Literature DB >> 25111324

Major predictors of fibrous adherences in transvenous implantable cardioverter-defibrillator lead extraction.

Luca Segreti1, Andrea Di Cori2, Ezio Soldati2, Giulio Zucchelli2, Stefano Viani2, Luca Paperini2, Raffaele De Lucia2, Giovanni Coluccia2, Sergio Valsecchi3, Maria Grazia Bongiorni2.   

Abstract

BACKGROUND: Percutaneous removal of implantable cardioverter-defibrillator (ICD) leads is a difficult procedure because of the consequence of massive fibrous tissue growth along the lead.
OBJECTIVE: The purpose of this study was to describe the occurrence and location of fibrous adherences in ICD lead extraction and to identify potential predictors among patient and lead characteristics.
METHODS: We studied 637 consecutive patients who underwent transvenous extraction of 678 ICD leads from 1997 to 2013.
RESULTS: Procedural success rate was 99%, without major complications. Areas of adherence were found in the subclavian vein (78%), innominate vein (65%), superior vena cava (66%), and heart (73%). Dwell-time, passive fixation, and dual-coil lead design were independently associated with adherences. Dual-coil lead design was associated with adherences in the innominate vein and superior vena cava, whereas coil treatment (eg, expandable polytetrafluoroethylene-coated or medical adhesive back-filled strategies) prevented adherences. Passive fixation mechanism was associated with adherences in the heart.
CONCLUSION: ICD leads, after long dwell-time, are affected by fibrous adherences uniformly distributed along the lead course. Lead features represent major predictors of the phenomenon. Careful lead selection is recommended at the time of implantation to prevent adherences. In addition, lead-related risk stratification is mandatory before a transvenous extraction procedure.
Copyright © 2014 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Adherence; Implantable cardioverter-defibrillator; Internal jugular vein; Lead; extraction

Mesh:

Year:  2014        PMID: 25111324     DOI: 10.1016/j.hrthm.2014.08.011

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  17 in total

Review 1.  Considerations for cardiac device lead extraction.

Authors:  Oussama Wazni; Bruce L Wilkoff
Journal:  Nat Rev Cardiol       Date:  2016-01-29       Impact factor: 32.419

Review 2.  Transvenous Lead Extractions: Current Approaches and Future Trends.

Authors:  Adryan A Perez; Frank W Woo; Darren C Tsang; Roger G Carrillo
Journal:  Arrhythm Electrophysiol Rev       Date:  2018-08

Review 3.  [ICD leads].

Authors:  Carsten W Israel; Mohamed Karim Sheta
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06

Review 4.  A Questionable Indication For ICD Extraction After Successful VT Ablation.

Authors:  Luca Segreti; Andrea Di Cori; Giulio Zucchelli; Ezio Soldati; Giovanni Coluccia; Stefano Viani; Luca Paperini; Maria Grazia Bongiorni
Journal:  J Atr Fibrillation       Date:  2015-04-30

Review 5.  State-of-the-art narrative review: multimodality imaging in electrophysiology and cardiac device therapies.

Authors:  Balint Laczay; Divyang Patel; Richard Grimm; Bo Xu
Journal:  Cardiovasc Diagn Ther       Date:  2021-06

6.  Effectiveness and Safety of Transvenous Removal of Cardiac Pacing and Implantable Cardioverter-defibrillator Leads in the Real Clinical Scenario.

Authors:  Roberto Costa; Katia Regina da Silva; Elizabeth Sartori Crevelari; Wagner Tadeu Jurevicius Nascimento; Marcia Mitie Nagumo; Martino Martinelli Filho; Fabio Biscegli Jatene
Journal:  Arq Bras Cardiol       Date:  2020-12       Impact factor: 2.000

7.  Transvenous Lead Extraction via the Inferior Approach Using a Gooseneck Snare versus Simple Manual Traction.

Authors:  Uk Jo; Jun Kim; You-Mi Hwang; Ji-Hyun Lee; Min-Su Kim; Hyung-Oh Choi; Woo-Seok Lee; Chang-Hee Kwon; Gi-Young Ko; Hyun-Ki Yoon; Gi-Byoung Nam; Kee-Joon Choi; You-Ho Kim
Journal:  Korean Circ J       Date:  2016-03-21       Impact factor: 3.243

8.  Retrieval of a transcatheter pacemaker in sheep after a mid-term implantation time.

Authors:  Maria Grazia Bongiorni; Luca Segreti; Andrea Di Cori; Matthew Bonner; Michael Eggen; Pamela Omdahl
Journal:  HeartRhythm Case Rep       Date:  2015-10-23

9.  Frequency and clinical impact of retained implantable cardioverter defibrillator lead materials in heart transplant recipients.

Authors:  Jun Kim; Jongmin Hwang; Jin Hee Choi; Hyo-In Choi; Min-Seok Kim; Sung-Ho Jung; Gi-Byoung Nam; Kee-Joon Choi; Jae Won Lee; You-Ho Kim; Jae-Joong Kim
Journal:  PLoS One       Date:  2017-05-02       Impact factor: 3.240

10.  Transvenous Extraction and Removal of Pacing Leads Placed after Cardiac Transplantation.

Authors:  Caleb Norton; Benjamin Holmes; Asad Al Aboud; Eun-Jeong Kim; Holly Gonzales; Christopher Ellis; Roy John; George H Crossley; Jay Montgomery
Journal:  Case Rep Cardiol       Date:  2019-01-03
View more