| Literature DB >> 26699330 |
Jennifer B Smilowitz1, Indra J Das, Vladimir Feygelman, Benedick A Fraass, Stephen F Kry, Ingrid R Marshall, Dimitris N Mihailidis, Zoubir Ouhib, Timothy Ritter, Michael G Snyder, Lynne Fairobent.
Abstract
The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education and professional practice of medical physics. The AAPM has more than 8,000 members and is the principal organization of medical physicists in the United States. The AAPM will periodically define new practice guidelines for medical physics practice to help advance the science of medical physics and to improve the quality of service to patients throughout the United States. Existing medical physics practice guidelines will be reviewed for the purpose of revision or renewal, as appropriate, on their fifth anniversary or sooner. Each medical physics practice guideline represents a policy statement by the AAPM, has undergone a thorough consensus process in which it has been subjected to extensive review, and requires the approval of the Professional Council. The medical physics practice guidelines recognize that the safe and effective use of diagnostic and therapeutic radiology requires specific training, skills, and techniques, as described in each document. Reproduction or modification of the published practice guidelines and technical standards by those entities not providing these services is not authorized. The following terms are used in the AAPM practice guidelines:• Must and Must Not: Used to indicate that adherence to the recommendation is considered necessary to conform to this practice guideline.• Should and Should Not: Used to indicate a prudent practice to which exceptions may occasionally be made in appropriate circumstances.Entities:
Mesh:
Year: 2015 PMID: 26699330 PMCID: PMC5690154 DOI: 10.1120/jacmp.v16i5.5768
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Workflow of TPS dose algorithm commissioning, validation, and routine QA. The numbers refer to sections of this report.
Detectors suitable for TPS commissioning and validation of photon and electron beams
|
|
|
|
|
|---|---|---|---|
| Scanning ion chambers | Beam scanning for photons and electrons | Typical scanning chambers have an air cavity of | TG‐106 (Das et al. |
| Electron diodes and film | Beam scanning for electrons, output factors (film) | QMP must confirm the effective point of measurement | TG‐25 (Khan et al. |
| Small field detectors | • Small field scanning & output factors | Carefully select the detector type and size to fit the application. | TG‐106 (Das et al. |
| • IMRT/VMAT point measurement | When scanning for penumbra, diodes are recommended. | ||
| • MLC intraleaf measurement & penumbra | |||
| Large ion chamber | Aggregate MLC transmission factors | Interleaf transmission | LoSasso et al. |
| Film and/or array detector | 2D dose distributions, including dynamic/virtual wedge and planar fluence | • Absolute dosimetry preferred; relative dosimetry adequate. | TG‐106 (Das et al. |
| • Desirable if the device can be mounted on the gantry and/or in a phantom at different geometries |
If a diode detector is used for small field measurements, a “daisy chain” approach is recommended to minimize the energy dependence effects; the diode is first cross‐compared with an ion chamber for a field and then is used to measure the smaller fields.
Using film for intraleaf transmission is usually less precise than interleaf transmission.
Equipment required for TPS commissioning of photon and electron beams
|
|
|
|
|
|---|---|---|---|
| 3D water phantom | Beam scanning | Must have sufficient scanning range and lateral/depth scatter | TG‐106 (Das et al. |
| Electrometers and cables | Beam scanning, output calibration, relative and absolute dosimetry | ADCL calibration, low noise and leakage with wide dynamic range and linear response | TG‐106 (Das et al. |
| Buildup cap or miniphantom | In‐air output factor measurement | Measurements required for some planning systems, most second check systems | Yunice, et al. |
| Water‐equivalent phantom material in slab form | Buildup and backscatter for measurements |
| TG‐106 (Das et al. |
| CT density phantom | CT number to electron or mass density calibration | Should include tissue‐equivalent materials spanning the clinical range of low‐density lung to high‐density bone. | TG‐66 (Mutic et al. |
| Heterogeneity phantom with lung‐equivalent material | End‐to‐end testing | Include cavities for detectors, useful for annual QA reference test | TG‐65 (Papanikolaou & Stathakis |
| Anthropomorphic phantom | Anatomic model testing, end‐to‐end testing, use testing | Include cavities for detectors | IAEA TRS‐430 |
| Software for data processing | Processing, comparing, and analyzing profiles, depth‐dose curves, and other beam data | May be included with the 3D water tank scanning software | TG‐106 (Das et al. |
| IMRT/VMAT or arc therapy phantom | VMAT or arc therapy | Options include a solid phantom holding a planar array, 3D detector arrays, film inside a phantom, other | TG‐120 (Low et al. |
TPS model comparison tests and tolerances
|
|
|
|
|
|---|---|---|---|
| 5.1 | Dose distributions in planning module vs. modeling (physics) module | Comparison of dose distribution for large ( | Identical |
| 5.2 | Dose in test plan vs. clinical calibration condition | Reference calibration condition check | 0.5% |
| 5.3 | Dose distribution calculated in planning system vs. commissioning data | PDD and off axis output factors for a large and a small field size | 2% |
Identical to within the expected statistical uncertainty (considering noise and calculation grid size).
TPS absolute dose at reference point.
Basic photon beam validation tests summary
|
|
|
|
|---|---|---|
| 5.4 | Small MLC‐shaped field (non SRS) | Photon Test 1 |
| 5.5 | Large MLC‐shaped field with extensive blocking (e.g., mantle) | Photon Test 3 |
| 5.6 | Off‐axis MLC shaped field, with maximum allowed leaf over travel | Photon Test 2 |
| 5.7 | Asymmetric field at minimal anticipated SSD | Photon Test 6 |
| 5.8 |
| Photon Test 10 |
| 5.9 | Large ( | – |
For all tests, measurements in the high‐dose region, penumbra, and low‐dose tail regions should be compared to calculated values at various depths (including slightly beyond dmax, midrange/10–15 cm, and deep/25–30 cm). SSDs, other than those used at commissioning and that reflect the clinically expected range, should be used. The MLC should be used for tests 5.4–5.6. The MLC or jaws may be used for tests 5.7–5.9.
Tests 5.4–5.8 are intended for each open and (hard) wedged field. Nonphysical wedges are considered an extension of the corresponding open field in terms of spectra and only require the addition of Test 5.9.
Basic TPS photon beam evaluation methods and tolerances
|
|
|
|
|---|---|---|
| High dose | Relative dose with one parameter change from reference conditions | 2% |
| Relative dose with multiple parameter changes | 5% | |
| Penumbra | Distance to agreement | 3 mm |
| Low‐dose tail | Up to 5 cm from field edge | 3% of maximum field dose |
Tolerances are relative to local dose unless otherwise noted.
For example, off‐axis with physical wedge.
Heterogeneous TPS photon beam validation tests
|
|
|
|
|
|
|---|---|---|---|---|
| 6.1 | Validate planning system reported electron (or mass) densities against known values | CT‐density calibration for air, lung, water, dense bone, and possibly additional tissue types | – | TG 65, |
| 6.2 | Heterogeneity correction distal to lung tissue |
| 3% | IAEA TRS‐430, |
Tolerances are relative to local dose unless otherwise noted.
VMAT/IMRT test summary
|
|
|
|
|
|
|---|---|---|---|---|
| 7.1 | Verify small field PDD |
| Diode or plastic scintillator | Yunice et al.(16) |
| 7.2 | Verify output for small MLC‐defined fields | Use small square and rectangular MLC‐defined segments, measuring output at a clinically relevant depth for each | Diode, plastic scintillator, minichamber or microion chamber | Cadman et al. |
| 7.3 | TG‐119 tests | Plan, measure, and compare planning and QA results to the TG119 report for both the Head and Neck and C‐shape cases | Ion chamber, film and/or array | TG‐119 (Ezzell et al. |
| 7.4 | Clinical tests | Choose at least 2 relevant clinical cases; plan, measure, and perform an in‐depth analysis of the results | Ion chamber, film and/or array | Nelms et al. |
| 7.5 | External review | Simulate, plan, and treat an anthropomorphic phantom with embedded dosimeters. | Various options exist | Kry et al. |
A bar pattern scanned with a diode can be used to obtain additional absolute dose profile comparison in the direction perpendicular to MLC movement
If IROC Houston service is used, they typically employ TLDs and radiochromic film. Certain commercial phantoms can accommodate ion chambers for point dose measurements
VMAT/IMRT evaluation methods and tolerances
|
|
|
|
|---|---|---|
| Ion Chamber | Low‐gradient target region OAR region | 2% of prescribed dose |
| 3% of prescribed dose | ||
| Planar/Volumetric Array | All regions |
|
| End‐to‐End | Low‐gradient target region | 5% of prescribed dose |
Application of a gamma criterion can result in the discovery of easily correctable problems with IMRT commissioning that may be hidden in the higher (and ubiquitous) passing rates.
Basic TPS validation tests for electron beams and minimum tolerance values
|
|
|
|
|
|---|---|---|---|
| 8.1 | Basic model verification with shaped fields | Custom cutouts at standard and extended SSDs |
|
| 8.2 | Surface irregularities obliquity | Oblique incidence using reference cone and nominal clinical SSD | 5% |
| 8.3 | Inhomogeneity test | Reference cone and nominal clinical SSD | 7% |
|
|
|
|
|---|---|---|
| 1 | QMP understands algorithms and has received proper training. | |
| 3 | Manufacturer's guidance for data acquisition was consulted and followed. | |
| 3.B | Appropriate CT calibration data acquired. | |
| 3.D | Review of raw data (compare with published data, check for error, confirm import into TPS). | |
| 4 | Beam modeling process completed according to manufacturer's instructions. | |
| 4 | Beam models evaluated qualitatively and quantitatively using metrics within the modeling software. | |
| 5 | Basic photon beam validation: Tests 5.1–5.8 (5.9 for nonphysical wedge). | |
| 6 | Heterogeneity correction validation for photon beams: Tests 6.1–6.2 | |
| 7 | IMRT/VMAT validation: Tests 7.1–7.4 | |
| 7 | IMRT/VMAT End‐to‐End test with external review: Test 7.5 | |
| 7 | Understand and document limitations of IMRT/VMAT modeling and dose algorithms. | |
| 8 | Electron validation: Tests 8.1–8.3 | |
| 9 | Baseline QA plan(s) (for model constancy) identified for each configured beam and routine QA established. | |