Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients.

K Bendjelid1, G Marx, N Kiefer, T P Simon, M Geisen, A Hoeft, N Siegenthaler, C K Hofer.   

Abstract

BACKGROUND: A new calibrated pulse wave analysis method (VolumeView™/EV1000™, Edwards Lifesciences, Irvine, CA, USA) has been developed to continuously monitor cardiac output (CO). The aim of this study was to compare the performance of the VolumeView method, and of the PiCCO2™ pulse contour method (Pulsion Medical Systems, Munich, Germany), with reference transpulmonary thermodilution (TPTD) CO measurements.
METHODS: This was a prospective, multicentre observational study performed in the surgical and interdisciplinary intensive care units of four tertiary hospitals. Seventy-two critically ill patients were monitored with a central venous catheter, and a thermistor-tipped femoral arterial VolumeView™ catheter connected to the EV1000™ monitor. After initial calibration by TPTD CO was continuously assessed using the VolumeView-CCO software (CCO(VolumeView)) during a 72 h period. TPTD was performed in order to obtain reference CO values (COREF). TPTD and arterial wave signals were transmitted to a PiCCO2™ monitor in order to obtain CCO(PiCCO) values. CCO(VolumeView) and CCO(PiCCO) were recorded over a 5 min interval before assessment of CO(TPTD). Bland-Altman analysis, %(errors), and concordance (trend analysis) were calculated.
RESULTS: A total of 338 matched sets of data were available for comparison. Bias for CCO(VolumeView)-CO(REF) was -0.07 litre min(-1) and for CCO(PiCCO)-CO(REF) +0.03 litre min(-1). Corresponding limits of agreement were 2.00 and 2.48 litre min(-1) (P<0.01), %(errors) 29 and 37%, respectively. Trending capabilities were comparable for both techniques.
CONCLUSIONS: The performance of the new VolumeView™-CCO method is as reliable as the PiCCO2™-CCO pulse wave analysis in critically ill patients. However, an improved precision was observed with the VolumeView™ technique. CLINICALTRIALS.GOV IDENTIFIER: NCT01405040.

Entities:  

Keywords:  continuous cardiac output; pulse wave analysis; transpulmonary thermodilution

Mesh:

Year:  2013        PMID: 23625132     DOI: 10.1093/bja/aet116

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  17 in total

1.  Comparison of cardiac output measures by transpulmonary thermodilution, pulse contour analysis, and pulmonary artery thermodilution during off-pump coronary artery bypass surgery: a subgroup analysis of the cardiovascular anaesthesia registry at a single tertiary centre.

Authors:  Youn Joung Cho; Chang-Hoon Koo; Tae Kyong Kim; Deok Man Hong; Yunseok Jeon
Journal:  J Clin Monit Comput       Date:  2015-10-01       Impact factor: 2.502

Review 2.  Reproducibility of transpulmonary thermodilution cardiac output measurements in clinical practice: a systematic review.

Authors:  Raphaël Giraud; Nils Siegenthaler; Paolo Merlani; Karim Bendjelid
Journal:  J Clin Monit Comput       Date:  2016-01-11       Impact factor: 2.502

3.  The Diagnosis and Hemodynamic Monitoring of Circulatory Shock: Current and Future Trends.

Authors:  Adham Hendy; Şerban-Ion Bubenek-Turconi
Journal:  J Crit Care Med (Targu Mures)       Date:  2016-08-10

Review 4.  Journal of Clinical Monitoring and Computing 2016 end of year summary: cardiovascular and hemodynamic monitoring.

Authors:  Bernd Saugel; Karim Bendjelid; Lester A Critchley; Steffen Rex; Thomas W L Scheeren
Journal:  J Clin Monit Comput       Date:  2017-01-07       Impact factor: 2.502

Review 5.  Newer methods of cardiac output monitoring.

Authors:  Yatin Mehta; Dheeraj Arora
Journal:  World J Cardiol       Date:  2014-09-26

6.  Femoral indicator injection for transpulmonary thermodilution using the EV1000/VolumeView(®): do the same criteria apply as for the PiCCO(®)?

Authors:  Wolfgang Huber; Veit Phillip; Josef Höllthaler; Caroline Schultheiss; Bernd Saugel; Roland M Schmid
Journal:  J Zhejiang Univ Sci B       Date:  2016-07       Impact factor: 3.066

Review 7.  Minimally invasive or noninvasive cardiac output measurement: an update.

Authors:  Lisa Sangkum; Geoffrey L Liu; Ling Yu; Hong Yan; Alan D Kaye; Henry Liu
Journal:  J Anesth       Date:  2016-03-09       Impact factor: 2.078

Review 8.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

9.  Assessment of method agreement between two minimally invasive hemodynamic measurements in septic shock patients on high doses of vasopressor drugs. A preliminary study.

Authors:  Oana Antal; Mihai Mărginean; Natalia Hagău
Journal:  Rom J Anaesth Intensive Care       Date:  2017-10

10.  Use of a Minimally Invasive Cardiac Output Monitor to Optimise Haemodynamics in a Patient with Mitral Valve Disease Undergoing Cerebrovascular Surgery.

Authors:  Ali M Al-Mashani; Niranjan D Waje; Neeraj Salhotra; Samaresh Das; Neelam Suri; Rashid A Al-Sheheimi; Nilay Chatterjee
Journal:  Sultan Qaboos Univ Med J       Date:  2017-10-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.