Repeatability of apparent diffusion coefficient and intravoxel incoherent motion parameters at 3.0 Tesla in orbital lesions.

Augustin Lecler1,2, Julien Savatovsky3, Daniel Balvay4, Mathieu Zmuda5, Jean-Claude Sadik3, Olivier Galatoire5, Frédérique Charbonneau3, Olivier Bergès3, Hervé Picard6, Laure Fournier4,7.   

Abstract

OBJECTIVES: To evaluate repeatability of intravoxel incoherent motion (IVIM) diffusion-weighted imaging (DWI) parameters in the orbit.
METHODS: From December 2015 to March 2016, 22 patients were scanned twice using an IVIM sequence with 15b values (0-2,000 s/mm2) at 3.0T. Two readers independently delineated regions of interest in an orbital mass and in different intra-orbital and extra-orbital structures. Short-term test-retest repeatability and inter-observer agreement were assessed using the intra-class correlation coefficient (ICC), the coefficient of variation (CV) and Bland-Altman limits of agreements (BA-LA).
RESULTS: Test-retest repeatability of IVIM parameters in the orbital mass was satisfactory for ADC and D (mean CV 12% and 14%, ICC 95% and 93%), poor for f and D*(means CV 43% and 110%, ICC 90% and 65%). Inter-observer repeatability agreement was almost perfect in the orbital mass for all the IVIM parameters (ICC = 95%, 93%, 94% and 90% for ADC, D, f and D*, respectively).
CONCLUSIONS: IVIM appeared to be a robust tool to measure D in orbital lesions with good repeatability, but this approach showed a poor repeatability of f and D*. KEY POINTS: • IVIM technique is feasible in the orbit. • IVIM has a good-acceptable repeatability of D (CV range 12-25 %). • IVIM interobserver repeatability agreement is excellent (ICC range 90-95 %). • f or D* provide higher test-retest and interobserver variabilities.

Entities:  

Keywords:  DWI; Feasibility; IVIM; MRI; Orbit

Mesh:

Year:  2017        PMID: 28677061      PMCID: PMC5674133          DOI: 10.1007/s00330-017-4933-6

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


Introduction

Diffusion-weighted imaging (DWI) using magnetic resonance imaging (MRI) was described by Le Bihan in 1986 as being sensitive to displacement of water protons in tissues due to both random Brownian motion as well as capillary perfusion [1, 2]. Intravoxel incoherent motion (IVIM) imaging uses multiple b values with a biexponential model to quantify both phenomena. The DW signal is the result of a ‘pure’ diffusion fraction and a perfusion-dominated pseudo-diffusion fraction, yielding three parameters: the ‘pure’ diffusion coefficient (D), the perfusion fraction (f) and the pseudo-diffusion coefficient (D*) [1]. This model has been shown to provide a pertinent description of the DW signal measured in highly vascular organs [3]. Due to both improvement of MR devices and post-treatment softwares, IVIM has become more available in clinical practice. It has been tested and proved useful in a variety of organs, such as prostate, liver, abdomen, kidney and pelvis, and also in the head and neck [4-8], to diagnose liver diseases, characterise tumours, or assess and monitor tissue response to treatment [4-8]. However, one issue is repeatability, which has been evaluated over time in brain imaging [9, 10] or abdominal imaging [11-15]. In the orbit, only a few studies evaluated the DWI, showing its usefulness when visualising and diagnosing tumours [16, 17]. An ADC value <1 10-3 s/mm2 at b=1,000 s/mm2 were considered optimal thresholds to predict overall malignancy [17] and a low ADC <0.6 10-3 s/mm2 at b=1,000 s/mm2 was reported to have good accuracy in distinguishing benign from malignant orbital lymphoproliferative disorders [18]. Also, two recent studies combining morphological characteristics to DWI and dynamic contrast-enhanced perfusion showed that ADC alone yielded the optimal sensitivity in differentiating malignant from benign orbital lymphoproliferative disorders [19, 20]. However, IVIM has never been tested in the orbits, and it makes sense to test IVIM repeatability in orbital anatomy/disease in order to understand technical limitations before its use in future studies. The aim of our study was to prospectively evaluate robustness of IVIM-derived parameters in the orbit by evaluating the short-term test-retest and inter-observer repeatability of IVIM parameters and apparent diffusion coefficient (ADC) of orbital lesions and normal intra-orbital and extra-orbital structures at 3.0T.

Materials and methods

Study design

We conducted a prospective study in a tertiary referral centre specialising in ophthalmic diseases (NCT02401906). This study was prospectively approved by our institutional Research Ethics Board and adhered to the tenants of the Declaration of Helsinki (IRB 2015-A00364-45). Signed informed consent was obtained from all subjects. From December 2015 through March 2016, 22 patients were included in the study. Inclusion criteria were: (a) age over 18 years; (b) presence of an orbital mass. Patients with an MR contraindication such as implanted pacemakers, the presence of other metallic foreign bodies or claustrophobia were excluded.

MR imaging

All MRIs were performed on the same 3 Tesla Philips INGENIA device with a 32-channel head coil (Philips Medical Systems, Best, The Netherlands). All patients had the same MRI protocol including two IVIM acquisitions acquired with 15 b values, ranging from 0 to 2,000 s/mm2. The b values distribution was chosen to cover both the initial pseudodiffusion decay (≤200 s/mm2) and the molecular diffusion decay (>200 s/mm2). We used a large number of lower b values for more accurate calculation of IVIM parameters. The first acquisition was acquired at the beginning of the examination and the second one immediately before contrast injection, with a median delay between the two scanning procedures of 17 min, providing two distinct data sets of IVIM. Patients were asked to look at a fixed point during the acquisitions in order to prevent kinetic artefacts generated from eye movements. They were also asked to move between the two sequences and were repositioned for the second IVIM acquisition. Technical specifications of the MRI protocol are provided in Table 1.
Table 1

Specifications of the intravoxel incoherent motion (IVIM) sequence parameters

IVIM
TR/TE (ms)2,639 / 74
Section thickness (mm)3
Number of slices210
Gap (mm)1
Bandwith (Herz)1,695
Number of excitations1
Number of b-values15
b values (s/mm2 )0, 10, 20, 40, 80, 110, 140, 170, 200, 300, 400, 600, 800, 1,000, 2,000
Field of view (mm2)122 × 122
Matrix176 × 122
Parallel acceleration factorSENSE = 2
Acquisition duration (s)143
Specifications of the intravoxel incoherent motion (IVIM) sequence parameters

Image analysis

Two radiologists, blinded to patient ID, medical history and laboratory results, read independently and in random order the IVIM sequences. The first senior neuroradiologist was specialized in orbital imaging with 7 years of experience (AL), and the second was a senior radiologist with no experience in orbital imaging (LF). All the post-processing steps were performed using the Olea Sphere® software (v3.0, Olea Medical, La Ciotat, France), implementing a Bayesian probability-based algorithm using all 15 b-values to fit a biexponential diffusion model to each voxel within a region of interest (ROI) for the calculation of IVIM parameters (D, D* and f), and a monoexponential diffusion model for the calculation of ADC (Fig. 1).
Fig. 1

A 47-year-old man with a cavernous haemangioma in his left orbit (asterisk), displacing the eye anteriorly, as shown on the axial post-contrast T1 WI (a). Placement of a freehand large region of interest (ROI) (yellow line) and of a circular small ROI (red circle) inside the mass are shown (b). Test (c) and retest (d) bi-exponential fitting curves using b values are shown. Coloured intravoxel incoherent motion (IVIM) parametric map is displayed (e)

A 47-year-old man with a cavernous haemangioma in his left orbit (asterisk), displacing the eye anteriorly, as shown on the axial post-contrast T1 WI (a). Placement of a freehand large region of interest (ROI) (yellow line) and of a circular small ROI (red circle) inside the mass are shown (b). Test (c) and retest (d) bi-exponential fitting curves using b values are shown. Coloured intravoxel incoherent motion (IVIM) parametric map is displayed (e) The operators independently drew seven ROIs on each imaging set: two ROIs inside the orbital mass, the first one encompassing the maximum area of the orbital mass (ROI 1) and the second one a circular ROI at the centre of the orbital mass without consideration for the zones with restricted diffusion or with enhancing nodules if present (ROI 2) (Fig. 1), three ROIs in non-tumoral orbital structures: in the contralateral lacrymal gland (ROI 3), in the contralateral medial (ROI 4) and lateral (ROI 5) extra-ocular muscles, and two ROIs outside the orbit: one ROI in the temporal muscle (ROI 6) and one ROI in the contralateral temporal lobe white matter (ROI 7). The five last ROIs were drawn in order to analyse their repeatability, in case they might be used in further studies as structures of reference to normalise IVIM values (as is done sometimes in brain studies [10, 21]). Size and location of the ROIs in the orbital and encephalic locations was kept constant across patients to every possible extent, except for ROI 1, which encompassed the largest possible portion of the mass on the slice including its largest diameter. ROIs were placed on the transverse b0 image and propagated to all b-values, ADC and IVIM parametric maps. Readers were free to review the other structural images to ease the drawing of the ROI. A direct registration between the structural and the IVIM images was not possible due to the distortions seen on the IVIM images. In addition, given that parallel imaging was used for DWI, we used the difference method to estimate the signal to noise ratio (eSNR) in the orbital mass [22].

Statistical analysis

Data were analysed using the R software package (R Foundation for Statistical Computing, Vienna, Austria). Mean IVIM values and CVs were computed for each reader individually, then averaged for both readers. Test-retest repeatability of IVIM parameters was assessed by calculating the coefficient of variation (CV, computed as standard deviation (SD) divided by the mean), the intra-class correlation coefficient (ICC), and the 95% Bland-Altman limits of agreements (BA-LA) [23, 24] for each reader individually, then averaged for both readers, as per the recommendations of the Quantitative Imaging Biomarkers Alliance (QIBA) [25]. The parameters’ repeatability was defined as excellent when CV was ≤10%, good when CV was between 10–20%, acceptable when CV was between 20–30%, and poor when CV was >30% [24]. The inter-observer repeatability agreement was assessed by calculating the ICC and was interpreted as follows: 0.0–0.2:poor correlation; 0.21–0.4:fair correlation ;0.41–0.6:moderate correlation ; 0.61–0.8:good correlation ;0.81–1:almost perfect correlation [23].

Results

Twenty-two consecutive patients were included in the study (nine males and 13 females, median age 51 years, range 41–62). Six patients had a histologically-proven orbital lymphoma, six patients had an orbital inflammation, four patients had a cavernous haemangioma, two had a dacryoadenitis, two had an orbital metastasis, one had a muscular granuloma and one had an orbital sarcoma. The quality of ADC and IVIM parametric maps was considered good for subsequent analysis for all patients, and no patient had to be secondarily excluded from the study because of artefacts masking or distorting orbital tumours. Mean (SD) ROI size was 86 (36) pixels for ROI 1, 23.5 (9) for ROI 2, 21 (9) for ROI 3, 27 (10) for ROI 4, 22.5 (8) for ROI 5, 127.5 (48) for ROI 6 and 601 (114) for ROI 7. Mean (SD) IVIM parameter values inside all orbital lesions were calculated as follows: 1.26 (0.35), 1.02 (0.27), 0.18 (0.07) and 14.6 (7.9) for ADC, D, f and D*, respectively. The mean IVIM parameters values and SDs inside the other ROIs are shown in Table 2. The average eSNR inside the orbital mass at b=2,000 s/mm2 was 56.6 (SD 3.5)
Table 2

Mean intravoxel incoherent motion (IVIM) parameter values (of all measurements)

Mean values
Anatomical regionADC (10-3 s/mm2) (SD)D (10-3 s/mm2) (SD) f (%) (SD)D*(10-3 s/mm2) (SD)
ROI 1. Freehand large ROI inside the mass1.26 (0.35)1.02 (0.27)0.18 (0.07)14.6 (7.9)
ROI 2. Circular small ROI inside the mass1.25 (0.38)1.02 (0.3)0.17 (0.08)14.2 (9.4)
ROI 3. Lacrymal gland1.5 (0.68)1.13 (0.28)0.2 (0.08)15.4 (7.8)
ROI 4. Medial extra-ocular muscle1.74 (0.57)1.27 (0.21)0.21 (0.08)11.6 (5.1)
ROI 5. Lateral extra-ocular muscle1.54 (0.27)1.17 (0.21)0.2 (0.06)9.8 (4.9)
ROI 6. Temporal muscle1.63 (0.29)1.16 (0.13)0.23 (0.07)14.6 (5.6)
ROI 7. Temporal lobe0.89 (0.07)0.75 (0.04)0.13 (0.05)20.2 (6.7)

ADC apparent diffusion coefficient, D ‘true’ diffusion coefficient, D* pseudodiffusion coefficient, f perfusion fraction, SD standard deviation, ROI region of interest

Mean intravoxel incoherent motion (IVIM) parameter values (of all measurements) ADC apparent diffusion coefficient, D ‘true’ diffusion coefficient, D* pseudodiffusion coefficient, f perfusion fraction, SD standard deviation, ROI region of interest

Test-retest repeatability

Repeatability of IVIM parameters in the orbital mass was good for ADC and D (mean CV 12% and 14%, ICC 95% and 93%, BA-LA : [−0.35; 0.35] and [−0.36; 0.27]) and poor for f and D*(mean CV 43% and 110%, ICC 90% and 65%, BA-LA : [−0.10; 0.10] and [−23.1; 20.9]) (Fig. 2.a). There was a significant correlation between ADC and D values in the orbital mass (p<0.001).
Fig. 2

Bland-Altman plots showing test-retest repeatability (a) and inter-observer repeatability agreement (b) of apparent diffusion coefficient (ADC), ‘true’ diffusion coefficient (D), pseudodiffusion coefficient (D*), perfusion fraction (f), for all regions of interest (ROIs)

Bland-Altman plots showing test-retest repeatability (a) and inter-observer repeatability agreement (b) of apparent diffusion coefficient (ADC), ‘true’ diffusion coefficient (D), pseudodiffusion coefficient (D*), perfusion fraction (f), for all regions of interest (ROIs) Regarding the structures which could be used as references, repeatability was good in the lacrymal gland for ADC and D (mean CV 18% and 19%, ICC 94% and 86%, BA-LA : [−0.61; 0.75] and [−0.36; 0.45]), poor for f and D*(mean CV 51% and 130%, ICC 91% and 75%, BA-LA : [−0.09; 0.10] and [−17.6; 12.0]). Repeatability was acceptable in the extra-ocular muscles and the extra-orbital structures for ADC and D and poor for f and D*, with a slightly better repeatability for the lateral extra-ocular muscle (Supplementary Fig. 1). All detailed values are shown in Table 3.
Table 3

Test-retest intraclass correlation coefficients (ICCs), Bland-Altman limits of agreements (BA-Las), Bias and coefficients of variation (CVs) of intravoxel incoherent motion (IVIM)–diffusion-weighted imaging (DWI) values for each anatomical region

ROI 1ROI 2ROI 3ROI 4ROI 5ROI 6ROI 7Mean of all ROI
ICC (mean)[95% CI]ADC0.96[0.91–0.98]0.95[0.87–0.98]0.94[0.85–0.97]0.54[0–0.81]0.75[0.39–0.90]0.89[0.70–0.96]0.77[0.45–0.91]0.87[0.82–0.90]
D0.94[0.86–0.98]0.93[0.83–0.97]0.86[0.67–0.94]0.72[0.35–0.88]0.71[0.28–0.88]0.94[0.85–0.97]0.59[0; 0.83]0.90[0.87–0.93]
PF0.91[0.78–0.96]0.90[0.77–0.96]0.91[0.79–0.96]0.75[0.40–0.90]0.76[0.41–0.90]0.86[0.64–0.94]0.90[0.77–0.96]0.87[0.83–0.91]
D*0.84[0.62–0.93]0.65[0.15–0.86]0.75[0.41–0.90]0.35[0; 0.73]0.32[0–0.72]0.76[0.43–0.90]0.78[0.46–0.91]0.73[0.63–0.80]
BA-LAADC[-0.29–0.26][-0.35–0.35][-0.61–0.75][-1.48–1.60][-0.52–0.55][-0.44–0.26][-0.13–0.13][-0.69–0.70]
D[-0.28–0.25][-0.36–0.27][-0.36–0.45][-0.48–0.35][-0.45–0.45][-0.12–0.13][-0.10–0.11][-0.33–0.31]
PF[-0.08–0.08][-0.10–0.10][-0.09–0.10][-0.15–0.17][-0.12–0.13][-0.12–0.08][-0.06–0.07][-15.9–14.7]
D*[-11.8–13.2][-23.1–20.9][-17.6–12.0][-14.7–17.7][-15.6–16.3][-12.3–8.0][-13.2–11.9][-0.10–0.11]
Bias (mean)[95% CI]ADC-0.01[-0.08–0.05]0.00[-0.08–0.08]0.07[-0.09–0.22]0.06[-0.28–0.41]0.01[-0.11–0.14]-0.09[-0.17–-0.01]0[-0.03–0.03]0.01[-0.05–0.06]
D-0.02[-0.08–0.04]-0.04[-0.11–0.03]0.05[-0.04–0.14]-0.06[-0.16–0.03]0[-0.10–0.10]0[-0.03–0.03]0[-0.02–0.03]-0.01[-0.04–0.02]
PF0[-0.02–0.02]0.00[-0.02–0.03]0.00[-0.02–0.03]0.01[-0.02–0.05]0.01[-0.02–0.03]-0.02[-0.04–0.00]0.01[-0.01–0.02]0[-0.01–0.01]
D*0.71[-2.13–3.54]-1.1[-6.08–3.89]-2.78[-6.14–0.57]1.50[-2.17–5.16]0.35[-3.26–3.96]-2.15[-4.45–0.15]-0.64[-3.48–2.19]-0.59[-1.83–0.65]
CV (mean) (SD)ADC0.21 (0.06)0.12 (0.05)0.18 (0.06)0.22 (0.06)0.22 (0.07)0.33 (0.13)0.23 (0.06)0.22 (0.09)
D0.23 (0.06)0.14 (0.05)0.19 (0.06)0.25 (0.07)0.25 (0.06)0.29 (0.06)0.22 (0.05)0.22 (0.07)
PF0.62 (0.18)0.43 (0.17)0.51 (0.15)0.51 (0.13)0.54 (0.1)0.66 (0.11)0.75 (0.14)0.57 (0.17)
D*1.5 (0.4)1.1 (0.4)1.3 (0.4)1.3 (0.4)1.3 (0.4)1.6 (0.2)1.2 (0.2)1.3 (0.4)

CI confidence interval, ADC apparent diffusion coefficient, D ‘true’ diffusion coefficient, D* pseudodiffusion coefficient, f perfusion fraction, ROI region of interest

Test-retest intraclass correlation coefficients (ICCs), Bland-Altman limits of agreements (BA-Las), Bias and coefficients of variation (CVs) of intravoxel incoherent motion (IVIM)–diffusion-weighted imaging (DWI) values for each anatomical region CI confidence interval, ADC apparent diffusion coefficient, D ‘true’ diffusion coefficient, D* pseudodiffusion coefficient, f perfusion fraction, ROI region of interest

Inter-observer repeatability

Inter-observer repeatability was almost perfect in the orbital mass (ICC = 95%, 93%, 94% and 90% and BA-LA = [−0.30; 0.29], [−0.26 0.27], [−0.06; 0.07] and [−10.5; 8.6] for ADC, D, f and D*, respectively) (Fig. 2.b). It was almost perfect in the lacrymal gland for all the IVIM parameters (ICC = 80%, 84%, 90% and 84% and BA-LA = [−0.56; 0.44], [−32; 0.23], [−0.06; 0.10] and [−12.3; 12.1] for ADC, D, f and D*, respectively). It was almost perfect in the extra-ocular muscles and the temporal muscle for ADC, almost perfect in the lateral extra-ocular muscle and the temporal muscle for D and in the temporal muscle for f. This level of inter-observer repeatability was only seen for the extra-ocular muscles and the temporal muscle for D* (Supplementary Fig. 2). All detailed values are shown in Table 4.
Table 4

Interobserver repeatability agreement intraclass correlation coefficients (ICCs), Bland-Altman limits of agreements (BA-Las), Bias and coefficients of variation (CVs) of intravoxel incoherent motion (IVIM)–diffusion-weighted imaging (DWI) values for each anatomical region

ROI 1ROI 2ROI 3ROI 4ROI 5ROI 6ROI 7Mean of all ROI
ICC (mean)[95% CI]ADC0.95[0.86–0.98]0.88[0.70–0.95]0.80[0.52–0.92]0.97[0.91–0.99]0.88[0.70–0.95]0.88[0.62–0.96]0.57[0–0.82]0.95[0.93–0.96]
D0.93[0.83–0.97]0.82[0.56–0.93]0.84[0.60–0.93]0.43[0–0.77]0.88[0.70–0.95]0.84[0.39–0.94]0.63[0.12–0.85]0.88[0.83–0.91]
f 0.94[0.85–0.98]0.86[0.65–0.94]0.90[0.71–0.96]0.78[0.46–0.91]0.66[0.18–0.86]0.83[0–0.95]0.81[0.54–0.92]0.85[0.78–0.89]
D*0.90[0.77–0.96]0.71[0.31–0.88]0.84[0.60–0.93]0.66[0.18–0.86]0.61[0.01–0.84]0.62[0–0.87]0.66[0.18–0.86]0.76[0.67–0.83]
BA-LAADC[-0.30–0.29][-0.48–0.38][-0.56–0.44][-0.45–0.27][-0.39–0.39][-0.25–0.48][-0.15–0.21][-0.39–0.38]
D[-0.26–0.27][-0.44–0.39][-0.32–0.23][-0.61–0.48][-0.30–0.28][-0.23–0.10][-0.07–0.11][-0.35–0.30]
f [-0.06–0.07][-0.12–0.12][-0.06–0.10][-0.14–0.17][-0.12–0.17][-0.03–0.12][-0.08–0.09][-0.09–0.13]
D*[-10.5–8.6][-22.9–16.5][-12.3–12.1][-8.8–13.3][-7.2–13.8][-5.8–16.7][-17.2–13.2][-13.4–14.8]
Bias (mean)[95% CI]ADC0[-0.07–0.07]-0.05[-0.15–0.05]-0.06[-0.18–0.06]-0.09[-0.18–-0.01]0[-0.09–0.09]0.11[0.03–0.20]0.03[-0.01–0.07]-0.01[-0.04–0.02]
D0.01[-0.05–0.07]-0.02[-0.12–0.07]-0.05[-0.11–0.02]-0.06[-0.19–0.06]-0.01[-0.08–0.06]-0.06[-0.1–-0.03]0.02[0.00–0.04]-0.03[-0.05–0.00]
f 0[-0.01–0.02]0[-0.03–0.03]0.02[0–0.04]0.02[-0.02–0.05]0.02[-0.01–0.05]0.05[0.03–0.06]0.01[-0.01–0.03]0.02[0.01–0.03]
D*-0.98[-3.20–1.24]-3.20[-7.78–1.38]-0.10[-2.94–2.74]2.28[-0.29–4.84]3.31[0.87–5.76]5.43[2.82–8.04]-1.96[-5.49–1.56]0.68[-0.49–1.85]

CI confidence interval, ADC apparent diffusion coefficient, D ‘true’ diffusion coefficient, D* pseudodiffusion coefficient, f perfusion fraction, ROI region of interest

Interobserver repeatability agreement intraclass correlation coefficients (ICCs), Bland-Altman limits of agreements (BA-Las), Bias and coefficients of variation (CVs) of intravoxel incoherent motion (IVIM)–diffusion-weighted imaging (DWI) values for each anatomical region CI confidence interval, ADC apparent diffusion coefficient, D ‘true’ diffusion coefficient, D* pseudodiffusion coefficient, f perfusion fraction, ROI region of interest

Discussion

Our study allowed evaluation of the robustness of quantitative assessment of IVIM parameters in the orbit. Test-retest repeatability of IVIM in the orbital tumours was good for ADC and D, but was poor for f and especially D*. Inter-observer repeatability agreement was almost perfect for all IVIM values in orbital tumours. Our study was compliant with the Quantitative Imaging Biomarkers Alliance (QIBA) guidelines [25-27] to best evaluate technical performance of IVIM in orbital imaging. IVIM is a relatively new imaging technique and seems promising in the characterisation and evaluation of tumours, but proportionally few studies have evaluated its accuracy, repeatability or reproducibility. Therefore, we decided to perform a study to quantify the variability and possible source of error related to the technique in the orbit before further evaluation of its clinical relevance or impact in clinical practice [26, 27]. DWI and ADC repeatability have already been evaluated over time in head and neck or brain imaging [21, 28], but only few studies have evaluated the IVIM repeatability in these domains [9, 10] and none in the orbit to the best of our knowledge. A large majority of previously published studies evaluating the IVIM accuracy, repeatability or reproducibility were performed in the abdomen, especially in the liver. They showed a higher repeatability for ADC and D parameters than for perfusion-related f and D* parameters. The pseudo-diffusion coefficient D* appeared to be the least reproducible parameter among IVIM metrics, with the coefficient of variation ranging from 24.8% to 193.8% [11–14, 29], whereas the true diffusion coefficient D appeared to be the most reproducible parameter in the liver and in the kidney [14, 30, 31]. Although a recent multicentre study showed significant discrepancies of IVIM values between different MR imagers across multiple centres [15], studies evaluating the use of this technique in the abdomen concluded that IVIM parameters could be clinically relevant [14, 30, 31]. Our results are in agreement with these studies, with an acceptable test-retest repeatability for D in both orbital and extra-orbital structures (mean 22%, range 14–29%) similar to ADC (mean 22%, range 12–33%), and a poor repeatability for f and D* (mean 57%, range 43–75% and mean 130%, range 110–160%) [14, 30, 31]. A few hypotheses may explain these results. First, the IVIM model results from an oversimplification as it does not consider the exchanges between intra and extra-vascular compartments nor the possibility of a diffusion anisotropy and it does not distinguish arterial from venous flows, which may impair the accurate estimation of the two parameters related to blood flow, f and D* [2, 32]. Second, blood water comprises less than 5% of the total tissue water in the orbit as in many organs, therefore most of the IVIM signal arises from extravascular water and hence is theoretically uninformative in assessing blood flow by the IVIM method. The dynamic range for the measurement of f and D* is poor and the quality of parameter estimates depends upon a number of factors including the number of b values obtained, especially of low b values, and the signal-to-noise ratio, which has to be high enough, therefore requiring a high-field MRI [32]. Third, there is no direct measurement of the IVIM parameters but different methods to calculate them from exponential decay data, such as the nonlinear least squares or the Bayesian probability theory. Although reported to be more accurate, the Bayesian approach provides only a representation of the uncertainty in the parameters estimates in the form of a probability density function [33]. ADC and D fulfil the QIBA quality criteria [34]. In the literature, a 20% within-subject coefficient of variation is considered as a good threshold to use a biomarker in a clinical practice, because it suggests that a change of approximately 40% is required in a single subject to be considered that this difference may be related to a biological mechanism and not to the intrinsic variability of the technique [26]. Some technical challenges are common to abdominal and orbital imaging and affect the quality of IVIM, such as the presence of motion and susceptibility artefacts [13, 15, 30, 35]. The use of a high-field MRI is preferable in orbital and ocular imaging because of its higher signal-to-noise ratio and contrast-to-noise ratio as well as its better resolution, fully adapted for a small structure like the eye [35, 36]. However, large soft-tissue and air interfaces resulting in susceptibility effects are more pronounced at 3T and can provoke image distortions, leading to a loss of image quality [35]. In order to minimize motion artefacts in the orbit, we asked the patients to keep their eyes open during the exam and to stare at a fixed point. The use of a cued-blinking protocol including a regular break every 3 s, in which acquisition is automatically paused and the subject is instructed to blink, might be another strategy to reduce motion artefacts [37]. We also used parallel imaging techniques to decrease these artefacts, and bipolar diffusion gradients to attenuate eddy currents intrinsic to the process. Good visual quality of the sequence was determined in all of our patients by the two readers, and the estimated SNR at b= 2,000 s/mm2 was good, almost twice the minimal SNR recommended for reliable IVIM imaging [10]. No patient had to be secondarily excluded from the study because of artefacts masking or distorting orbital tumours. Regarding the optimal type of ROI for analysing an orbital mass, the test-retest repeatability and the inter-observer agreements were slightly better with the use of a large freehand ROI encompassing the largest area inside the lesion versus a small circular ROI, especially for the D* parameter. The careful delineation of the entire lesion appeared to be more robust for analysis. These results are in agreement with other studies which evaluated the influence of the type, size, and position of tumour ROI on perfusion values, with a greater reliability and repeatability of large ROIs outlining entire tumours [38]. We investigated which intra-orbital or extra-orbital structures might be a good structure of reference for the quantitative use of IVIM values when diagnosing orbital diseases, if normalization of parameters between subjects was needed, such as it has been proposed in gynaecological tumours (using the outer myometrium) [39] or brain tumours (using a contralateral healthy area) [40]. The repeatability of the ADC and D parameters in the temporal muscle was excellent, with a low CV and a high ICC. This muscle is bigger than extra-ocular muscles or the lacrimal gland and is therefore easier to see. It is an immobile structure and it is less prone to image distortion compared to orbital structures. Moreover, it is almost always preserved, even in case of orbital diseases involving both eyes, thus being more reliable as a structure of reference. In our study, D values were consistently lower than the ADC values despite a clear linear relation between these two values, strongly supporting the existence of both a diffusional component in the orbital tumours evaluated by the D value and a non-negligible perfusional component evaluated by f and D*. However, perfusion-related parameter changes will require caution in their interpretation in future studies considering their poor repeatabilities. On the other hand, ADC and D showed a good repeatability in our study, suggesting that these parameters might be more reliable and reproducible quantitative biomarkers to characterise orbital tumours, quantify the severity of the disease and predict its course, but also to evaluate therapeutic responses under treatments like chemotherapy for lymphomas or prototherapy for ocular melanomas. Our study had some limitations. First, we included a relatively small number of patients recruited from a single centre, thus preventing us from performing analysis of the reproducibility of the IVIM technique on different MR devices or centres, which would be an important step before further large multicentre studies. Second, the time delay between the first and second IVIM acquisition was short, limiting the ability to evaluate all potential factors of variability and minimising the effects of other variance components. Third, we considered the orbit contralateral to the side of pathology as healthy, although most of the cases imaged were inflammatory diseases or lymphomas, which can affect both sides. In conclusion, our study showed the feasibility and robustness of quantitative assessment of IVIM parameters in the orbit. Test-retest repeatability in orbital tumours was good for ADC and D, but was poor for f and especially for D*, whereas the inter-observer agreement repeatability was almost perfect for all values. Further research would be needed to improve MR acquisition and to increase the repeatability of perfusion-related IVIM parameters, and clinical studies must be performed to assess its potential utility in clinical practice. Below is the link to the electronic supplementary material. (DOCX 19997 kb)
  39 in total

1.  Bring on the biomarkers.

Authors:  George Poste
Journal:  Nature       Date:  2011-01-13       Impact factor: 49.962

2.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

3.  Intravoxel incoherent motion diffusion-weighted MR imaging of the liver: effect of triggering methods on regional variability and measurement repeatability of quantitative parameters.

Authors:  Yedaun Lee; Seung Soo Lee; Namkug Kim; Eunki Kim; Yeong Jae Kim; Sung-Cheol Yun; Bernd Kühn; In Seong Kim; Seong Ho Park; So Yeon Kim; Moon-Gyu Lee
Journal:  Radiology       Date:  2014-09-17       Impact factor: 11.105

4.  Diagnosis of cirrhosis with intravoxel incoherent motion diffusion MRI and dynamic contrast-enhanced MRI alone and in combination: preliminary experience.

Authors:  Jignesh Patel; Eric E Sigmund; Henry Rusinek; Marcel Oei; James S Babb; Bachir Taouli
Journal:  J Magn Reson Imaging       Date:  2010-03       Impact factor: 4.813

5.  Primary central nervous system lymphoma and atypical glioblastoma: multiparametric differentiation by using diffusion-, perfusion-, and susceptibility-weighted MR imaging.

Authors:  Philipp Kickingereder; Benedikt Wiestler; Felix Sahm; Sabine Heiland; Matthias Roethke; Heinz-Peter Schlemmer; Wolfgang Wick; Martin Bendszus; Alexander Radbruch
Journal:  Radiology       Date:  2014-05-03       Impact factor: 11.105

6.  Differentiation of pancreas carcinoma from healthy pancreatic tissue using multiple b-values: comparison of apparent diffusion coefficient and intravoxel incoherent motion derived parameters.

Authors:  Andreas Lemke; Frederik B Laun; Miriam Klauss; Thomas J Re; Dirk Simon; Stefan Delorme; Lothar R Schad; Bram Stieltjes
Journal:  Invest Radiol       Date:  2009-12       Impact factor: 6.016

7.  Orbital lymphoproliferative disorders (OLPDs): value of MR imaging for differentiating orbital lymphoma from benign OPLDs.

Authors:  K Haradome; H Haradome; Y Usui; S Ueda; T C Kwee; K Saito; K Tokuuye; J Matsubayashi; T Nagao; H Goto
Journal:  AJNR Am J Neuroradiol       Date:  2014-05-29       Impact factor: 3.825

8.  Benign and malignant orbital lymphoproliferative disorders: Differentiating using multiparametric MRI at 3.0T.

Authors:  Xiao-Quan Xu; Hao Hu; Hu Liu; Jiang-Fen Wu; Peng Cao; Hai-Bin Shi; Fei-Yun Wu
Journal:  J Magn Reson Imaging       Date:  2016-06-14       Impact factor: 4.813

9.  Liver cirrhosis: intravoxel incoherent motion MR imaging--pilot study.

Authors:  Alain Luciani; Alexandre Vignaud; Madeleine Cavet; Jeanne Tran Van Nhieu; Ariane Mallat; Lucile Ruel; Alexis Laurent; Jean-François Deux; Pierre Brugieres; Alain Rahmouni
Journal:  Radiology       Date:  2008-12       Impact factor: 11.105

10.  Clinical evaluation of ultra-high-field MRI for three-dimensional visualisation of tumour size in uveal melanoma patients, with direct relevance to treatment planning.

Authors:  Jan-Willem M Beenakker; Teresa A Ferreira; Karina P Soemarwoto; Stijn W Genders; Wouter M Teeuwisse; Andrew G Webb; Gregorius P M Luyten
Journal:  MAGMA       Date:  2016-02-25       Impact factor: 2.310

View more
  8 in total

1.  A signature of structural MRI features at 3 Tesla allows an accurate characterization of orbital cavernous venous malformation.

Authors:  Simon Elbaze; Loïc Duron; Natasha Mambour; Mathieu Zmuda; Sidney Krystal; Jessica Guillaume; Julien Savatovsky; Augustin Lecler
Journal:  Eur Radiol       Date:  2022-10-20       Impact factor: 7.034

2.  Intravoxel incoherent motion (IVIM) 3 T MRI for orbital lesion characterization.

Authors:  Augustin Lecler; Loïc Duron; Mathieu Zmuda; Kevin Zuber; Olivier Bergès; Marc Putterman; Julien Savatovsky; Laure Fournier
Journal:  Eur Radiol       Date:  2020-08-01       Impact factor: 5.315

3.  Intravoxel incoherent motion perfusion in patients with Moyamoya disease: comparison with 15O-gas positron emission tomography.

Authors:  Shoko Hara; Masaaki Hori; Ryo Ueda; Akifumi Hagiwara; Shihori Hayashi; Motoki Inaji; Yoji Tanaka; Taketoshi Maehara; Kenji Ishii; Shigeki Aoki; Tadashi Nariai
Journal:  Acta Radiol Open       Date:  2019-05-23

4.  Differentiation between orbital malignant and benign tumors using intravoxel incoherent motion diffusion-weighted imaging: Correlation with dynamic contrast-enhanced magnetic resonance imaging.

Authors:  Xiao-Quan Xu; Hao Hu; Guo-Yi Su; Hu Liu; Fei-Yun Wu; Hai-Bin Shi
Journal:  Medicine (Baltimore)       Date:  2019-03       Impact factor: 1.889

5.  Reference Interval for the Axis-Shield Clinical Chemistry Heparin-Binding Protein Assay.

Authors:  Sumi Yoon; Mina Hur; Hanah Kim; Hee-Won Moon; Yeo-Min Yun
Journal:  Diagnostics (Basel)       Date:  2022-08-10

Review 6.  Perfusion-driven Intravoxel Incoherent Motion (IVIM) MRI in Oncology: Applications, Challenges, and Future Trends.

Authors:  Mami Iima
Journal:  Magn Reson Med Sci       Date:  2020-06-15       Impact factor: 2.471

7.  Intravoxel incoherent motion and ADC measurements for differentiating benign from malignant thyroid nodules: utilizing the most repeatable region of interest delineation at 3.0 T.

Authors:  Minghui Song; Yunlong Yue; Yanfang Jin; Jinsong Guo; Lili Zuo; Hong Peng; Queenie Chan
Journal:  Cancer Imaging       Date:  2020-01-22       Impact factor: 3.909

8.  How Reproducible Is the Data from Sysmex DI-60 in Leukopenic Samples?

Authors:  Sumi Yoon; Mina Hur; Gun Hyuk Lee; Minjeong Nam; Hanah Kim
Journal:  Diagnostics (Basel)       Date:  2021-11-23
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.