Literature DB >> 20399424

A biomechanical comparison of the Surgical Implant Generation Network (SIGN) tibial nail with the standard hollow nail.

L A Calafi1, T Antkowiak, S Curtiss, C P Neu, D Moehring.   

Abstract

OBJECTIVE: In developing countries, tibial shaft fractures are frequently stabilised using Surgical Implant Generation Network (SIGN) nails. Despite widespread use throughout the world, little is known regarding their biomechanical properties. This study aimed to compare the mechanical stiffness of the SIGN tibial nail with a standard hollow tibial nail.
METHODS: A fracture gap model was created to simulate a comminuted mid-shaft tibia fracture (AO/OTA42-C3) using synthetic composite bones. The constructs were stabilised with either a 9 mm solid SIGN nail or a 10 mm hollow Russell-Taylor nail. Both nail systems were interlocked proximally and distally. Following fixation, the specimens were loaded in axial, torsional, and cyclical axial modes to calculate construct stiffness and irreversible (plastic) deformation.
RESULTS: The mean axial stiffness for the SIGN nail constructs was 47% higher than mean stiffness for the RT nail constructs (p<0.001). The difference in torsional stiffness was not statistically significant. However, the SIGN group demonstrated 159% more irreversible deformation than the Russell-Taylor group (p=0.006) for the loading parameters studied.
CONCLUSION: The SIGN tibial nail, despite its slightly smaller diameter, can provide similar construct stiffness and stability, when compared to a larger hollow nail for stabilisation of tibial shaft fractures. 2010 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20399424     DOI: 10.1016/j.injury.2010.03.003

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  4 in total

1.  Achieving locked intramedullary fixation of long bone fractures: technology for the developing world.

Authors:  Jonathan Phillips; Lewis G Zirkle; Richard A Gosselin
Journal:  Int Orthop       Date:  2012-07-31       Impact factor: 3.075

2.  Functional outcome of patients of tibial fracture treated with solid nail (SIGN nail) versus conventional hollow nail - A randomized trial.

Authors:  Rajiv Maharjan; Bikram Prasad Shrestha; Pashupati Chaudhary; Raju Rijal; Rosan Prasad Shah Kalawar
Journal:  J Clin Orthop Trauma       Date:  2020-08-01

3.  Prospects of implant with locking plate in fixation of subtrochanteric fracture: experimental demonstration of its potential benefits on synthetic femur model with supportive hierarchical nonlinear hyperelastic finite element analysis.

Authors:  Mohammed Hadi Latifi; Kunalan Ganthel; Shanmugam Rukmanikanthan; Azura Mansor; Tunku Kamarul; Mehmet Bilgen
Journal:  Biomed Eng Online       Date:  2012-04-30       Impact factor: 2.819

4.  Externalised locking compression plate as an alternative to the unilateral external fixator: a biomechanical comparative study of axial and torsional stiffness.

Authors:  B F H Ang; J Y Chen; A K S Yew; S K Chua; S M Chou; S L Chia; J S B Koh; T S Howe
Journal:  Bone Joint Res       Date:  2017-04       Impact factor: 5.853

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.