| Literature DB >> 33073897 |
Vivian U Y Chow1, Monica W K Kan1,2, Anthony T C Chan1,2.
Abstract
An in-house trajectory log analysis program (LOGQA) was developed to evaluate the delivery accuracy of volumetric-modulated arc therapy (VMAT) for stereotactic body radiation therapy (SBRT). Methods have been established in LOGQA to provide analysis on dose indices, gantry angles, and multi-leaf collimator (MLC) positions. Between March 2019 and May 2020, 120 VMAT SBRT plans of various treatment sites using flattening filter-free (FFF) mode were evaluated using both LOGQA and phantom measurements. Gantry angles, dose indices, and MLC positions were extracted from log and compared with each plan. Integrated transient fluence map (ITFM) was reconstructed from log to examine the deviation of delivered fluence against the planned one. Average correlation coefficient of dose index versus gantry angle and ITFM for all patients were 1.0000, indicating that the delivered beam parameters were in good agreement with planned values. Maximum deviation of gantry angles and monitor units (MU) of all patients were less than 0.2 degree and 0.03 % respectively. Regarding MLC positions, maximum and root-mean-square (RMS) deviations from planned values were less than 0.6 mm and 0.3 mm respectively, indicating that MLC positions during delivery followed planned values in precise manner. Results of LOGQA were consistent with measurement, where all gamma-index passing rates were larger than 95 %, with 2 %/2 mm criteria. Three types of intentional errors were introduced to patient plan for software validation. LOGQA was found to recognize the introduced errors of MLC positions, gantry angles, and dose indices with magnitudes of 1 mm, 1 degree, and 5 %, respectively, which were masked in phantom measurement. LOGQA was demonstrated to have the potential to reduce or even replace patient-specific QA measurements for SBRT plan delivery provided that the frequency and amount of measurement-based machine-specific QA can be increased to ensure the log files record real values of machine parameters.Entities:
Keywords: log file analysis; patient-specific QA; plan delivery accuracy; stereotactic body radiation therapy; volumetric-modulated arc therapy
Mesh:
Year: 2020 PMID: 33073897 PMCID: PMC7700944 DOI: 10.1002/acm2.13053
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
Fig. 1Reconstruction of fluence maps between two consecutive control points (CPs).
Fig. 2Dose index versus gantry angle.
Fig. 3Gantry angle deviation versus control point.
Fig. 4Monitor unit (MU) deviation versus control point.
Fig. 5Multileaf collimator (MLC) leaf position deviation.
Fig. 6Planned and delivered integrated transient fluence map (ITFM) comparison.
Parameters for verifying the accuracy of plan delivery
| Parameters to be checked by LOGQA | Quantitative Indicators with passing criteria |
|---|---|
| (1) Dose index (fractional monitor unit delivered) versus gantry angle | Correlation coefficient (CC) ≥0.985 |
| (2) Gantry angle deviation versus control point | Maximum deviation ≤0.3 degree |
| (3) Monitor unit (MU) deviation versus control point | Maximum deviation ≤0.04 % |
| (4) Multileaf collimator (MLC) leaf position deviation |
Maximum deviation ≤1 mm Root‐mean‐square (RMS) ≤0.5 mm |
| (5) Integrated transient fluence map (ITFM) | Correlation coefficient (CC) ≥0.985 |
Average error of MLC leaf positions, gantry angles, and monitor unit of 120 VMAT SBRT plans with various treatment sites
| Treatment Site | MLC error (mm) | Gantry angle error (0) | Monitor unit error (%) |
|---|---|---|---|
| Abdomen | 0.1318 ± 0.0184 | 0.1321 ± 0.0268 | 0.0152 ± 0.0075 |
| Liver | 0.1470 ± 0.0182 | 0.1263 ± 0.0127 | 0.0160 ± 0.0044 |
| Lung | 0.1445 ± 0.0200 | 0.1275 ± 0.0158 | 0.0142 ± 0.0022 |
| Pelvis | 0.1339 ± 0.0217 | 0.1287 ± 0.0222 | 0.0126 ± 0.0045 |
| Prostate | 0.1514 ± 0.0078 | 0.0999 ± 0.0165 | 0.0075 ± 0.0040 |
| Spine | 0.1276 ± 0.0112 | 0.0899 ± 0.0056 | 0.0063 ± 0.0012 |
Results of LOGQA and gamma analysis with the intentional leaf positioning errors of 1 mm and 3 mm introduced to a leaf pair of a patient plan for all the control points
| LOGQA | ArcCheckTM measurement | ||||||
|---|---|---|---|---|---|---|---|
|
Maximum leaf deviation (mm) [Preset tolerance ≤ 1 mm] |
RMS (mm) [Preset tolerance ≤ 0.5 mm] |
Passing rate of gamma analysis (%) [Preset tolerance ≥ 95 %] | |||||
| Arc 1 | Arc 2 | Arc 3 | Arc 1 | Arc 2 | Arc 3 | ||
| Original plan compared with delivered plan |
0.5295 [Pass] |
0.5263 [Pass] |
0.5372 [Pass] |
0.1267 [Pass] |
0.1235 [Pass] |
0.1201 [Pass] |
97.8 [Pass] |
| Original plan compared with delivered plan with |
1.5247 [Fail] |
1.4450 [Fail] |
1.5080 [Fail] |
1.0252 [Fail] |
1.0149 [Fail] |
0.9521 [Fail] |
96.9 [Pass] |
| Original plan compared with delivered plan with |
3.5275 [Fail] |
3.4450 [Fail] |
3.5044 [Fail] |
2.8983 [Fail] |
2.7766 [Fail] |
2.6747 [Fail] |
92.7 [Fail] |
Fig. 7Result of LOGQA on ITFM analysis for a patient plan (a) without the introduction of intentional errors, (b) with an intentional MLC positioning error of 1 mm and (c) with an intentional MLC positioning error of 3 mm for all the control points of a leaf pair.
Results of LOGQA and gamma analysis with the introduction of 1‐degree and 2‐degree gantry angle errors in a patient plan for all the control points
| LOGQA | ArcCheckTM measurement | |||
|---|---|---|---|---|
|
Maximum deviation (Degree) [Preset tolerance ≤ 0.3 degree] |
Passing rate of gamma analysis (%) [Preset tolerance ≥ 95 %] | |||
| Arc1 | Arc2 | Arc3 | ||
| Original plan compared with delivered plan |
0.1120 [Pass] |
0.0682 [Pass] |
0.1149 [Pass] |
97.8 [Pass] |
| Original plan compared with delivered plan with |
1.0749 [Fail] |
1.1271 [Fail] |
1.0748 [Fail] |
95.8 [Pass] |
| Original plan compared with delivered plan with |
2.0812 [Fail] |
2.1277 [Fail] |
2.0815 [Fail] |
84.7 [Fail] |
Results of LOGQA and gamma analysis with the introduction of 5 % and 10 % differential dose index errors in a patient plan for 60 control points
| LOGQA | ArcCheckTM measurement | |||
|---|---|---|---|---|
|
Maximum deviation of cumulative dose index (%) [Preset tolerance ≤ 0.04 %] |
Passing rate of gamma analysis (%) [Preset tolerance ≥ 95 %] | |||
| Arc1 | Arc2 | Arc3 | ||
| Original plan compared with delivered plan |
0.0132 [Pass] |
0.0122 [Pass] |
0.0124 [Pass] |
97.8 [Pass] |
| Original plan compared with delivered plan with |
0.3776 [Fail] |
0.3681 [Fail] |
0.3787 [Fail] |
97.1 [Pass] |
| Original plan compared with delivered plan with |
0.7674 [Fail] |
0.7624 [Fail] |
0.7794 [Fail] |
94.4 [Fail] |