Hand Surgery and Fluoroscopic Eye Radiation Dosage: A Prospective Pilot Comparison of Large Versus Mini C-Arm Fluoroscopy Use.

Mark L Wang1, C Edward Hoffler1, Asif M Ilyas1, William H Kirkpatrick1, Pedro K Beredjiklian1, Charles F Leinberry1.   

Abstract

Background: The purpose of this study is to (1) perform a prospective pilot comparison of the impact of large versus mini C-arm fluoroscopy on resultant eye radiation exposure and (2) test the hypothesis that the use of either modality during routine hand surgery does not exceed the current recommended limits to critical eye radiation dosage.
Methods: Over a 12-month period, eye radiation exposure was prospectively measured by a board-certified hand surgeon using both large and mini C-arm fluoroscopy. For each modality, accumulated eye radiation dosage was measured monthly, while fluoroscopic radiation output was recorded, including total exposure time and dose rate.
Results: A total of 58 cases were recorded using large C-arm and 25 cases using mini C-arm. Between the 2 groups, there was not a significant difference with total exposure time (P = .88) and average dose rate per case (P = .10). With the use of either modality, average monthly eye radiation exposure fell within the undetectable range (<30 mrem), significantly less than the current recommended limit of critical eye radiation (167 mrem/month). Conclusions: The impact of various fluoroscopic sources on eye radiation exposure remains relatively unexplored. In this study, the minimal detectable eye radiation dosages observed in both groups were reliably consistent. Our findings suggest that accumulated eye radiation dosage, from the use of either fluoroscopic modality, does not approach previously reported levels of critical radiation loads.

Entities:  

Keywords:  cataracts; eye radiation; fluoroscopy; mini C-arm; radiation; radiation exposure

Mesh:

Year:  2016        PMID: 28082838      PMCID: PMC5207279          DOI: 10.1177/1558944716643279

Source DB:  PubMed          Journal:  Hand (N Y)        ISSN: 1558-9447


  23 in total

1.  Development of light dissemination in the rat lens after in vivo exposure to radiation in the 300-nm wavelength region.

Authors:  P G Söderberg
Journal:  Ophthalmic Res       Date:  1990       Impact factor: 2.892

2.  Radiation cataract risk in interventional cardiology personnel.

Authors:  Eliseo Vano; Norman J Kleiman; Ariel Duran; Madan M Rehani; Dario Echeverri; Mariana Cabrera
Journal:  Radiat Res       Date:  2010-10       Impact factor: 2.841

3.  Radiation exposure with use of the mini-C-arm for routine orthopaedic imaging procedures.

Authors:  Brian L Badman; Lynn Rill; Bradley Butkovich; Manuel Arreola; Robert A Vander Griend
Journal:  J Bone Joint Surg Am       Date:  2005-01       Impact factor: 5.284

4.  Radiation exposure in hand surgery: mini versus standard C-arm.

Authors:  George S Athwal; Reuben A Bueno; Scott W Wolfe
Journal:  J Hand Surg Am       Date:  2005-11       Impact factor: 2.230

5.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

6.  Exposure from the large C-arm versus the mini C-arm using hand/wrist and elbow phantoms.

Authors:  Gordon Singer; Brent Herron; David Herron
Journal:  J Hand Surg Am       Date:  2011-04       Impact factor: 2.230

Review 7.  Radiation and cataract.

Authors:  Madan M Rehani; Eliseo Vano; Olivera Ciraj-Bjelac; Norman J Kleiman
Journal:  Radiat Prot Dosimetry       Date:  2011-07-14       Impact factor: 0.972

8.  Diagnostic x-ray exposure and lens opacities: the Beaver Dam Eye Study.

Authors:  B E Klein; R Klein; K L Linton; T Franke
Journal:  Am J Public Health       Date:  1993-04       Impact factor: 9.308

9.  Risk of cataract after exposure to low doses of ionizing radiation: a 20-year prospective cohort study among US radiologic technologists.

Authors:  Gabriel Chodick; Nural Bekiroglu; Michael Hauptmann; Bruce H Alexander; D Michal Freedman; Michele Morin Doody; Li C Cheung; Steven L Simon; Robert M Weinstock; André Bouville; Alice J Sigurdson
Journal:  Am J Epidemiol       Date:  2008-07-29       Impact factor: 4.897

10.  Oxidative DNA damage in patients with cataract.

Authors:  Orhan Ates; Hakan Hamit Alp; Ibrahim Kocer; Orhan Baykal; Ilknur A Salman
Journal:  Acta Ophthalmol       Date:  2010-12       Impact factor: 3.761

View more
  2 in total

1.  Elevated Radiation Exposure Associated With Above Surface Flat Detector Mini C-Arm Use.

Authors:  Dennis P Martin; Talia Chapman; Christopher Williamson; Brian Tinsley; Asif M Ilyas; Mark L Wang
Journal:  Hand (N Y)       Date:  2017-11-22

2.  The Effect of Metal Instrumentation on Patient and Surgical Team Scatter Radiation Exposure Using Mini C-Arm in a Simulated Forearm Fracture Fixation Model.

Authors:  Michael T Groover; Jacob R Hinkley; Daniel E Gerow; H Brent Bamberger; Jennifer Evans; Roland E Gazaille
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2019-06-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.