Sensitivity of volumetric modulated arc therapy patient specific QA results to multileaf collimator errors and correlation to dose volume histogram based metrics.

Linda Coleman1, Christina Skourou.   

Abstract

PURPOSE: This study investigates the impact of systematic multileaf collimator (MLC) positional errors on gamma analysis results used for quality assurance (QA) of Rapidarc treatments. In addition, this study evaluates the relationship of these gamma analysis results and clinical dose volume histogram metrics (DVH) for Rapidarc treatment plans.
METHODS: Five prostate plans were modified by the introduction of systematic MLC errors. The MLC shifts to each individual active leaf introduced were 0.25, 0.5, 0.75, and 1 mm. All QA verification plans were delivered and estimated 3D patient dose or high density phantom dose were obtained based on the ArcCHECK measurement files. QA gamma analysis of 3%/3 mm and 2%/2 mm were implemented and relationships to dose differences in DVH metrics encountered due to MLC errors were determined. Tolerances of 3% and 5% for DVH metric were implemented to determine the sensitivity of gamma analysis to MLC errors. A calculation of sensitivity was determined from the number of incidences of false negative and false positive cases in gamma analysis results.
RESULTS: The sensitivity of global gamma analysis for criteria of 3%/3 mm was 0.78 and for 2%/2 mm was 0.82. A number of instances occurred for an acceptable VMAT QA gamma index which did not indicate a DVH metric dose error greater than 5%. The correlation between global gamma analysis using criteria 3%/3 mm and DVH metric dose error were all <0.8 indicating less than a strong correlation.
CONCLUSIONS: There is a greater sensitivity for detection of dosimetric errors occurring in a Rapidarc plan using gamma criteria of 2%/2 mm than 3%/3 mm. However, there is lack of consistently strong correlation between global gamma indexes and clinical DVH metrics for PTV and bladder and rectum for Rapidarc plans. It is recommended that the sole use of gamma index for Rapidarc QA plan evaluation could be insufficient and a methodology for evaluation of delivered dose to patient is required.

Entities:  

Mesh:

Year:  2013        PMID: 24320423     DOI: 10.1118/1.4824433

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Evaluation of the ability of three commercially available dosimeters to detect systematic delivery errors in step-and-shoot IMRT plans.

Authors:  Alison Gray; Omemh Bawazeer; Sankar Arumugam; Philip Vial; Joseph Descallar; David Thwaites; Lois Holloway
Journal:  Rep Pract Oncol Radiother       Date:  2021-09-30

2.  Comparisons of volumetric modulated arc therapy (VMAT) quality assurance (QA) systems: sensitivity analysis to machine errors.

Authors:  Bin Liang; Bo Liu; Fugen Zhou; Fang-Fang Yin; Qiuwen Wu
Journal:  Radiat Oncol       Date:  2016-11-07       Impact factor: 3.481

3.  Breaking bad IMRT QA practice.

Authors:  Strahinja Stojadinovic; Luo Ouyang; Xuejun Gu; Arnold Pompoš; Qinan Bao; Timothy D Solberg
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

4.  Evaluation of Dose Distribution in Intensity Modulated Radiosurgery for Lung Cancer under Condition of Respiratory Motion.

Authors:  Mee Sun Yoon; Jae-Uk Jeong; Taek-Keun Nam; Sung-Ja Ahn; Woong-Ki Chung; Ju-Young Song
Journal:  PLoS One       Date:  2016-09-20       Impact factor: 3.240

5.  Dosimetric and radiobiological comparison for quality assurance of IMRT and VMAT plans.

Authors:  Nava Raj Paudel; Ganesh Narayanasamy; Eun Young Han; Jose Penagaricano; Panayiotis Mavroidis; Xin Zhang; Anil Pyakuryal; Dongwook Kim; Xiaoying Liang; Steven Morrill
Journal:  J Appl Clin Med Phys       Date:  2017-08-03       Impact factor: 2.102

6.  Tuning of AcurosXB source size setting for small intracranial targets.

Authors:  Stephen J Gardner; Siming Lu; Chang Liu; Ning Wen; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2017-05-04       Impact factor: 2.102

7.  Patient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy.

Authors:  Yoshiyuki Katsuta; Noriyuki Kadoya; Yukio Fujita; Eiji Shimizu; Kenichi Matsunaga; Kinya Sawada; Haruo Matsushita; Kazuhiro Majima; Keiichi Jingu
Journal:  Technol Cancer Res Treat       Date:  2017-12-05

8.  Accuracy of one algorithm used to modify a planned DVH with data from actual dose delivery.

Authors:  Tianjun Ma; Matthew B Podgorsak; Lalith K Kumaraswamy
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

9.  Practical implications for the quality assurance of modulated radiation therapy techniques using point detector arrays.

Authors:  Steffi Kantz; Almut Troeller McDermott; Matthias Söhn; Sabine Reinhardt; Claus Belka; Katia Parodi; Michael Reiner
Journal:  J Appl Clin Med Phys       Date:  2017-08-30       Impact factor: 2.102

10.  To propose adding index of achievement (IOA) to IMRT QA process.

Authors:  Dong-Su Kim; Siyong Kim; Seong-Hee Kang; Tae-Ho Kim; So-Hyun Park; Kyeong-Hyeon Kim; Min-Seok Cho; Dong-Seok Shin; Yu-Yun Noh; Jin-Beom Chung; Tae Suk Suh
Journal:  Radiat Oncol       Date:  2018-06-15       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.