Low-dose dental computed tomography: significant dose reduction without loss of image quality.

P Rustemeyer1, U Streubühr, J Suttmoeller.   

Abstract

PURPOSE: To measure and reduce the patient dose during computed tomography (CT) for dental applications.
MATERIAL AND METHODS: Lithium fluoride thermoluminescent dosimeters were implanted in a tissue-equivalent humanoid phantom (Alderson-Rando-Phantom) to determine doses to the thyroid gland, the active bone marrow, the salivary glands, and the eye lens. Dental CT was performed with spiral CT and a dental software package. The usual dental CT technique was compared with a new dose-reduced protocol, which delivered best image quality at lowest possible radiation dose, as tested in a preceding study. Image quality was analysed using a human anatomic head preparation. In addition, the radiation dose was compared with panoramic radiography and digital volume tomography (DVT). Eight radiologists evaluated all images in a blinded fashion. A Wilcoxon rank pair test was used for statistical evaluation.
RESULTS: Radiation dose could be reduced by a factor of 9 (max.) with the new dose-reduced protocol (e.g. bone marrow dose from 23.6 mSv to 2.9 mSv; eye lens from 0.5 mSv to 0.3 mSv; thyroid gland from 2.5 mSv to 0.5 mSv; parotid glands from 2.3 mSv to 0.4 mSv). Dose reduction did not reduce image quality or diagnostic information.
CONCLUSION: A considerable dose reduction without loss of diagnostic information is achievable in dental CT. As radiation exposure of the presented low-dose protocol is expected to be in the same range as DVT, low-dose dental CT might be superior to DVT, because CT can be used to evaluate soft tissues as well.

Entities:  

Mesh:

Year:  2004        PMID: 15690615     DOI: 10.1080/02841850410001402

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  10 in total

1.  Evaluation of the effective dose of cone beam CT and multislice CT for temporomandibular joint examinations at optimized exposure levels.

Authors:  N Kadesjö; D Benchimol; B Falahat; K Näsström; X-Q Shi
Journal:  Dentomaxillofac Radiol       Date:  2015-06-02       Impact factor: 2.419

2.  Evaluation of different low-dose multidetector CT and cone beam CT protocols in maxillary sinus imaging: part I-an in vitro study.

Authors:  Abeer A Almashraqi; Eman A Ahmed; Nashwa S Mohamed; Imad H Barngkgei; Nader A Elsherbini; Esam S Halboub
Journal:  Dentomaxillofac Radiol       Date:  2017-04-12       Impact factor: 2.419

3.  Image quality optimization using a narrow vertical detector dental cone-beam CT.

Authors:  Danieli Moura Brasil; Ruben Pauwels; Wim Coucke; Francisco Haiter-Neto; Reinhilde Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2019-01-31       Impact factor: 2.419

4.  Comparative study of image quality and radiation dose of cone beam and low-dose multislice computed tomography--an in-vitro investigation.

Authors:  Elisabeth Hofmann; Matthias Schmid; Martin Sedlmair; Rosemarie Banckwitz; Ursula Hirschfelder; Michael Lell
Journal:  Clin Oral Investig       Date:  2013-03-05       Impact factor: 3.573

5.  Cone beam computed tomography and low-dose multislice computed tomography in orthodontics and dentistry: a comparative evaluation on image quality and radiation exposure.

Authors:  E Hofmann; M Schmid; M Lell; U Hirschfelder
Journal:  J Orofac Orthop       Date:  2014-08-28       Impact factor: 1.938

6.  Validity of linear measurements of the jaws using ultralow-dose MDCT and the iterative techniques of ASIR and MBIR.

Authors:  Asma'a A Al-Ekrish; Reema Al-Shawaf; Peter Schullian; Ra'ed Al-Sadhan; Romed Hörmann; Gerlig Widmann
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-06-02       Impact factor: 2.924

7.  Effects of exposure parameters and slice thickness on detecting clear and unclear mandibular canals using cone beam CT.

Authors:  Gainer R Jasa; Mayumi Shimizu; Kazutoshi Okamura; Kenji Tokumori; Yohei Takeshita; Warangkana Weerawanich; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

8.  Impact of Cone Beam Computed Tomography Dose in Pre-Surgical Implant Analysis.

Authors:  Nabil El Sahili; Ibrahim Nasseh; Antoine Berberi; Sandra David-Tchouda; Sophie Thoret; Thomas Fortin
Journal:  Open Dent J       Date:  2018-01-31

9.  Localization of the inferior alveolar canal using ultralow dose CT with iterative reconstruction techniques.

Authors:  Asmaa A Al-Ekrish; Wafa Alfaleh; Romed Hörmann; Ameera Alabdulwahid; Wolfgang Puelacher; Gerlig Widmann
Journal:  Dentomaxillofac Radiol       Date:  2018-07-23       Impact factor: 2.419

10.  Comparability of dental implant site ridge measurements using ultra-low-dose multidetector row computed tomography combined with filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction.

Authors:  Asma'a Abdurrahman Al-Ekrish; Reema Al-Shawaf; Wafa Alfaleh; Romed Hörmann; Wolfgang Puelacher; Gerlig Widmann
Journal:  Oral Radiol       Date:  2018-10-13       Impact factor: 1.852

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.