Xavier Monnet1,2,3, Jean-Louis Teboul4,5.
Abstract
BACKGROUND: For complex patients in the intensive care unit or in the operating room, many questions regarding their haemodynamic management cannot be answered with simple clinical examination. In particular, arterial pressure allows only a rough estimation of cardiac output. Transpulmonary thermodilution is a technique that provides a full haemodynamic assessment through cardiac output and other indices. MAIN BODY: Through the analysis of the thermodilution curve recorded at the tip of an arterial catheter after the injection of a cold bolus in the venous circulation, transpulmonary thermodilution intermittently measures cardiac output. This measure allows the calibration of pulse contour analysis. This provides continuous and real time monitoring of cardiac output, which is not possible with the pulmonary artery catheter. Transpulmonary thermodilution provides several variables beyond cardiac output. It estimates the end-diastolic volume of the four cardiac cavities, which is a marker of cardiac preload. It provides an estimation of the systolic function of the combined ventricles. It is more direct than the pulmonary artery catheter, but does not allow the distinct estimation of right and left cardiac function. It is easier and faster to perform than echocardiography, but does not provide a full evaluation of the cardiac structure and function. Transpulmonary thermodilution has the unique advantage of being able to estimate at the bedside extravascular lung water, which quantifies the volume of pulmonary oedema, and pulmonary vascular permeability, which quantifies the degree of a pulmonary capillary leak. Both indices are helpful for guiding fluid strategy, especially in case of acute respiratory distress syndrome.Entities:
Keywords: Cardiac output; Cardiac preload; Extravascular lung water; Fluid responsiveness; Haemodynamic monitoring
Mesh:
Year: 2017 PMID: 28625165 PMCID: PMC5474867 DOI: 10.1186/s13054-017-1739-5
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Fig. 1Haemodynamic variables provided by transpulmonary thermodilution and calibrated pulse contour analysis, with their meaning, utility and normal values (N). ARDS acute respiratory distress syndrome
Advantages and drawbacks of measurements performed by transpulmonary thermodilution and calibrated pulse contour analysis
| Variable | Main advantages | Main drawbacks |
|---|---|---|
| Cardiac output measured by TPTD | As reliable as pulmonary thermodilution | Does not provide a continuous measurement |
| Cardiac output measured by pulse contour analysis | Continuous measurement | Requires regular recalibration |
| Global end-diastolic volume | Better reflects cardiac preload than pressure markers of preload | Does not distinguish between the right and left ventricles |
| Stroke volume variation | Continuous automated assessment of fluid responsiveness | Cannot be used in case of spontaneous breathing, cardiac arrhythmias and ARDS |
| Cardiac function index, global ejection fraction | Can be used as an alarm for decreased LV systolic function | Overestimate LV systolic function in case of right ventricular dilation |
| Extravascular lung water | Directly estimates the volume of lung oedema | Unreliable in case of pulmonary embolism, lung resection, large pleural effusions |
| Pulmonary vascular permeability index | Directly estimates lung permeability | Same as for extravascular lung water |
ARDS acute respiratory distress syndrome, LV left ventricular, PAOP pulmonary artery occlusion pressure, TPTD transpulmonary thermodilution
Fig. 2Assessment of intrathoracic volume by the PiCCO device. With the VolumeView device, the geometrical analysis of the thermodilution curve directly estimates the global end-diastolic volume. Dt downslope time, MTt mean transit time. For explanation, see the text
Fig. 3Indication for haemodynamic monitoring in the intensive care unit. A-line arterial line, CV-line central venous line. The choice also depends on team experience