Juan Huang1, Benjamin Sheffield2,3, Payton Lin4, Fan-Gang Zeng5.
Abstract
For cochlear implant users, combined electro-acoustic stimulation (EAS) significantly improves the performance. However, there are many more users who do not have any functional residual acoustic hearing at low frequencies. Because tactile sensation also operates in the same low frequencies (<500 Hz) as the acoustic hearing in EAS, we propose electro-tactile stimulation (ETS) to improve cochlear implant performance. In ten cochlear implant users, a tactile aid was applied to the index finger that converted voice fundamental frequency into tactile vibrations. Speech recognition in noise was compared for cochlear implants alone and for the bimodal ETS condition. On average, ETS improved speech reception thresholds by 2.2 dB over cochlear implants alone. Nine of the ten subjects showed a positive ETS effect ranging from 0.3 to 7.0 dB, which was similar to the amount of the previously-reported EAS benefit. The comparable results indicate similar neural mechanisms that underlie both the ETS and EAS effects. The positive results suggest that the complementary auditory and tactile modes also be used to enhance performance for normal hearing listeners and automatic speech recognition for machines.Entities:
Mesh:
Year: 2017 PMID: 28526871 PMCID: PMC5438362 DOI: 10.1038/s41598-017-02429-1
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1(A) Average speech reception threshold (SRT) between the cochlear implant only condition (CI: open bar) and the combined electro-tactile stimulation (ETS: filled bar). The error bars represent one standard deviation of the mean. The asterisk indicates a significant improvement of the ETS condition over the CI condition. (B) Individual enhancement in terms of the SRT difference between the CI and ETS conditions, ranked from low (Subject 1) to high (Subject 10).
Figure 2The ETS experimental setup. A personal computer (PC) controls both electric and tactile stimulation. Electric stimulation is delivered to a cochlear implant (CI) through an audiometer and speaker. Tactile stimulation is delivered to the index finger through an amplifier and a tactile transducer.