Jinyu Xue1, Hesheng Wang1, David Barbee1, Matthew Schmidt2, Indra J Das1.
Abstract
Dosimetric accuracy of a volumetric modulated arc therapy (VMAT) plan is directly related to the beam model, particularly with multileaf collimator characterization. Inappropriate dosimetric leaf gap (DLG) value can lead to a suboptimal beam model, with significant failure in patient-specific quality assurance (PSQA) of VMAT plans. This study addressed the systematic issue of beam modeling and developed a practical method to determine the optimal DLG value for a beam model. Several complex VMAT plans were selected for the quality assurance analysis using the variable DLG values. The results of three-dimensional (3D) Gamma analysis as a function of the DLG at 3%/3 mm, 2%/2 mm, and 1%/1 mm criteria were fitted by a polynomial curve. The DLG value corresponding to the maximum Gamma passing rate for each polynomial fitting function was derived, and the average was calculated to be the optimal DLG value for each model. The 3D Gamma analysis was repeated with the optimal DLG value to verify the dosimetric accuracy of each VMAT case by PSQA. Gamma passing rates are seen to vary considerably with the DLG values and different analysis criteria (3%/3 mm, 2%/2 mm, and 1%/1 mm) for each case. The optimal DLG derived for each model was 1.16 mm and 1.10 mm, much larger than the measured value (about 0.3 mm). The beam models with the optimal DLG was able to produce an average Gamma passing rate of 97.1% (range, 94.6%- 99.1%) at 3%/3 mm and 93.5% (range, 89.0%- 96.5%) at 2%/2 mm for one beam model, and 97.1% (range, 94.8%- 99.1%) at 3%/3 mm, and 93.3% (range, 88.8%- 96.7%) at 2%/2 mm for another. The overall accuracy of dose calculation for VMAT plans should be optimized with a compromise of varied modulation complexities in a beam model. We have developed a practical method to derive the optimal DLG value for each beam model based on the Gamma passing criterion. This technique should be applicable in general for all beam energies and patient cases.Entities:
Keywords: Beam model; dosimetric leaf gap; quality assurance; treatment planning system; volumetric modulated arc therapy
Year: 2018 PMID: 29962688 PMCID: PMC6020624 DOI: 10.4103/jmp.JMP_144_17
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Summary of some dosimetric parameters for two models
Summary of volumetric modulated arc therapy plans
Numeric Gamma passing rate in percentage for case #1
Figure 1Gamma passing rate for 4 lung volumetric modulated arc therapy cases plotted together (blue: #1, burgundy: #2, green: #3 and purple: #4) at 3%/3 mm, 2%/2 mm and 1%/1 mm criteria. Beam model #1 was used
Figure 2Gamma passing rate for the same lung volumetric modulated arc therapy case at 3%/3 mm, 2%/2 mm and 1%/1 mm for both models #1 and #2, respectively. Solid circles (blue) for model #1, diamond (orange) for model #2. Solid line for 3%/3 mm, dash for 2%/2 mm, dotted for 1%/1 mm
Example of polynomial fit to the data
Dosimetric leaf gap (mm) for the maximum Gamma passing rate of each test set with the average and the standard deviation