Yoshiyuki Katsuta1,2, Noriyuki Kadoya2, Yukio Fujita3, Eiji Shimizu1, Kenichi Matsunaga1, Kinya Sawada1, Haruo Matsushita2, Kazuhiro Majima1, Keiichi Jingu2.
Abstract
Log file-based methods are attracting increasing interest owing to their ability to validate volumetric-modulated arc therapy outputs with high resolution in the leaf and gantry positions and in delivered dose. Cross-validation of these methods for comparison with measurement-based methods using the ionization chamber/ArcCHECK-3DVH software (version 3.2.0) under the same conditions of treatment anatomy and plan enables an efficient evaluation of this method. In this study, with the purpose of cross-validation, we evaluate the accuracy of a log file-based method using Elekta log files and an X-ray voxel Monte Carlo dose calculation technique in the case of leaf misalignment during prostate volumetric-modulated arc therapy. In this study, 10 prostate volumetric-modulated arc therapy plans were used. Systematic multileaf collimator leaf positional errors (±0.4 and ±0.8 mm for each single bank) were deliberately introduced into the optimized plans. Then, the delivered 3-dimensional doses to a phantom with a certain patient anatomy were estimated by our system. These doses were compared with the ionization chamber dose and the ArcCHECK-3DVH dose. For the given phantom and patient anatomy, the estimated dose strongly coincided with the ionization chamber/ArcCHECK-3DVH dose ( P < .01). In addition, good agreement between the estimated dose and the ionization chamber/ArcCHECK-3DVH dose was observed. The dose estimation accuracy of our system, which combines Elekta log files and X-ray voxel Monte Carlo dose calculation, was evaluated.Entities:
Keywords: Monte Carlo; log file; oncology; patient-specific QA; radiotherapy; volumetric-modulated arc therapy
Mesh:
Year: 2017 PMID: 29334027 PMCID: PMC5762095 DOI: 10.1177/1533034617745250
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
Figure 1.Flowchart of the cross-validation of the log file dose and ionization chamber/ArcCHECK-3DVH dose.
Figure 2.For the phantom geometry, the mean dose to the ionization chamber volume estimated using the log file–based method versus the ionization chamber measurement.
Figure 3.For the patient anatomy, the gamma index distributions (1%/2 mm, global, 20% dose threshold) between the log file and ArcCHECK-3DVH doses for typical cases; (A) axial plane, (B) coronal plane, and (C) sagittal plane. The planning target volume was contoured in green line, bladder was blue line, and rectum was orange line.
Figure 4.For the phantom geometry, the mean dose to tumor/normal tissue volumes estimated using the log file–based method versus that obtained using ArcCHECK-3DVH; (A) planning target volume, (B) bladder, and (C) rectum.