Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability?

M Ahmad1, R Nanda, A S Bajwa, J Candal-Couto, S Green, A C Hui.   

Abstract

AIM: To investigate in vitro the mechanical stability of a locking compression plate (LCP) construct in a simulated diaphyseal fracture of the humerus at increasing distances between the plate and bone. MATERIALS AND
METHOD: A series of biomechanical in vitro experiments were performed using composite humerus sawbone as the bone model. Osteotomy created in the mid-diaphyseal region. A 10mm osteotomy gap was bridged with a seven-hole 4.5 stainless steel plate with one of four methods: a control group consisted of a dynamic compression plate applied flush to the bone and three study groups which comprised of a LCP applied flush to the bone, at 2mm and at 5mm from the bone. Standard AO technique used with locking head screws used for LCP fixation. Static and dynamic loading tests performed in a custom made jig in which the bone model was fixed both proximally and distally. Samples were subjected to cyclical compression, compression load to failure, cyclical torque and torque to failure. Plastic deformation and failure was assessed using three-dimensional measurements. Scanning electron microscopy of the plate and screw surface allowed detailed inspection of micro-fracture in areas of fatigue.
RESULTS: Comparable results were achieved in both the DCP and LCP constructs in which the plate was applied at or less than 2mm from the bone. When applied 5mm from the bone the LCP demonstrated significantly increased plastic deformation during cyclical compression and required lower loads to induce construct failure.
CONCLUSION: At a distance 5 mm we observe an inferior performance in the mechanical properties of the LCP construct with decrease in axial stiffness and torsional rigidity. If it is desirable for an LCP to be used the distance between plate and bone should be <or=2 mm.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17296199     DOI: 10.1016/j.injury.2006.08.058

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  35 in total

1.  Effects of construct stiffness on healing of fractures stabilized with locking plates.

Authors:  Michael Bottlang; Josef Doornink; Trevor J Lujan; Daniel C Fitzpatrick; J Lawrence Marsh; Peter Augat; Brigitte von Rechenberg; Maren Lesser; Steven M Madey
Journal:  J Bone Joint Surg Am       Date:  2010-12       Impact factor: 5.284

2.  Computational modelling of long bone fractures fixed with locking plates - How can the risk of implant failure be reduced?

Authors:  M Nassiri; B Macdonald; J M O'Byrne
Journal:  J Orthop       Date:  2013-03-07

3.  The dynamic locking screw (DLS) can increase interfragmentary motion on the near cortex of locked plating constructs by reducing the axial stiffness.

Authors:  Stefan Döbele; Carsten Horn; Stefan Eichhorn; Arne Buchholtz; Andreas Lenich; Rainer Burgkart; Andreas K Nüssler; Martin Lucke; Daniel Andermatt; Rudolf Koch; Ulrich Stöckle
Journal:  Langenbecks Arch Surg       Date:  2010-04-01       Impact factor: 3.445

4.  Morphometry of Proximal Femur in Indian Population.

Authors:  Minakshi Verma; Sheetal Joshi; Anita Tuli; Shashi Raheja; Priyanka Jain; Priyanka Srivastava
Journal:  J Clin Diagn Res       Date:  2017-02-01

5.  Biomechanical comparison of gourd-shaped LCP versus LCP for fixation of comminuted tibial shaft fracture.

Authors:  Guo-Hui Xu; Bo Liu; Qi Zhang; Juan Wang; Wei Chen; Yue-Ju Liu; A-Qin Peng; Ying-Ze Zhang
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2013-04-17

6.  Mismatch of AO anatomically shaped distal humeral plate with humeral shaft forward flexion angulation in adult Chinese population.

Authors:  Zhiyong Zhou; Zhifeng Tang; Xueqin Zhao; Wentao Chen; Xu Chen; Mingzhang Mu; Shudong Zhang; Bin Ning; Tanghong Jia
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-03-22

7.  [Mechanobiology of fracture healing part 2 : Relevance for internal fixation of fractures].

Authors:  L Claes
Journal:  Unfallchirurg       Date:  2017-01       Impact factor: 1.000

8.  Mechanical evaluation of fourth-generation composite femur hybrid locking plate constructs.

Authors:  Tarun Goswami; Vinit Patel; David J Dalstrom; Michael J Prayson
Journal:  J Mater Sci Mater Med       Date:  2011-07-17       Impact factor: 3.896

9.  Dual plating of humeral shaft fractures: orthogonal plates biomechanically outperform side-by-side plates.

Authors:  Victor Kosmopoulos; Arvind D Nana
Journal:  Clin Orthop Relat Res       Date:  2013-11-12       Impact factor: 4.176

10.  Biomechanical investigation of an alternative concept to angular stable plating using conventional fixation hardware.

Authors:  Markus Windolf; Kajetan Klos; Dirk Wähnert; Bas van der Pol; Roman Radtke; Karsten Schwieger; Roland P Jakob
Journal:  BMC Musculoskelet Disord       Date:  2010-05-21       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.