B F H Ang1, J Y Chen2, A K S Yew2, S K Chua2, S M Chou3, S L Chia2, J S B Koh2, T S Howe2. 1. Singapore General Hospital, The Academia, 20 College Road, 169856 Singapore benjamin.ang@mohh.com.sg. 2. Singapore General Hospital, The Academia, 20 College Road, 169856 Singapore. 3. Nanyang Technological University, School of Mechanical and Aerospace Engineering, 50 Nanyang Avenue, Singapore 639798.
Abstract
OBJECTIVES: External fixators are the traditional fixation method of choice for contaminated open fractures. However, patient acceptance is low due to the high profile and therefore physical burden of the constructs. An externalised locking compression plate is a low profile alternative. However, the biomechanical differences have not been assessed. The objective of this study was to evaluate the axial and torsional stiffness of the externalised titanium locking compression plate (ET-LCP), the externalised stainless steel locking compression plate (ESS-LCP) and the unilateral external fixator (UEF). METHODS: A fracture gap model was created to simulate comminuted mid-shaft tibia fractures using synthetic composite bones. Fifteen constructs were stabilised with ET-LCP, ESS-LCP or UEF (five constructs each). The constructs were loaded under both axial and torsional directions to determine construct stiffness. RESULTS: The mean axial stiffness was very similar for UEF (528 N/mm) and ESS-LCP (525 N/mm), while it was slightly lower for ET-LCP (469 N/mm). One-way analysis of variance (ANOVA) testing in all three groups demonstrated no significant difference (F(2,12) = 2.057, p = 0.171).There was a significant difference in mean torsional stiffness between the UEF (0.512 Nm/degree), the ESS-LCP (0.686 Nm/degree) and the ET-LCP (0.639 Nm/degree), as determined by one-way ANOVA (F(2,12) = 6.204, p = 0.014). A Tukey post hoc test revealed that the torsional stiffness of the ESS-LCP was statistically higher than that of the UEF by 0.174 Nm/degree (p = 0.013). No catastrophic failures were observed. CONCLUSION: Using the LCP as an external fixator may provide a viable and attractive alternative to the traditional UEF as its lower profile makes it more acceptable to patients, while not compromising on axial and torsional stiffness.Cite this article: B. F. H. Ang, J. Y. Chen, A. K. S. Yew, S. K. Chua, S. M. Chou, S. L. Chia, J. S. B. Koh, T. S. Howe. Externalised locking compression plate as an alternative to the unilateral external fixator: a biomechanical comparative study of axial and torsional stiffness. Bone Joint Res 2017;6:216-223. DOI: 10.1302/2046-3758.64.2000470.
OBJECTIVES: External fixators are the traditional fixation method of choice for contaminated open fractures. However, patient acceptance is low due to the high profile and therefore physical burden of the constructs. An externalised locking compression plate is a low profile alternative. However, the biomechanical differences have not been assessed. The objective of this study was to evaluate the axial and torsional stiffness of the externalised titanium locking compression plate (ET-LCP), the externalised stainless steel locking compression plate (ESS-LCP) and the unilateral external fixator (UEF). METHODS: A fracture gap model was created to simulate comminuted mid-shaft tibia fractures using synthetic composite bones. Fifteen constructs were stabilised with ET-LCP, ESS-LCP or UEF (five constructs each). The constructs were loaded under both axial and torsional directions to determine construct stiffness. RESULTS: The mean axial stiffness was very similar for UEF (528 N/mm) and ESS-LCP (525 N/mm), while it was slightly lower for ET-LCP (469 N/mm). One-way analysis of variance (ANOVA) testing in all three groups demonstrated no significant difference (F(2,12) = 2.057, p = 0.171).There was a significant difference in mean torsional stiffness between the UEF (0.512 Nm/degree), the ESS-LCP (0.686 Nm/degree) and the ET-LCP (0.639 Nm/degree), as determined by one-way ANOVA (F(2,12) = 6.204, p = 0.014). A Tukey post hoc test revealed that the torsional stiffness of the ESS-LCP was statistically higher than that of the UEF by 0.174 Nm/degree (p = 0.013). No catastrophic failures were observed. CONCLUSION: Using the LCP as an external fixator may provide a viable and attractive alternative to the traditional UEF as its lower profile makes it more acceptable to patients, while not compromising on axial and torsional stiffness.Cite this article: B. F. H. Ang, J. Y. Chen, A. K. S. Yew, S. K. Chua, S. M. Chou, S. L. Chia, J. S. B. Koh, T. S. Howe. Externalised locking compression plate as an alternative to the unilateral external fixator: a biomechanical comparative study of axial and torsional stiffness. Bone Joint Res 2017;6:216-223. DOI: 10.1302/2046-3758.64.2000470.
External fixators are the traditional fixation method of choice for contaminated open fractures. However, patient acceptance is low due to the high profile and therefore physical burden of the constructs.An externalised locking compression plate is a low profile alternative. However, the biomechanical differences have not been assessed.The objective of this study was to evaluate the axial and torsional stiffness of the externalised titanium locking compression plate (ET-LCP), the externalised stainless steel locking compression plate (ESS-LCP) and the unilateral external fixator (UEF).The externalised LCP has a similar axial stiffness to that of the UEF.The torsional stiffness of the ESS-LCP is higher than that of the UEF.Using the LCP as an external fixator may provide a viable and attractive alternative to the traditional UEF as its lower profile makes it more acceptable to patients, while not compromising on axial and torsional stiffness.First biomechanical study comparing the axial and torsional stiffness of externalised locking compression plate with an external fixator.Four-point bending and cyclic fatigue testing not performed due to equipment limitations.
Introduction
External fixators are the traditional fixation method of choice for contaminated open fractures and in certain closed fractures with severe soft-tissue injuries. External fixators allow for better soft-tissue management and preservation of blood supply to the fractured bone.[1,2] They can be used for temporary or definitive fixation. However, acceptance of external fixation is low due to the high profile of the constructs, which makes it obstructive and inconvenient during ambulation and dressing. The high profile of the external fixator is secondary to the length of the Schanz pins and the size of the clamps, which also limit how close to the bone the rod can be placed. Besides the obstructive nature of the high profile construct, some patients are also self-conscious of these fixators and find them less socially acceptable.[2-8]Various locking plates have been used as a low profile alternative, particularly the locking compression plate (LCP) which has the advantage of a lower profile and therefore poses less of an inconvenience during dressing and ambulation. The externalised LCP is also a less costly option compared with the external fixator.[8] The main drawback of using the LCP, compared with the external fixator, is the inability to perform subsequent adjustments of the bone-to-plate offset and distraction/compression at the fracture site.Stiffness (the measurement of force per unit displacement in Newtons/metre) of the construct is extremely important because it is related to the relative micromotion at the fracture site, which has a direct effect on the biology of fracture healing. Too much movement at the fracture site, which can be caused inadvertently by too high an offset between the bone and the plate, can result in fibrous union. Delayed union or even nonunion will eventually lead to fatigue failure of the implant.Bottlang et al[9] report that, even for the same implant, the reported stiffness values can vary greatly as they are affected by the test setup. This is particularly the case when the stiffness is calculated from the displacement of the loading actuator. Actuator displacement represents deformation along the entire test specimen and can grossly overestimate the actual motion at the fracture site.[9] Studies which demonstrate a higher stiffness used an extensometer to determine displacement at the fracture site which is more accurate.[10,11]The stiffness of the LCP used in a traditional manner has been quoted as being as low as 42 N/mm to as high as 3300 N/mm.[9,12-17] The methodologies in these studies are different, including the loading conditions, as are the measurements of displacement (either across the entire construct or at the fracture gap) used in calculating the stiffness of the construct. Other differences include the use of monocortical screws in some studies,[16] the use of different implants (length, size, type), and different screw configurations/densities.To the best of our knowledge, there has only been one study performed investigating the mechanical properties of a LCP when used as an external fixation device. However, no attempt was made to compare it with the traditional carbon rod and Schanz pin external fixator construct.[18] This direct comparison under standardised conditions gives us important data on the relative mechanical properties of the externalised LCP and external fixator constructs. Lower limb biomechanics studies reveal that forces at the fracture site are predominantly compressive during two-legged stance, while transverse and torsional contributions to load bearing are relatively low.[16] The objective of this study is to compare the axial and torsional stiffness of the externalised LCP and the unilateral external fixator (UEF).We hypothesise that the externalised LCP has a similar axial and torsional stiffness to that of the UEF, which favours its use as an external fixation device.
Materials and Methods
Fracture model
The tibia was chosen as it is the most suitable bone for using the externalised LCP as an external fixator, being superficial and having only a thin layer of soft tissue overlying it. A fracture gap model was created to simulate comminuted mid-shaft tibia fractures using synthetic composite bones.[19,20] Fourth-generation, large-sized left tibia synthetic composite bones, model #3402 (Sawbones; Pacific Research Laboratories Inc., Vashon, Washington), 405 mm in length, 84 mm wide at the proximal tibia, 58 mm wide at the distal tibia, 28 mm wide at the middle of the tibial shaft, and 10 mm in canal diameter, were used for this study.The specimens were cut at the mid point to create a 20 mm gap, simulating a comminuted diaphyseal fracture and ensuring that there was no contact between the two ends of the fracture gap during axial loading.
Implants
A power analysis was undertaken before this study was carried out, based on the results from a preliminary study. To detect a 16% reduction in axial stiffness from a baseline of 530 ± 20 N/mm at a power of 0.80, a sample size of at least four specimens in each group would be required. This calculation was made for a two-sided test with a type I error of 0.05. In view of possible sampling errors in the preliminary study, five specimens in each group were chosen to provide for these possible sampling errors.[21]Fifteen tibias were divided into three groups of five, with fixation either by externalised titanium locking compression plate (ET-LCP), the externalised stainless steel locking compression plate (ESS-LCP) or the unilateral external fixator (UEF).Five tibias were stabilised with a titanium 260 mm 14-hole broad 4.5/5 mm LCP (Synthes, Switzerland), with 5 mm diameter titanium locking screws in the third and seventh locking holes from the middle of the LCP, on both sides of the fracture. Five tibias were stabilised with a stainless steel 260 mm 14-hole broad 4.5/5 mm LCP (Synthes), with 5 mm diameter stainless steel locking screws in the third and seventh locking holes from the middle of the LCP, on both sides of the fracture. Both LCPs are 260 mm long, 5.2 mm thick, 17.5 mm wide and have a hole spacing of 18 mm.In our control model, five tibias were stabilised with external fixation (Synthes). Two stainless steel standard 5 mm diameter Schanz pins were used per fragment and attached with standard clamps to one standard 11 mm diameter stainless steel rod. The positions of the Schanz pins correspond to the locking screws in the third and seventh locking holes from the middle of the LCP. The choice of the third and seventh locking holes from the middle of the LCP adheres to the Arbeitsgemeinschaft für Osteosynthesefragen (AO) principles of external fixation in having a pin near (third locking hole from the middle of the LCP) and far (seventh locking hole from the middle of the LCP) from the fracture site in both fragments, while being limited by the zone of injury and the length of the tibia.[22] This same configuration is used in the only other study[18] in the literature investigating the stiffness of the externalised LCP.The AO recommendation for traditional LCP fixation is to have a minimum of three screws in each fragment on either side of the fracture but, for the purposes of comparison, the same two-screws/Schanz pin positions were used in each fragment on either side of the fracture. We recognise that this is a limitation and an additional screw may influence the axial or torsional stiffness of the construct.Standard AO techniques were used for the fixation of the constructs, where all screw holes were prepared with drilling using a 4.3 mm drill bit through a locking barrel, and all screws that were used were fully threaded self-tapping 5 mm diameter locking screws (Synthes). The torque applied to each screw was standardised to 4 Nm using a torque-controlling screwdriver to lock the screw to the plate as recommended by the manufacturer. The screws used were slightly longer than the measured length to ensure that there was adequate purchase of the far cortex. The Schanz pins were advanced three ‘turns’ or approximately 3 mm once they had penetrated the far cortex to ensure adequate purchase.A 20 mm offset distance is kept between the bone and the plates/external fixator rods at the fracture site to account for soft tissue. This also allows for soft-tissue swelling without any interference with the construct.[23] We chose a 20 mm offset rather than a 30 mm offset in order to increase the stability of the construct and to prevent excessive micromotion at the fracture site. The lower profile plate also had the previously mentioned advantages of less obstruction during ambulation, and ease of fitting under clothing. Our choice of 20 mm was able to fit with adequate clearance from the soft tissue due to the subcutaneous nature of the anterior tibial border.
Mechanical testing
For mechanical testing, the proximal and distal ends of each tibia were potted in dental plaster (Fig. 1). Including the dental plaster, the total length of the construct was 465 mm. The position of the tibia was such that the line of action for the load went through the central axis of the construct, simulating the mechanical axis of the tibia. To limit all unwanted movement during testing, both ends of the tibia sawbone were potted.
Photograph of constructs potted in dental plaster. Left: unilateral external fixator; middle: externalised stainless steel locking compression plate; right: externalised titanium locking compression plate.Following fixation, the constructs were axially loaded using an Instron 5565 uniaxial mechanical testing machine (Instron, Norwood, Massachusetts) (Fig. 2). Axial loading was applied by directly placing the construct under the load cell.
Fig. 2
Diagram of constructs undergoing loading.
Diagram of constructs undergoing loading.In a pilot study, the static load to failure was determined to be comfortably in excess of 2100 N. Hence, an axial load of up to 2100 N was chosen, corresponding to three times the body weight of a 70 kg person, which is more than the world average weight of 62 kg.[24] The stiffness testing within this chosen loading range would not be confounded by catastrophic implant failure, as defined by implant failure or screw cut-out.For the quasi-static testing, axial loading force was gradually increased from 0 N to 2100 N at 20 N/s for six cycles. Interfragmentary motion at the fracture site was recorded at the far cortex, opposite, furthest from, the implant, using a contact extensometer with 0.01 mm resolution.[25] The far cortex was preferred over the near cortex as there is greater interfragmentary motion, and thus resulted in a lower calculated stiffness than the near cortex.Torsional testing was performed using a customised jig. Torque was applied up to 5 Nm, with every increment of 0.2 Nm and rotation of the specimen at the fracture site tracked using electromagnetic sensors (3D Guidance Model 130; Ascension Technology Corporation, Shelburne, Vermont). Torsional stiffness is determined as the average slope of the torque-rotation curve, and is expressed in Nm/degree. Each specimen was tested six times in axial compression and six times in torsion.Data were collected on a personal computer at a rate of 100 Hz. The first three cycles each in axial compression and in torsion were discarded to account for system settling. The gradient of the load-displacement curves of the final three cycles was analysed and the average axial stiffness for each specimen was calculated. Torsional stiffness was calculated by dividing the torque by the degree of rotation.Statistical analysis was performed using SPSS 17.0 (SPSS Inc., Chicago, Illinois). One-way analysis of variance (ANOVA) was used to compare the three groups for difference in axial and torsional stiffness, and post hoc testing was performed using Tukey if necessary. One-way ANOVA was chosen as it was the appropriate test to determine significant differences between the means of the three independent groups. A level of significance of p < 0.05 was used as the threshold for statistical significance.
Results
The mean axial stiffness was very similar for UEF (528, standard deviation (sd) 42 N/mm) and ESS-LCP (525, sd 69 N/mm), while it was slightly lower for ET-LCP (469, sd 37 N/mm). Table I shows the results in table form. Figure 3 shows the typical graph of compressive load (N) against displacement at the fracture site (mm). One-way ANOVA testing of the stiffness in all three groups demonstrated no significant difference (F(2,12) = 2.057, p = 0.171).
Table I.
Results for axial stiffness
Construct
Specimen
Axial stiffness (N/mm)
Mean axial stiffness (N/mm)
Standard deviation (N/mm)
p-value
0.171
Unilateral external fixator
1
488
528
42
2
493
3
576
4
511
5
571
Externalised Stainless Steel LCP
1
427
525
69
2
479
3
591
4
553
5
573
Externalised Titanium LCP
1
446
469
37
2
489
3
476
4
419
5
515
LCP, locking compression plate
Fig. 3
Typical graph of compressive load (N) against displacement at fracture site (mm) (ET-LCP, externalised titanium locking compression plate; ESS-LCP, externalised stainless steel locking compression plate; UEF, unilateral external fixator).
Results for axial stiffnessLCP, locking compression plateTypical graph of compressive load (N) against displacement at fracture site (mm) (ET-LCP, externalised titanium locking compression plate; ESS-LCP, externalised stainless steel locking compression plate; UEF, unilateral external fixator).There was a significant difference in mean torsional stiffness between the UEF (0.512, sd 0.104 Nm/degree), the ESS-LCP (0.686, sd 0.032 Nm/degree) and the ET-LCP (0.639, sd 0.089 Nm/degree), as determined by one-way ANOVA (F(2,12) = 6.204, p = 0.014). Table II shows the results in table form.
Table II.
Results for torsional stiffness
Construct
Specimen
Torsional stiffness (Nm/°)
Average torsional stiffness (Nm/°)
Standard deviation (Nm/°)
p-value
0.014
Unilateral external fixator
1
0.462
0.512
0.104
2
0.615
3
0.583
4
0.543
5
0.358
Externalised stainless steel LCP
1
0.646
0.686
0.032
2
0.673
3
0.710
4
0.727
5
0.675
Externalised titanium LCP
Externalised titanium LCP
1
0.493
0.639
0.089
2
0.690
3
0.656
4
0.632
5
0.723
LCP, locking compression plate
Results for torsional stiffnessLCP, locking compression plateA Tukey post hoc test revealed that the torsional stiffness of the ESS-LCP was statistically higher than that of UEF by 0.174, sd 0.051 Nm/degree (p = 0.013). However, there were no statistically significant differences between the UEF and ET-LCP or between the ESS-LCP and ET-LCP.No catastrophic failures, as defined by implant failure or screw cut-out, were observed in any of the three groups under both testing protocols.
Discussion
The technique of placing a plate outside the skin was first reported as the Zespol osteosynthesis system which had been developed in Poland in the 1970s.[26-28] There have been several case series reporting the use of the externalised LCP as an external fixator in various regions of the body.[7,29,30] However, the most common region is the tibia.[2,4-6,8,31,32] Our experience with the externalised LCP as an external fixator is also in the tibia. There have been several comparisons made between the externalised LCP and traditional external fixators.[1,4] However, the stiffness of external locked plating is not clear[2] as mechanical testing was not reported in the case reports and case series.Studies involving the mechanical properties of fracture-fixation constructs provide important information, which helps with our clinical decision-making. Few papers report on the stiffness of external fixators, with several historical studies indicating the range from 50 N/mm to 400 N/mm.[34-36] According to a study by Peindl et al,[12] the LCP had a stiffness of 264 N/mm, which was greater than a hybrid construct (a 165 mm “3/4” ring, two 2.0 mm cross wires tensioned to 130 kg oriented at 65° to each other, and one 5 mm Schanz “drop pin” and a “V-configured” carbon fiber rod frame), although the difference was not statistically significant. Therefore, by externalising the LCP and increasing the length of the screw between the plate and the bone, we believe the stiffness would decrease,[1,13,16] and we postulate that it would be comparable with that of an external fixator.In the fracture model we have presented, a comminuted fracture with a shaft defect was simulated without the possibility of bony contact. Therefore, our loading model primarily tested the stability of the bone-implant construct, where interfacing of the proximal and distal fragment was achieved through the UEF or the externalised LCP (ESS-LCP or ET-LCP). Based on this model system, results of this present study support the hypothesis that there was no significant difference in axial stiffness between the UEF, the ESS-LCP and the ET-LCP. There is also no significant difference in torsional stiffness between the UEF and the ET-LCP. However, the torsional stiffness of the ESS-LCP is significantly higher than that of the UEF.The axial stiffness of the UEF in our study is slightly higher but comparable with that in the literature, reported to be in the range of 50 N/mm to 400 N/mm.[32-35] We believe that the use of only one rod and a short distance of 20 mm between the bone and the rod may explain our findings of slightly higher construct stiffness.[1,13,16] Another possible explanation for the higher construct stiffness in our study could be our use of an extensometer to measure fracture site displacement, which results in a more accurate but higher stiffness value compared with the other studies which used actuator displacement instead.The axial stiffness of the ESS-LCP in our study (525 N/mm) is higher than that found in the only other study (112 N/mm) in the literature which investigated the biomechanical properties of an externalised LCP.[18] The reasons for this include the use of a greater offset of 30 mm between the bone and the plate in this other study, and the use of actuator displacement instead of an extensometer, resulting in a lower measured construct stiffness. That study made no attempt to measure the torsional stiffness of the externalised LCP.We acknowledge a high variability in the stiffness of the different specimens of the same construct. While every effort was made to produce identical specimens, variability in specimen stiffness is expected due to inadvertent small differences in fixation positions and variability in sawbones/implants. Multiple specimens were tested multiple times to reduce any error.To our knowledge, there is no study in the literature documenting the torsional stiffness of the externalised LCP. The torsional stiffness of the LCP used in the traditional manner is reported in the literature to be in the range of 0.1 Nm/degree to 6 Nm/degree.[11,16,17,36] The torsional stiffness of all of the constructs in our study is within this range, but at the lower end of the spectrum. This is to be expected as the torsional stiffness is anticipated to be lower as the offset between the bone and the plate is increased.Our study shows that the ESS-LCP has a significantly higher torsional stiffness than the UEF. This, however, does not adversely affect the use of the ESS-LCP as an alternative to the UEF, as the increased stiffness is still well within the range of torsional stiffness of the LCP and thus should not compromise fracture healing.Although not statistically significant, the observed lower axial and torsional stiffness of ET-LCP when compared with those of ESS-LCP was expected and this is due to titanium's lower Young's modulus in contrast with that of stainless steel. As the Young’s modulus of stainless steel is approximately twice that of titanium, we would expect the stiffness of stainless steel to be approximately twice that of titanium. However, our study showed that the axial stiffness of the ESS-LCP is only about 1.12 times that of ET-LCP while the torsional stiffness of the ESS-LCP is 1.21 times that of the ET-LCP. Another study, by Hoffmeier et al[14] also showed that their stainless steel implant has 1.3 times the stiffness of their titanium implant. We postulate that there are several other factors which may affect the calculations, and thus the difference in implant material alone does not account for a difference in the calculated stiffness.There are several similarities between the UEF and the externalised LCP. Both systems rely on fixed-angle stabilisation between Schanz pins or screws and the rod or plate that spans the fracture. The rod or plate is suspended at a fixed distance over the bone and thus there is neither compression nor resultant compromise on the periosteal blood supply.[37-39] Both techniques adopt a minimally invasive implantation technique and rely on secondary bone healing with callus formation.[41] To achieve a suitable mechanical environment for secondary bone healing, both techniques allow modulation of construct stiffness to a certain degree.[41]However, the extent, range and methods to which stiffness can be modulated differ between the two constructs. Changing the inter-pin distance and the distance of the pin to the fracture site can change the stiffness of the fixation construct.[42] It is more difficult to modify the stiffness of the externalised LCP as there is a limited number of locking screw holes which may be used. The removal of screws, the usage of titanium instead of stainless steel, and the use of a longer bridge span are all methods of decreasing the stiffness of the externalised LCP. Increasing the number of rods in the construct to increase the stiffness of the external fixators is not possible for the externalised LCP. Modifying the height of the rod/plate is also much easier with the external fixator.Another difference is that use of the externalised LCP is technically challenging as compared with external fixation as it requires a good reduction before fixation, due to the limited adjustments that can be made after fixation. The high flexibility due to the adjustable clamps and the easy implantation technique are advantages of the external fixator.In 1996, Kowalski et al[23] concluded that when the non-contact plate is elevated 20 mm from the bone surface it behaves similarly to a small external fixator, but does not provide the stiffness and strength necessary for supporting the loads applied to lower extremity. However, in their study, the configuration of the non-contact plate and external fixator is vastly different, including the use of a short plate with all the screws close to the fracture site for the non-contact plate, compared with a longer external fixator with Schanz pins near and far from the fracture site. The current LCP in the market is also vastly different from the non-contact plate used in that study.Our study has shown that the externalised LCP is able to withstand up to three times the bodyweight of an average 70 kg adult on axial loading only, without catastrophic failure, opening up the possibility that it may provide enough stiffness and strength to allow partial weight bearing initially, and progression to full weight- bearing. However, further mechanical testing is required to ascertain the fatigue strength.Ahmad et al[15] reported that a 5 mm plate elevation decreased construct strength in axial compression by 63%. Their study, as with most other biomechanical studies that compared LCP elevated up to 20 mm with LCPs fixed to bone, observed decreased strength and stiffness in axial compression. However, no studies have made direct comparison with external fixators. Furthermore, the configurations are mostly with shorter plates with many screws close to the fracture site.In a similar experimental set-up performed by Bottlang et al,[13] the axial stiffness of a LCP with a 1 mm offset from the bone was 2900 N/mm. With the offset increased to 20 mm in our study, our calculated axial stiffness is correspondingly much lower.This study had several limitations. First, it only tests the axial and torsional stiffness of the constructs. Ideally, we would like to test the constructs in four-point bending as well. However, limitations of laboratory equipment did not allow us to do so. Second, as with all biomechanical studies, the artificially applied load used in this model may not represent the multifaceted manner of loading that occurs in vivo. During fatigue tests, a combined load regime of both axial compression and torsion acting on the construct is likely to decrease the life of the implant.Synthetic bone was used instead of cadaver tibias to eliminate variations in geometry and in material properties associated with humantibias, so that any differences found in testing could be attributed to the fixation devices. The use of Sawbones for mechanical testing is well established[43,44] and minimises the variation in stiffness found in cadaveric bones with differences in age and bone quality.[45] These large-size fourth-generation composite replicate bones exhibited intra-specimen variations of less than 10% for all cases and were also found to have similar mechanical properties to healthy adult bones, and thus are close to ideal replicas for standardisation in biomechanical analyses.[19,20]Although this model did not take into account the actual muscle forces acting on the tibial diaphysis, we feel that it was appropriate for comparing the relative stability and stiffness of the three construct groups.There are several case reports and case series of the externalised LCP being used as an external fixator in current practice. However, there has been limited mechanical evidence supporting this practice, with only one study[18] in the literature evaluating the biomechanical properties of the externalised LCP and none comparing it with an external fixator.This study provides biomechanical evidence to support the use of LCP as an external fixator. We found that the axial and torsional stiffness of the externalised LCP are comparable with those of the external fixator and thus may provide a viable alternative.In conclusion, using the externalised LCP as an external fixator may provide a viable and attractive alternative to traditional UEF as its lower profile makes it more acceptable to patients, while not compromising on axial and torsional stiffness.
Authors: P L N Fernando; Aravinda Abeygunawardane; Pci Wijesinghe; Parakrama Dharmaratne; Pujitha Silva Journal: Med Eng Phys Date: 2021-11-04 Impact factor: 2.242