| Literature DB >> 24710455 |
Juan F Calvo-Ortega1, Tony Teke, Sandra Moragues, Miquel Pozo, Joan Casals-Farran.
Abstract
In the present study, we describe a method based on the analysis of the dynamic MLC log files (DynaLog) generated by the controller of a Varian linear accelerator in order to perform patient-specific IMRT QA. The DynaLog files of a Varian Millennium MLC, recorded during an IMRT treatment, can be processed using a MATLAB-based code in order to generate the actual fluence for each beam and so recalculate the actual patient dose distribution using the Eclipse treatment planning system. The accuracy of the DynaLog-based dose reconstruction procedure was assessed by introducing ten intended errors to perturb the fluence of the beams of a reference plan such that ten subsequent erroneous plans were generated. In-phantom measurements with an ionization chamber (ion chamber) and planar dose measurements using an EPID system were performed to investigate the correlation between the measured dose changes and the expected ones detected by the reconstructed plans for the ten intended erroneous cases. Moreover, the method was applied to 20 cases of clinical plans for different locations (prostate, lung, breast, and head and neck). A dose-volume histogram (DVH) metric was used to evaluate the impact of the delivery errors in terms of dose to the patient. The ionometric measurements revealed a significant positive correlation (R² = 0.9993) between the variations of the dose induced in the erroneous plans with respect to the reference plan and the corresponding changes indicated by the DynaLog-based reconstructed plans. The EPID measurements showed that the accuracy of the DynaLog-based method to reconstruct the beam fluence was comparable with the dosimetric resolution of the portal dosimetry used in this work (3%/3 mm). The DynaLog-based reconstruction method described in this study is a suitable tool to perform a patient-specific IMRT QA. This method allows us to perform patient-specific IMRT QA by evaluating the result based on the DVH metric of the planning CT image (patient DVH-based IMRT QA).Entities:
Mesh:
Year: 2014 PMID: 24710455 PMCID: PMC5875466 DOI: 10.1120/jacmp.v15i2.4665
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Workflow of the patient dose‐volume histogram‐based IMRT QA procedure.
Comparison of the measured changes vs. the detected ones using the DynaLog‐based method for the isocenter dose. A significant correlation was obtained
|
|
|
|
|
|---|---|---|---|
| 1 |
| 1.038 | 1.039 |
| 2 | 2 mm MLC opening | 1.077 | 1.077 |
| 3 | 4 mm MLC opening | 1.154 | 1.153 |
| 4 | 0.5 mm shift bank A | 1.019 | 1.019 |
| 5 | 1 mm shift bank A | 1.037 | 1.036 |
| 6 | 2 mm shift bank A | 1.074 | 1.071 |
| 7 | 0.5 mm shift bank B | 1.019 | 1.020 |
| 8 | 1 mm shift bank B | 1.039 | 1.040 |
| 9 | 2 mm shift bank B | 1.079 | 1.077 |
| 10 | Segments missing | 0.997 | 0.993 |
Figure 2Predicted fluence (left) for a modulated field reconstructed according to the DynaLog files. The gamma analysis (middle) shows that the DynaLog‐based method is accurate enough to reproduce the delivered fluence (right).
Differences in parameter minimum dose (D98%), maximum dose (D2%), and mean dose (Dmean) for the target volume between the reference plan (original) and the DynaLog‐based reconstructed plan (recons.). Doses are expressed as percentage of the prescription dose. Analysis was performed for 20 clinical cases and four sites (head and neck, prostate, breast, and lung). Maximum root mean square error (RMS) is shown for each plan according DFV application
|
|
|
| |||||
|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
| H&N#1 | 0.035 | 92.4 | 92.3 | 105.6 | 105.5 | 100.0 | 100.0 |
| H&N#2 | 0.059 | 92.6 | 92.6 | 103.0 | 102.9 | 100.3 | 100.2 |
| H&N#3 | 0.056 | 94.7 | 94.6 | 103.7 | 103.7 | 100.0 | 100.1 |
| H&N#4 | 0.056 | 90.5 | 90.5 | 105.1 | 105.1 | 100.0 | 100.0 |
| H&N#5 | 0.032 | 93.7 | 93.4 | 105.3 | 105.0 | 100.0 | 99.8 |
| Pros#1 | 0.033 | 96.5 | 96.5 | 102.0 | 102.1 | 100.0 | 100.1 |
| Prost#2 | 0.053 | 94.9 | 95.0 | 101.9 | 102.0 | 100.0 | 100.1 |
| Prost#3 | 0.028 | 96.0 | 95.9 | 102.3 | 102.3 | 100.0 | 99.9 |
| Prost#4 | 0.049 | 95.3 | 95.4 | 102.6 | 102.6 | 100.0 | 100.0 |
| Prost#5 | 0.044 | 95.3 | 95.3 | 102.8 | 102.8 | 100.0 | 100.0 |
| Breast#1 | 0.024 | 91.8 | 91.6 | 104.7 | 104.6 | 100.0 | 99.9 |
| Breast#2 | 0.022 | 92.5 | 92.7 | 105.8 | 106.0 | 100.0 | 100.2 |
| Breast#3 | 0.030 | 94.0 | 93.8 | 104.6 | 104.4 | 100.0 | 99.8 |
| Breast#4 | 0.017 | 90.6 | 90.8 | 105.9 | 106.3 | 100.0 | 100.5 |
| Breast#5 | 0.030 | 92.5 | 92.4 | 103.2 | 103.1 | 100.0 | 100.0 |
| Lung#1 | 0.037 | 90.4 | 90.3 | 106.0 | 105.9 | 100.0 | 99.9 |
| Lung#2 | 0.007 | 102.4 | 102.4 | 126.0 | 126.0 | 117.0 | 117.1 |
| Lung#3 | 0.020 | 95.1 | 96.1 | 103.5 | 103.4 | 100.0 | 100.0 |
| Lung#4 | 0.015 | 101.6 | 101.7 | 123.4 | 123.5 | 113.7 | 113.8 |
| Lung#5 | 0.021 | 97.5 | 97.3 | 120.8 | 120.8 | 109.3 | 109.2 |
Figure 3DVH‐based analysis between the reference plan (squares) and the DynaLog‐based reconstructed plan (triangles) for an intended error consisting of 0.5 mm MLC opening. Blue: right femoral head; green: left femoral head; brown: rectum; yellow: bladder; orange: target volume.