| Literature DB >> 28464008 |
Jun Kim1, Jongmin Hwang1, Jin Hee Choi1, Hyo-In Choi1, Min-Seok Kim1, Sung-Ho Jung2, Gi-Byoung Nam1, Kee-Joon Choi1, Jae Won Lee2, You-Ho Kim1, Jae-Joong Kim1.
Abstract
End-stage heart failure patients with implantable cardioverter-defibrillator (ICD) with/without cardiac resynchronization therapy (CRT-D) often require heart transplantation (HTPL) as a last-resort treatment. We aimed to assess the frequency and clinical impact of retained ICD lead materials in HTPL patients. In this retrospective single center study, we examined the clinical records and chest radiographs of patients with ICD and CRT-D who underwent HTPL between January 1992 and July 2014. Of 40 patients with ICD and CRT-D at HTPL, 19 (47.5%) patients had retained ICD lead materials within the central venous system. Retained ICD lead materials following HTPL were more frequently noted in patients with longer implantation durations until HTPL. None of the patients underwent extraction procedures after HTPL. All patients were asymptomatic and did not exhibit significant complications or death related to the retained ICD lead materials. Seven (7/40, 17.5%) patients without any retained ICD lead materials underwent magnetic resonance imaging (MRI) during the follow-up period (median, 29.5 months); none of the patients with retained lead materials were given MRI. Considering the common use of MRI in HTPL patients, further studies on the prophylactic extraction of retained ICD lead materials and safety of MRI in these patients are needed.Entities:
Mesh:
Year: 2017 PMID: 28464008 PMCID: PMC5413001 DOI: 10.1371/journal.pone.0176925
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Baseline characteristics of patients at the time of heart transplantation.
| Overall (n = 40) | With retained ICD/CRT-D lead materials (n = 19) | Without retained ICD/CRT-D lead materials (n = 21) | p value | |
|---|---|---|---|---|
| Age at heart transplantation, years | 51.6±9.6 | 52.3±9.5 | 50.9±9.9 | 0.650 |
| Male, n (%) | 30 (75%) | 15 (78.9%) | 15 (71.4%) | 0.583 |
| BMI | 22.8±4.2 | 22.6±3.7 | 22.9±4.7 | 0.830 |
| Etiology, n (%) | ||||
| Dilated CMP | 27 (67.5%) | 11 (57.9%) | 16 (76.2%) | 0.314 |
| Ischemic CMP | 7 (21.1%) | 4 (21.1%) | 3 (14.3%) | 0.574 |
| Hypertrophic CMP | 3 (7.5%) | 3 (15.8%) | 0 (0%) | 0.058 |
| Others | 3 (7.5%) | 1 (5.3%) | 2 (9.6%) | 0.335 |
| History of AF, n (%) | 14 (35.0%) | 9 (47.4%) | 5 (23.8%) | 0.186 |
| History of ventricular arrhythmia, n (%) | 31 (77.5%) | 17 (89.5%) | 14 (66.7%) | 0.085 |
| Type of device, n (%) | ||||
| ICD | 35 (87.5%) | 18 (94.7%) | 17 (81.0%) | 0.188 |
| CRT-D | 5 (12.5%) | 1 (5.3%) | 4 (19.0%) | 0.188 |
| Type of defibrillator lead, n (%) | 0.335 | |||
| Single coil | 1 (2.5%) | 0 (0%) | 1 (4.8%) | |
| Dual coil | 39 (97.5%) | 19 (100%) | 20 (95.2%) | |
| Number of leads at the time of heart transplantation, median (range) | 2 (1–4) | 2 (1–3) | 2 (1–3) | 0.415 |
| Duration from device implant to heart transplantation, median (range), months | 26 (1–154) | 31 (18–154) | 18 (1–146) | 0.010 |
Data are presented as mean ± SD, median (interquartile range), or number (%). Abbreviations: BMI, body mass index; CMP, cardiomyopathy; AF, atrial fibrillation; ICD, implantable cardioverter defibrillator; CRT-D, cardiac resynchronization therapy (CRT) with defibrillation capability.
Characteristics of the retained ICD/CRT-D lead materials following heart transplantation.
| Components of retained lead materials, n (%) | |
| Coil | 14 (73.7%) |
| Lead | 0 (0%) |
| Coil and lead | 5 (26.3%) |
| Extent of the retained lead materials, n (%), | |
| SVC | 7 (36.8%) |
| From SVC to justify or Right brachiocephalic vein | 7 (36.8%) |
| From SVC to justify or Right subclavian vein | 5 (26.3%) |
Data are presented as number (%). Abbreviations: ICD, implantable cardioverter defibrillator; CRT-D, cardiac resynchronization therapy (CRT) with defibrillation capability; SVC, superior vena cava.
MRI examination in HTPL patients.
| Patients without ICD/CRT-D at HTPL (n = 434) | Patients without remnant ICD/CRT-D lead materials after HTPL (n = 21) | Patients with remnant ICD/CRT-D lead materials after HTPL (n = 19) | |
|---|---|---|---|
| Number of patients who underwent MRI | 168 (38.7%) | 7 (17.5%) | 0 |
| Total number of MRI procedures | 294 | 14 | 0 |
| Site of MRI | 0 | ||
| Brain | 211 | 12 | 0 |
| Lumbar spine | 25 | 1 | 0 |
| Lower extremities | 17 | 0 | 0 |
| Hip | 15 | 0 | 0 |
| Abdomen | 12 | 1 | 0 |
| Cervical spine | 9 | 0 | 0 |
| Upper extremities | 4 | 0 | 0 |
| Heart | 1 | 0 | 0 |
Data are presented as number or number (%). Abbreviations: MRI, magnetic resonance imaging examination; HTPL, heart transplantation; ICD, implantable cardioverter defibrillator; CRT-D, cardiac resynchronization therapy (CRT) with defibrillation capability.