The Influence of Various Irrigants on the Accuracy of 2 Electronic Apex Locators in Locating Simulated Root Perforations.

Demet Altunbaş1, Alper Kuştarcı2, Mustafa Toyoğlu3.   

Abstract

INTRODUCTION: The aim of this study was to assess the accuracy of the Dentaport ZX (Morita Co, Kyoto, Japan) and the Rootor (Meta Biomed, Cheongwon-gun, Korea) electronic apex locators (EALs) in detecting root perforations in dry conditions and in the presence of the following irrigation solutions: 2.5% sodium hypochlorite (NaOCl), 0.9% saline solution (NaCl), and 17% EDTA.
METHODS: Twenty extracted, single-rooted human teeth were perforated artificially in the middle section. The actual canal lengths (ALs) up to the perforation site were determined, and then the teeth were embedded in an alginate mold. The electronic measurements of the perforations were obtained using a size 20 K-file by each EAL in various conditions. For each tooth, the AL was subtracted from the electronic length of the perforation. Statistical analyses were performed using the Friedman and Wilcoxon signed rank tests at a significance level of P < .05.
RESULTS: The most accurate measurements for both EALs were obtained in dry conditions. Among the irrigation solutions in the Dentaport ZX group, EDTA gave the most accurate results, and NaOCl gave the least accurate ones. However, measurements with NaCl were closer to the AL than those obtained with EDTA for the Rootor group. Significant differences were noted among the EALs when the measurements were taken with NaOCl, NaCl, and EDTA (P < .05).
CONCLUSIONS: The Dentaport ZX was more accurate compared with the Rootor in the presence of different irrigants. The content of the root canal affected the accuracy of both EALs. The most accurate measurements were obtained in dry canals.
Copyright © 2016 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Apex locator; Dentaport ZX; Rootor; irrigant; root perforation

Mesh:

Substances:

Year:  2017        PMID: 28104321     DOI: 10.1016/j.joen.2016.10.031

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  8 in total

1.  Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.

Authors:  Franziska Haupt; M Hülsmann
Journal:  Clin Oral Investig       Date:  2018-03-07       Impact factor: 3.573

2.  Effect of Sodium Hypochlorite Concentration on Electronic Apex Locator Reliability.

Authors:  Franck Diemer; Emma Plews; Marie Georgelin-Gurgel; Lora Mishra; Hyeon-Cheol Kim
Journal:  Materials (Basel)       Date:  2022-01-23       Impact factor: 3.623

3.  Comparative Evaluation of Accuracy of Two Electronic Apex Locators in the Presence of Contemporary Irrigants: An In vitro Study.

Authors:  Queentaj Baruah; Nidhi Sinha; Bijay Singh; P Narendranatha Reddy; Kaveri Baruah; Vaishak Augustine
Journal:  J Int Soc Prev Community Dent       Date:  2018-07-18

4.  Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants.

Authors:  Sakshi Bilaiya; Pallav Mahesh Patni; Pradeep Jain; Sanket Hans Pandey; Swadhin Raghuwanshi; Bhupesh Bagulkar
Journal:  Eur Endod J       Date:  2020-03-19

5.  Accuracy of two electronic apex locators in locating root perforations in curved canals in dry and wet conditions: A comparative in vitro study.

Authors:  Monisha Parshotam Khatri; Sheetal B Ghivari; Madhu Pujar; Reshma Faras; Pallavi Gopeshetti; Amulya Vanti
Journal:  Dent Res J (Isfahan)       Date:  2019-11-12

6.  Ex Vivo Evaluation of the Accuracy of 3 Electronic Apex Locators in Different Environments: A Micro-Computed Tomography Study.

Authors:  Firdevs Çınar; Yakup Üstün
Journal:  Eur Endod J       Date:  2020-12

7.  Effect of Orthodontic Brackets on the Accuracy of Apex Locators: A Pilot Study.

Authors:  Özgür Genç Şen; Özgür İlke Ulusoy; Yelda Nayır Paltun; Çağrı Ulusoy
Journal:  Biomed Res Int       Date:  2021-04-24       Impact factor: 3.411

8.  Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study.

Authors:  Malavika Mohan; Meenakshi R Verma; Ashish K Jain; Rahul D Rao; Priyanka Yadav; Sonal Agrawal
Journal:  J Conserv Dent       Date:  2022-05-02
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.