| Literature DB >> 29242816 |
Francisco Martins1, Patrícia Branco1, José Reis1, Ignacio Barbero Navarro2, Paulo Maurício1.
Abstract
Objective: Oral rehabilitation success is enhanced by an accurate and reproducible final impression. The purpose of this study is to evaluate the dimensional changes of a polyether and addition silicone subjected to disinfection and/or sterilization after a long storage period. Material and methods: Ninety samples were obtained from polyether ImpregumTM PentaTM (3M ESPETM, Seefeld, Germany) and 90 of addition silicone ImprintTM 4 PentaTM Putty (3M ESPETM, Seefeld, Germany) according to ISO 4823:2000. The samples of each material were split to form three groups with 30 samples each: a control group, a hypochlorite group (disinfection) and an autoclave group (sterilization). Samples were stored in the Portuguese Institute for Quality for six months at 23 °C. Samples were measured by laser interferometry, according to the Michelson technique before calculating dimensional stability according ISO 4823:2000. A statistical analysis via a three-way mixed ANOVA was performed.Entities:
Keywords: Elastomers; laser interferometry; sterilization
Year: 2017 PMID: 29242816 PMCID: PMC5724799 DOI: 10.1080/23337931.2017.1401933
Source DB: PubMed Journal: Acta Biomater Odontol Scand ISSN: 2333-7931
Figure 1.Block test and ring mold made according to ISO 4823:2000.
Figure 2.Superior view of the block test. Lines 1 and 2 have a width of 50 and 20 microns, respectively. Lines 3, 4 and 5 have a width of 75.
Figure 3.Descriptive statistics of PE and VPS groups with mean values and standard deviation at T0 and T1 time level.
Descriptive and three-way ANOVA analysis between the three variables.
| T0 (Mean % ± SD) | T1 (Mean % ± SD) | Shrinkage (Mean % ± SD) | ||
|---|---|---|---|---|
| PE | ||||
| Control | 0.13 ± 0.19 | 0.90 ± 0.26 | 0.77 ± 0.17 | |
| Hypochlorite | 0.22 ± 0.19 | 0.64 ± 023 | 0.42 ± 0.19 | |
| Autoclave | 0.30 ± 0.30 | 0.83 ± 0.32 | 0.52 ± 0.28 | |
| VPS | ||||
| Control | 0.12 ± 0.18 | 0.53 ± 0.19 | 0.42 ± 0.12 | |
| Hypochlorite | 0.16 ± 0.22 | 0.52 ± 0.24 | 0.36 ± 0.09 | |
| Autoclave | 0.30 ± 0.12 | 0.87 ± 0.13 | 0.58 ± 0.13 | |
Identifies a statistically significant differences, for a 95% confidence level.
Identifies a statistically significant interaction, for a 95% confidence level.
Three-way mixed ANOVA – between-subjects effects.
Three-way mixed ANOVA – within-subjects effects.
Statistically significant differences with autoclave group – pairwise comparisons.
Statistically significant differences with hypochlorite group – pairwise comparisons.
Statistically significant differences with VPS material – pairwise comparisons.