3D evaluation of 3DVH program using BANG3 polymer gel dosimeter.

Yoichi Watanabe1, Yuji Nakaguchi.   

Abstract

PURPOSE: With the recent introduction of intensity modulated arc therapy techniques, there is an increasing need for validation of treatment delivery in three-dimensional (3D) space. A commercial dosimetry device ArcCHECK™ (Sun Nuclear Corporation, Melbourne, FL, USA) can be used in conjunction with 3DVH program. With this system, one can reconstruct the 3D dose distribution produced in the actual patient. In this work the authors evaluate the relative accuracy of the ArcCHECK™-3DVH system using BANG3 (MGS Research, Guilford, CT, USA) polymer gel dosimeter.
METHODS: About 15-cm diameter and 20-cm long cylindrical phantoms filled with BANG3 was used to simulate a patient, to which a volumetrically modulated arc therapy plan was created with Pinnacle3 treatment planning software (Philips Healthcare, Andover, MA, USA). The plan (76 Gy total in 38 fractions) was designed for prostate radiotherapy using a 6 MV photon beam from an Elekta Synergy linear accelerator (Elekta AB, Stockholm, Sweden). The treatment was delivered to the simulated patient. The same plan was used to irradiate an ArcCHECK™ device with an insert plug. The point dose at the isocenter was measured using a Farmer-type ionization chamber. The measured dose data were imported into the 3DVH program, which generated the 3D dose distributions projected onto the simulated patient. The dose data recorded in the polymer gel were read out using a MRI scanner and the 3D dose distribution delivered to the simulated patient was analyzed and compared with those from the 3DVH program and the Pinnacle3 software. The comparison was accomplished by using the gamma index, overlaying the isodose lines for a set of data on selected planes, and computing dose-volume histogram of structures.
RESULTS: The dose at the center of the ArcCHECK™ device measured with an ionization chamber was 1.82% lower than the dose predicted by Pinnacle3. The 3D dose distribution generated by Pinnacle3 was compared with those obtained by the ArcCHECK™-3DVH system and BANG3. The gamma passing rates for criteria of 3% dose difference, 3 mm distance-to-agreement, and 25% lower dose threshold were 99.1% for the former and 95.7% for the latter. The mean and maximum PTV doses estimated by the 3DVH were 74.0 and 79.3 Gy in comparison to 74.4 and 76.5 Gy with Pinnacle3. Those values for BANG3 measurements were 74.7 and 79.5 Gy. The mean doses to rectum were 40.2, 39.8, and 38.8 Gy for Pinnacle3, 3DVH, and BANG3, whereas the mean doses to the bladder were 26.7, 25.7, and 21.7 Gy, respectively.
CONCLUSIONS: The ArcCHECK™-3DVH system provides an accurate estimation of 3D dose distribution in an actual patient within a clinically meaningful tolerance level. However, both 3DVH and BANG3 showed two noticeable differences from Pinnacle3. First, the measured dose throughout the PTV region was less uniform than Pinnacle3. Second, the dose gradient at the interface between PTV and rectum was steeper than Pinnacle3 prediction. Further investigation may be able to identify the cause for these findings.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23927338     DOI: 10.1118/1.4813301

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Validation of a quick three-dimensional dose verification system for pre-treatment IMRT QA.

Authors:  Yuji Nakaguchi; Fujio Araki; Takeshi Ono; Yuki Tomiyama; Masato Maruyama; Nozomu Nagasue; Yoshinobu Shimohigashi; Yudai Kai
Journal:  Radiol Phys Technol       Date:  2014-09-27

2.  Dosimetric characteristics of a reusable 3D radiochromic dosimetry material.

Authors:  Jong Min Park; So-Yeon Park; Chang Heon Choi; Minsoo Chun; Ji Hye Han; Jin Dong Cho; Jung-In Kim
Journal:  PLoS One       Date:  2017-07-13       Impact factor: 3.240

3.  Evaluation of patient DVH-based QA metrics for prostate VMAT: correlation between accuracy of estimated 3D patient dose and magnitude of MLC misalignment.

Authors:  Noriyuki Kadoya; Masahide Saito; Makoto Ogasawara; Yukio Fujita; Kengo Ito; Kiyokazu Sato; Kazuma Kishi; Suguru Dobashi; Ken Takeda; Keiichi Jingu
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

4.  Validation of an improved helical diode array and dose reconstruction software using TG-244 datasets and stringent dose comparison criteria.

Authors:  Saeed Ahmed; Benjamin Nelms; Jakub Kozelka; Geoffrey Zhang; Eduardo Moros; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

5.  Evaluation of Dose Distribution in Intensity Modulated Radiosurgery for Lung Cancer under Condition of Respiratory Motion.

Authors:  Mee Sun Yoon; Jae-Uk Jeong; Taek-Keun Nam; Sung-Ja Ahn; Woong-Ki Chung; Ju-Young Song
Journal:  PLoS One       Date:  2016-09-20       Impact factor: 3.240

Review 6.  Three-dimensional radiation dosimetry using polymer gel and solid radiochromic polymer: From basics to clinical applications.

Authors:  Yoichi Watanabe; Leighton Warmington; N Gopishankar
Journal:  World J Radiol       Date:  2017-03-28

7.  Comparison of DVH-based plan verification methods for VMAT: ArcCHECK-3DVH system and dynalog-based dose reconstruction.

Authors:  Masahide Saito; Noriyuki Kadoya; Kiyokazu Sato; Kengo Ito; Suguru Dobashi; Ken Takeda; Hiroshi Onishi; Keiichi Jingu
Journal:  J Appl Clin Med Phys       Date:  2017-06-26       Impact factor: 2.102

8.  Dosimetric and radiobiological comparison for quality assurance of IMRT and VMAT plans.

Authors:  Nava Raj Paudel; Ganesh Narayanasamy; Eun Young Han; Jose Penagaricano; Panayiotis Mavroidis; Xin Zhang; Anil Pyakuryal; Dongwook Kim; Xiaoying Liang; Steven Morrill
Journal:  J Appl Clin Med Phys       Date:  2017-08-03       Impact factor: 2.102

9.  Patient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy.

Authors:  Yoshiyuki Katsuta; Noriyuki Kadoya; Yukio Fujita; Eiji Shimizu; Kenichi Matsunaga; Kinya Sawada; Haruo Matsushita; Kazuhiro Majima; Keiichi Jingu
Journal:  Technol Cancer Res Treat       Date:  2017-12-05

10.  Correlation analysis between 2D and quasi-3D gamma evaluations for both intensity-modulated radiation therapy and volumetric modulated arc therapy.

Authors:  Jung-In Kim; Chang Heon Choi; Hong-Gyun Wu; Jin Ho Kim; Kyubo Kim; Jong Min Park
Journal:  Oncotarget       Date:  2017-01-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.