A patient-specific QA comparison between 2D and 3D diode arrays for single-lesion SRS and SBRT treatments.

Yongsook C Lee1, Yongbok Kim2.   

Abstract

The purpose of this study is to compare patient-specific quality assurance (PSQA) results between two dimensional (2D) diode (SRS MapCHECK®) and 3D diode (ArcCHECK®) arrays. Twenty-eight intracranial stereotactic radiosurgery (SRS) and 26 lung stereotactic body radiation therapy (SBRT) clinical plans with a single lesion were selected and categorized into 4 groups: 20 SRS dynamic conformal arc therapy (DCAT) plans (Group A), 8 SRS volumetric modulated arc therapy (VMAT) plans (Group B), 6 SBRT DCAT plans (Group C) and 20 SBRT VMAT plans (Group D). An individual field of each plan was delivered on SRS MapCHECK and ArcCHECK and QA analysis was performed using 4 gamma criteria of dose difference/distance-to-agreement of 3%/3 mm, 3%/2 mm, 2%/2 mm and 2%/1 mm. Statistical analysis was performed to compare PSQA results between the 2 QA devices. For all 4 groups and all 4 gamma criteria, average gamma passing rates were higher with SRS MapCHECK.
© 2021 Old City Publishing, Inc.

Entities:  

Keywords:  ArcCHECK; Patient-specific quality assurance (PSQA); SRS MapCHECK; stereotactic body radiation therapy (SBRT); stereotactic radiosurgery (SRS)

Year:  2021        PMID: 34631231      PMCID: PMC8492049     

Source DB:  PubMed          Journal:  J Radiosurg SBRT


  21 in total

1.  Comparison of 2D and 3D gamma analyses.

Authors:  Kiley B Pulliam; Jessie Y Huang; Rebecca M Howell; David Followill; Ryan Bosca; Jennifer O'Daniel; Stephen F Kry
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

2.  A study on correlation between 2D and 3D gamma evaluation metrics in patient-specific quality assurance for VMAT.

Authors:  Dhanabalan Rajasekaran; Prakash Jeevanandam; Prabakar Sukumar; Arulpandiyan Ranganathan; Samdevakumar Johnjothi; Vivekanandan Nagarajan
Journal:  Med Dosim       Date:  2014-06-06       Impact factor: 1.482

3.  Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance.

Authors:  Mark S Rose; Lena Tirpak; Kristin Van Casteren; Jeff Zack; Tom Simon; Andreas Schoenfeld; William Simon
Journal:  Med Phys       Date:  2020-04-27       Impact factor: 4.071

Review 4.  Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218.

Authors:  Moyed Miften; Arthur Olch; Dimitris Mihailidis; Jean Moran; Todd Pawlicki; Andrea Molineu; Harold Li; Krishni Wijesooriya; Jie Shi; Ping Xia; Nikos Papanikolaou; Daniel A Low
Journal:  Med Phys       Date:  2018-03-23       Impact factor: 4.071

5.  Evaluation and Performance of ArcCheck and Film using Gamma Criteria in Pre-treatment Quality Assurance of Stereotactic Ablative Radiotherapy.

Authors:  Prabhakar Ramachandran; Abdulrahman Tajaldeen; David Taylor; Derrick Wanigaratne; Karl Roozen; Moshi Geso
Journal:  J Med Phys       Date:  2017 Oct-Dec

6.  Comparison of volumetric-modulated arc therapy and dynamic conformal arc treatment planning for cranial stereotactic radiosurgery.

Authors:  Jessica Molinier; Christine Kerr; Sebastien Simeon; Norbert Ailleres; Marie Charissoux; David Azria; Pascal Fenoglietto
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

7.  Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.

Authors:  Saeed Ahmed; Geoffrey Zhang; Eduardo G Moros; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2019-09-03       Impact factor: 2.102

8.  Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D.

Authors:  James McCulloch; Jamie Pawlowski; Neil Kirby; Karl Rasmussen; Zheng Shi; Pamela Myers; Sotirios Stathakis; Niko Papanikolaou; Daniel L Saenz
Journal:  J Appl Clin Med Phys       Date:  2020-07-18       Impact factor: 2.102

9.  Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric-modulated arc therapy (VMAT).

Authors:  Guangjun Li; Sen Bai; Nianyong Chen; Lansdale Henderson; Kui Wu; Jianghong Xiao; Yingjie Zhang; Qingfeng Jiang; Xiaoqin Jiang
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

10.  Comparison of 3D and 2D gamma passing rate criteria for detection sensitivity to IMRT delivery errors.

Authors:  Dandan Zhang; Bin Wang; Guangshun Zhang; Charlie Ma; Xiaowu Deng
Journal:  J Appl Clin Med Phys       Date:  2018-06-15       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.