Elevated Radiation Exposure Associated With Above Surface Flat Detector Mini C-Arm Use.

Dennis P Martin1, Talia Chapman1, Christopher Williamson1, Brian Tinsley1, Asif M Ilyas1, Mark L Wang1.   

Abstract

Background: This study aims to test the hypothesis that: (1) radiation exposure is increased with the intended use of Flat Surface Image Intensifier (FSII) units above the operative surface compared with the traditional below-table configuration; (2) this differential increases in a dose-dependent manner; and (3) radiation exposure varies with body part and proximity to the radiation source.
Methods: A surgeon mannequin was seated at a radiolucent hand table, positioned for volar distal radius plating. Thermoluminescent dosimeters measured exposure to the eyes, thyroid, chest, hand, and groin, for 1- and 15-minute trials from a mini C-arm FSII unit positioned above and below the operating surface. Background radiation was measured by control dosimeters placed within the operating theater.
Results: At 1-minute of exposure, hand and eye dosages were significantly greater with the flat detector positioned above the table. At 15-minutes of exposure, hand radiation dosage exceeded that of all other anatomic sites with the FSII in both positions. Hand exposure was increased in a dose-dependent manner with the flat detector in either position, whereas groin exposure saw a dose-dependent only with the flat detector beneath the operating table. Conclusions: These findings suggest that the surgeon's hands and eyes may incur greater radiation exposure compared with other body parts, during routine mini C-arm FSII utilization in its intended position above the operating table. The clinical impact of these findings remains unclear, and future long-term radiation safety investigation is warranted. Surgeons should take precautions to protect critical body parts, particularly when using FSII technology above the operating with prolonged exposure time.

Keywords:  eye radiation; flat detector; fluoroscopy; hand radiation; mini C-arm; radiation

Mesh:

Year:  2017        PMID: 29166785      PMCID: PMC6760076          DOI: 10.1177/1558944717743600

Source DB:  PubMed          Journal:  Hand (N Y)        ISSN: 1558-9447


  17 in total

1.  Radiation exposure with use of the mini-C-arm for routine orthopaedic imaging procedures.

Authors:  Brian L Badman; Lynn Rill; Bradley Butkovich; Manuel Arreola; Robert A Vander Griend
Journal:  J Bone Joint Surg Am       Date:  2005-01       Impact factor: 5.284

2.  Radiation exposure to the hands from mini C-arm fluoroscopy.

Authors:  Gordon Singer
Journal:  J Hand Surg Am       Date:  2005-07       Impact factor: 2.230

3.  Exposure to direct and scatter radiation with use of mini-c-arm fluoroscopy.

Authors:  Brian D Giordano; Steven Ryder; Judith F Baumhauer; Benedict F DiGiovanni
Journal:  J Bone Joint Surg Am       Date:  2007-05       Impact factor: 5.284

Review 4.  New ICRP recommendations.

Authors:  A D Wrixon
Journal:  J Radiol Prot       Date:  2008-05-22       Impact factor: 1.394

5.  Exposure from the large C-arm versus the mini C-arm using hand/wrist and elbow phantoms.

Authors:  Gordon Singer; Brent Herron; David Herron
Journal:  J Hand Surg Am       Date:  2011-04       Impact factor: 2.230

6.  Measurements of surgeons' exposure to ionizing radiation dose: comparison of conventional and mini C-arm fluoroscopy.

Authors:  K H Sung; E Min; C Y Chung; B C Jo; M S Park; K Lee
Journal:  J Hand Surg Eur Vol       Date:  2015-06-26

7.  Fluoroscopic radiation exposure: are we protecting ourselves adequately?

Authors:  C Edward Hoffler; Asif M Ilyas
Journal:  J Bone Joint Surg Am       Date:  2015-05-06       Impact factor: 5.284

8.  Increased breast cancer prevalence among female orthopedic surgeons.

Authors:  Loretta B Chou; Sheena Chandran; Alex H S Harris; Joanna Tung; Lesley M Butler
Journal:  J Womens Health (Larchmt)       Date:  2012-03-20       Impact factor: 2.681

9.  Breast Radiation Exposure in Female Orthopaedic Surgeons.

Authors:  Lindsey C Valone; Monique Chambers; Lisa Lattanza; Michelle A James
Journal:  J Bone Joint Surg Am       Date:  2016-11-02       Impact factor: 5.284

10.  Risk of cataract after exposure to low doses of ionizing radiation: a 20-year prospective cohort study among US radiologic technologists.

Authors:  Gabriel Chodick; Nural Bekiroglu; Michael Hauptmann; Bruce H Alexander; D Michal Freedman; Michele Morin Doody; Li C Cheung; Steven L Simon; Robert M Weinstock; André Bouville; Alice J Sigurdson
Journal:  Am J Epidemiol       Date:  2008-07-29       Impact factor: 4.897

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.