Radiation exposure in hand surgery: mini versus standard C-arm.

George S Athwal1, Reuben A Bueno, Scott W Wolfe.   

Abstract

PURPOSE: The use of intraoperative fluoroscopy in hand surgery is common. Two types of fluoroscopic units are available: the mini C-arm and the standard C-arm. There is little literature on the radiation exposure from the mini C-arm, therefore, the primary goal of this study was to quantify and compare the amount of radiation exposure to members of the surgical team (surgeon, first assistant, nurse, anesthesiologist) using both standard and mini C-arms in a simulated wrist surgery setup. Mini C-arm positioning was also examined to determine the safest configuration to minimize radiation exposure to surgeons.
METHODS: Radiation dosimeters were used to test 2 commercially available fluoroscopy units in a simulated wrist surgery setup with a cadaveric upper extremity. Several different configurations of the C-arms were tested to determine radiation exposure rates to surgeons and the operating room staff.
RESULTS: The mean in-beam radiation exposures with the use of the mini and standard C-arms were 3,720 mR/h and 6,540 mR/h, respectively. The mini C-arm had universally less radiation exposure than the standard C-arm in the clinical configurations tested. The safest configuration of mini C-arm use to minimize radiation exposure was with the surgeon standing on the image intensifier side of the unit as compared with the source side. Mini C-arm radiation exposure to the hands, groin, chest, and thyroid of the operating surgeons were well below the National Council of Radiation Protection and Measurement's annual dose limits.
CONCLUSIONS: In the clinical configurations tested in this study the mini C-arm had lower radiation exposures than the standard C-arm. To reduce radiation exposure maximally surgeons should stand behind the lead-encased image intensifier and should use techniques to reduce exposure.

Mesh:

Year:  2005        PMID: 16344194     DOI: 10.1016/j.jhsa.2005.06.023

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  11 in total

1.  Closed reduction internal fixation rates and procedure times for metacarpal fractures treated in a minor surgery area before and after the introduction of a mini C-arm unit.

Authors:  R C Mahabir; C M DeCroff; L Thurgood; A R Harrop
Journal:  Can J Plast Surg       Date:  2008

2.  Mini C-Arm Fluoroscopy: Does Its Configuration Matter for Radiation Exposure to the Surgeon?

Authors:  Talia Chapman; Dennis P Martin; Christopher Williamson; Brian Tinsley; Mark L Wang; Asif M Ilyas
Journal:  Hand (N Y)       Date:  2017-06-23

3.  Hand Surgery and Fluoroscopic Eye Radiation Dosage: A Prospective Pilot Comparison of Large Versus Mini C-Arm Fluoroscopy Use.

Authors:  Mark L Wang; C Edward Hoffler; Asif M Ilyas; William H Kirkpatrick; Pedro K Beredjiklian; Charles F Leinberry
Journal:  Hand (N Y)       Date:  2016-04-06

4.  Radiation Exposure and Hand Dominance Using Mini C-Arm Fluoroscopy in Hand Surgery.

Authors:  Corey B Fuller; Montri D Wongworawat; Barth B Riedel
Journal:  Hand (N Y)       Date:  2016-02-02

5.  Elevated Radiation Exposure Associated With Above Surface Flat Detector Mini C-Arm Use.

Authors:  Dennis P Martin; Talia Chapman; Christopher Williamson; Brian Tinsley; Asif M Ilyas; Mark L Wang
Journal:  Hand (N Y)       Date:  2017-11-22

6.  Establishing local diagnostic reference levels for common orthopaedic procedures using the mini C-arm fluoroscope.

Authors:  Anna Kropelnicki; Rosemary Eaton; Alexandra Adamczyk; Jacqueline Waterman; Pegah Mohaghegh
Journal:  Br J Radiol       Date:  2020-10-22       Impact factor: 3.039

7.  Assessment of Radiation Protection in Hand-Shielding Products With Mini C-Arm Fluoroscopy.

Authors:  Matthew B Cantlon; Asif M Ilyas
Journal:  Hand (N Y)       Date:  2019-08-12

8.  Radiation Exposure to the Hand of a Spinal Interventionalist during Fluoroscopically Guided Procedures.

Authors:  Kazuta Yamashita; Hisanori Ikuma; Takuya Tokashiki; Takashi Maehara; Akihiro Nagamachi; Yoichiro Takata; Toshinori Sakai; Kosaku Higashino; Koichi Sairyo
Journal:  Asian Spine J       Date:  2017-02-17

9.  How to reduce radiation exposure during hand surgery: use of a kidney dish to protect the surgeon's hands.

Authors:  E A O Lindisfarne; A Baskaradas; K Kunasingam
Journal:  Ann R Coll Surg Engl       Date:  2012-05       Impact factor: 1.951

10.  Pulsation and Collimation During Fluoroscopy to Decrease Radiation: A Cadaver Study.

Authors:  Kazuta Yamashita; Kosaku Higashino; Hiroaki Hayashi; Fumio Hayashi; Yoshihiro Fukui; Koichi Sairyo
Journal:  JB JS Open Access       Date:  2017-12-19
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.