Evaluation of different low-dose multidetector CT and cone beam CT protocols in maxillary sinus imaging: part I-an in vitro study.

Abeer A Almashraqi1,2, Eman A Ahmed3,4, Nashwa S Mohamed3,5, Imad H Barngkgei6,7, Nader A Elsherbini8, Esam S Halboub2,9.   

Abstract

OBJECTIVES: This in vitro study aimed to assess radiation dose and image quality of different low-dose multidetector CT (MDCT) and CBCT imaging protocols in comparison with the standard MDCT protocol for maxillary sinus imaging.
METHODS: Effective dose (E) and image quality of 10 MDCT (changing effective milliampere second starting from 141.3 EmAs to 20 EmAs) and 3 CBCT protocols (changing milliampere second and voxel size) were assessed throughout scanning an anthropomorphic head and neck Alderson Rando phantom. E values were calculated using thermoluminescent dosemeters (TLDs) fixed at 6 sensitive organs (14 sites) on the Rando phantom. Image quality was assessed objectively (by calculating the standard deviation values of the radiographic density of water) and subjectively (by assessing the diagnostic image quality using a four-graded scale: 1 = very good, 2 = good, 3 = acceptable and 4 = unacceptable).
RESULTS: Two MDCT protocols (120 kV/32 EmA and 120 kV/25 EmA) had lower radiation doses with statistically significant differences (p < 0.001) compared with that of the standard MDCT protocol (120 kV/141.3 EmA), and they preserved a good diagnostic image quality. One CBCT protocol (120 kV/20 mA) had a reasonable radiation dose and good image quality. There were no statistically significant differences between the above-mentioned lower dose MDCT and CBCT protocols (p > 0.05) with respect to the radiation dose and image quality.
CONCLUSIONS: The low-dose MDCT and CBCT protocols are viable methods for maxillary sinus examination as evaluated using the above-mentioned phantom that yield a good diagnostic image quality using E approximately 7 and 11 times lower than that of the standard MDCT, respectively. These findings were evaluated in the in vivo part of this project.

Entities:  

Keywords:  CBCT; Effective dose; MDCT; dose reduction; maxillary sinus

Mesh:

Year:  2017        PMID: 28266870      PMCID: PMC5606275          DOI: 10.1259/dmfr.20160323

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  34 in total

1.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

Review 2.  Principles of CT: radiation dose and image quality.

Authors:  Lee W Goldman
Journal:  J Nucl Med Technol       Date:  2007-11-15

3.  Effective dose from cone beam CT examinations in dentistry.

Authors:  J A Roberts; N A Drage; J Davies; D W Thomas
Journal:  Br J Radiol       Date:  2008-10-13       Impact factor: 3.039

4.  What is cone-beam CT and how does it work?

Authors:  William C Scarfe; Allan G Farman
Journal:  Dent Clin North Am       Date:  2008-10

5.  The effect of decreasing mAs on image quality and patient dose in sinus CT.

Authors:  S A Sohaib; P D Peppercorn; J A Horrocks; M H Keene; G S Kenyon; R H Reznek
Journal:  Br J Radiol       Date:  2001-02       Impact factor: 3.039

6.  Low-dose dental computed tomography: significant dose reduction without loss of image quality.

Authors:  P Rustemeyer; U Streubühr; J Suttmoeller
Journal:  Acta Radiol       Date:  2004-12       Impact factor: 1.990

7.  Comparison between low-dose and standard-dose multidetector CT in patients with suspected chronic sinusitis.

Authors:  Denis Tack; Jacques Widelec; Viviane De Maertelaer; Jean-Marie Bailly; Christian Delcour; Pierre Alain Gevenois
Journal:  AJR Am J Roentgenol       Date:  2003-10       Impact factor: 3.959

Review 8.  CT radiation dose optimization and estimation: an update for radiologists.

Authors:  Hyun Woo Goo
Journal:  Korean J Radiol       Date:  2011-12-23       Impact factor: 3.500

9.  Analysis of location and prevalence of maxillary sinus septa.

Authors:  Won-Jin Lee; Seung-Jae Lee; Hyoung-Seop Kim
Journal:  J Periodontal Implant Sci       Date:  2010-04-15       Impact factor: 2.614

10.  Occurrence of maxillary sinus abnormalities detected by cone beam CT in asymptomatic patients.

Authors:  Inara Carneiro Costa Rege; Thiago Oliveira Sousa; Cláudio Rodrigues Leles; Elismauro Francisco Mendonça
Journal:  BMC Oral Health       Date:  2012-08-10       Impact factor: 2.757

View more
  3 in total

Review 1.  Radiation dose in non-dental cone beam CT applications: a systematic review.

Authors:  Cosimo Nardi; Sergio Salerno; Roberto Molteni; Mariaelena Occhipinti; Giulia Grazzini; Niccolò Norberti; Cesare Cordopatri; Stefano Colagrande
Journal:  Radiol Med       Date:  2018-06-05       Impact factor: 3.469

2.  Transfer learning in diagnosis of maxillary sinusitis using panoramic radiography and conventional radiography.

Authors:  Shinya Kotaki; Takahito Nishiguchi; Marino Araragi; Hironori Akiyama; Motoki Fukuda; Eiichiro Ariji; Yoshiko Ariji
Journal:  Oral Radiol       Date:  2022-09-27       Impact factor: 1.882

3.  Odontogenic causes complicating the chronic rhinosinusitis diagnosis.

Authors:  Annina Wuokko-Landén; Karin Blomgren; Anni Suomalainen; Hannamari Välimaa
Journal:  Clin Oral Investig       Date:  2020-06-04       Impact factor: 3.573

  3 in total