| Literature DB >> 28786239 |
Per H Halvorsen1, Eileen Cirino1, Indra J Das2, Jeffrey A Garrett3, Jun Yang4, Fang-Fang Yin5, Lynne A Fairobent6.
Abstract
The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education, and professional practice of medical physics. The AAPM has more than 8,000 members and is the principal organization of medical physicists in the United States. The AAPM will periodically define new practice guidelines for medical physics practice to help advance the science of medical physics and to improve the quality of service to patients throughout the United States. Existing medical physics practice guidelines will be reviewed for revision or renewal, as appropriate, on their fifth anniversary or sooner. Each medical physics practice guideline represents a policy statement by the AAPM, has undergone a thorough consensus process in which it has been subjected to extensive review, and requires the approval of the Professional Council. The medical physics practice guidelines recognize that the safe and effective use of diagnostic and therapeutic radiology requires specific training, skills, and techniques, as described in each document. Reproduction or modification of the published practice guidelines and technical standards by those entities not providing these services is not authorized. The following terms are used in the AAPM practice guidelines: Must and Must Not: Used to indicate that adherence to the recommendation is considered necessary to conform to this practice guideline. Should and Should Not: Used to indicate a prudent practice to which exceptions may occasionally be made in appropriate circumstances. Approved by AAPM Professional Council 3-31-2017 and Executive Committee 4-4-2017.Entities:
Mesh:
Year: 2017 PMID: 28786239 PMCID: PMC5874865 DOI: 10.1002/acm2.12146
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Minimum SRS‐SBRT relevant equipment QA and tolerances for C‐arm linac systems
| Frequency | Test | Tolerance |
|---|---|---|
| Daily | Laser localization — only if using SRS techniques relying on lasers for target localization (e.g., frame‐based SRS without X‐ray IGRT) | 1 mm |
| Collimator size indicator for clinically relevant aperture | 2 mm total | |
| Radiation isocentricity test (limited gantry and couch positions) — maximum deviation in center of target object relative to each projection's beam central axis | 1.0 mm SRS, 1.5 mm SBRT | |
| IGRT positioning/repositioning | 1 mm SRS, 2 mm SBRT | |
| Imaging subsystem interlocks | Functional | |
| Stereotactic interlocks — cone size, backup jaws | Functional | |
| Accelerator output constancy | ±3% | |
| Monthly |
Radiation isocentricity test — covering complete range of gantry, couch, collimator positions used clinically — | 1.0 mm SRS, 1.5 mm SBRT |
| Treatment couch position indicators: relative over the maximum clinical range | 1 mm/0.5° | |
| Output constancy at relevant dose rates | 2% | |
| Annually | SRS arc rotation mode (if used clinically) | 1 MU, 1° |
| MU linearity (≥5 MU to highest MU used clinically) | ±2% | |
| Accelerator output | ±1.5% | |
| Coincidence of radiation and mechanical isocenter | ±1.0 mm maximum 3‐D displacement from center of target object | |
| Verification of small‐field beam data — relative output factors for cones and/or MLC | ±2% from baseline for >1.0 cm apertures, ±5% from baseline for ≤1.0 cm apertures | |
| E2E localization assessment “hidden target test” using SRS frame and/or IGRT system | 1.0 mm | |
| E2E dosimetric evaluation using SRS frame and/or IGRT system | ±5% measured vs. calculated |
Tolerances are absolute accuracy, not variation from baseline, unless otherwise stated.
Minimum SBRT relevant equipment QA and tolerances for helical tomotherapy systems
| Frequency | Test | Tolerance |
|---|---|---|
| Daily | Red laser initialization (congruence with green laser) | 1 mm |
| Image/laser coordinate coincidence | 1 mm | |
| Image registration/alignment | 1 mm | |
| Accelerator output constancy (rotational or static) | ±3% | |
| Monthly | Transverse beam profile | 1% average difference in field core |
| Longitudinal beam profile (each slice width) | 1% of slice width FWHM | |
| Output constancy and rotational output variation | ±2% | |
| Beam quality constancy | ±1% PDD10 or TMR20 10 | |
| Red and green laser alignment | 1 mm | |
| Couch positioning accuracy | 1 mm | |
| CT dimensional accuracy | 1 mm | |
| Annually | Couch speed uniformity | ±2% dose nonuniformity |
| Couch translation per gantry rotation | 1 mm per 5 cm | |
| Accelerator output | ±1.5% | |
| Beam quality (each slice width) | ±1% PDD10 or TMR20 10 | |
| Verification of small‐field beam data | ±2% from baseline for >1.0 cm apertures, ±5% from baseline for ≤1.0 cm apertures | |
| CT imaging — treatment — laser coordinate coincidence | 1.0 mm | |
| E2E localization assessment “hidden target test” | 1.0 mm | |
| E2E dosimetric evaluation | ±5% measured vs. calculated |
SRS is not included in the scope of this document for helical tomotherapy.
Tolerances are absolute accuracy, not variation from baseline, unless otherwise stated.
Minimum equipment QA and tolerances for robotic linac systems
| Frequency | Test | Tolerance |
|---|---|---|
|
Daily* | Head laser alignment check | 1.0 mm |
| Safety interlocks | Functional | |
|
Automatic QA (AQA) test* | Total targeting ≤1.0 mm from baseline, not exceeding manufacturer's specification | |
| Accelerator output constancy | ±3% | |
| Monthly | Energy constancy | ±2% |
| Beam symmetry, relative | ±3% for 40 mm field, ±4% for 60 mm field | |
| Accelerator output constancy | ±2% | |
| Imager alignment | 1 mm or center pixels ±2 pixels | |
| IrisTM field size spot check (3 or more field sizes ≥10 mm) | 0.5 mm | |
| Picket fence for MLC (if applicable) | Visual check | |
| Quarterly |
E2E localization assessment | 1.0 mm static target, 1.5 mm motion tracking |
| Annually | Emergency Power Off (EPO) button, safety interlocks | Functional |
| Accelerator output | ±2.0% | |
| MU linearity (>10 MU to highest MU used clinically) | ±2% | |
| Path verification | ≤0.5 mm maximum per node, ≤0.3 mm average | |
| Imager kVp accuracy, mA station exposure linearity, isopost alignment with center pixel | ±10%, ±20%, and 1 mm, respectively | |
| Beam laser and radiation beam alignment for cone, IrisTM, and MLC | 0.5 mm from baseline | |
| AQA baseline | Recheck AQA baseline | |
| Beam data verification — Relative output factors for cones, IrisTM, and/or MLC covering the range used clinically | ±2% from baseline for >1.0 cm apertures, ±5% from baseline for ≤1.0 cm apertures |