact factor: 3.573

4.  Effect of Sodium Hypochlorite Concentration on Electronic Apex Locator Reliability.

Authors:  Franck Diemer; Emma Plews; Marie Georgelin-Gurgel; Lora Mishra; Hyeon-Cheol Kim
Journal:  Materials (Basel)       Date:  2022-01-23       Impact factor: 3.623

5.  In vitro evaluation of the accuracy of two electronic apex locators.

Authors:  Viresh Chopra; Shibani Grover; S Datta Prasad
Journal:  J Conserv Dent       Date:  2008-04

6.  Efficiency of 2 electronic apex locators on working length determination: A clinical study.

Authors:  Sibel Koçak; Mustafa Murat Koçak; Baran Can Sağlam
Journal:  J Conserv Dent       Date:  2013-05

7.  Evaluation of Dentaport ZX and Raypex 6 electronic apex locators: an in vivo study.

Authors:  S Moscoso; K Pineda; J Basilio; C Alvarado; M Roig; F Duran-Sindreu
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2014-03-01

8.  In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.

Authors:  Ugur Aydin; Emrah Karataslioglu; Fatih Aksoy; Cihan Yildirim
Journal:  J Conserv Dent       Date:  2015 Jan-Feb

9.  Comparison between two methods of working length determination and its effect on radiographic extent of root canal filling: a clinical study [ISRCTN71486641].

Authors:  L Smadi
Journal:  BMC Oral Health       Date:  2006-02-11       Impact factor: 2.757

10.  In vitro Evaluation of the Accuracy of the Root Zx in the Presence of Naocl 2.5% and Chlorhexidine 0.2.

Authors:  Marjan Bolbolian; Siamak Golchin; Seyedmatin Faegh
Journal:  J Clin Exp Dent       Date:  2018-11-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.