Radiation exposure with use of the mini-C-arm for routine orthopaedic imaging procedures.

Brian L Badman1, Lynn Rill, Bradley Butkovich, Manuel Arreola, Robert A Vander Griend.   

Abstract

BACKGROUND: The use of mobile fluoroscopic devices during orthopaedic procedures is associated with substantial concern with regard to the radiation exposure to surgeons and support staff. The perceived increased risks associated with large c-arm devices have been well documented. However, no study to date has documented the relative radiation risk associated with the use of a mini-c-arm device. The purpose of the current study was to determine the amount of radiation received by the surgeon during the use of a mini-c-arm device and to compare this amount with documented measurements associated with the large c-arm device.
METHODS: With use of a radiation dosimeter, measurements were carried out with tissue-equivalent anthropomorphic phantoms to quantitatively determine exposure rates at various locations and distances from the mini-c-arm for two common upper and lower extremity procedures.
RESULTS: Regardless of position, distance, or relative duration of exposure, exposure rates resulting from the use of the mini-c-arm device were one to two orders of magnitude lower than those reported in the literature in association with the use of the large c-arm device.
CONCLUSIONS: The mini-c-arm device should be utilized whenever feasible in order to eliminate many of the concerns associated with use of the large c-arm device, specifically those related to cumulative radiation hazards, positioning considerations, relative distance from the beam, and the need for protective shielding.

Mesh:

Year:  2005        PMID: 15634810     DOI: 10.2106/JBJS.D.02162

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  18 in total

1.  Occupational radiation exposure from C arm fluoroscopy during common orthopaedic surgical procedures and its prevention.

Authors:  Anupam Mahajan; Sumant Samuel; Atul K Saran; M K Mahajan; M K Mam
Journal:  J Clin Diagn Res       Date:  2015-03-01

2.  Closed reduction internal fixation rates and procedure times for metacarpal fractures treated in a minor surgery area before and after the introduction of a mini C-arm unit.

Authors:  R C Mahabir; C M DeCroff; L Thurgood; A R Harrop
Journal:  Can J Plast Surg       Date:  2008

3.  Mini C-Arm Fluoroscopy: Does Its Configuration Matter for Radiation Exposure to the Surgeon?

Authors:  Talia Chapman; Dennis P Martin; Christopher Williamson; Brian Tinsley; Mark L Wang; Asif M Ilyas
Journal:  Hand (N Y)       Date:  2017-06-23

4.  Hand Surgery and Fluoroscopic Eye Radiation Dosage: A Prospective Pilot Comparison of Large Versus Mini C-Arm Fluoroscopy Use.

Authors:  Mark L Wang; C Edward Hoffler; Asif M Ilyas; William H Kirkpatrick; Pedro K Beredjiklian; Charles F Leinberry
Journal:  Hand (N Y)       Date:  2016-04-06

Review 5.  C-arm fluoroscopy in orthopaedic surgical practice.

Authors:  Ishaq Ojodu; Ayodele Ogunsemoyin; Sascha Hopp; Tim Pohlemann; Oluwole Ige; Oluwaseun Akinola
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-05-23

6.  Radiation Exposure and Hand Dominance Using Mini C-Arm Fluoroscopy in Hand Surgery.

Authors:  Corey B Fuller; Montri D Wongworawat; Barth B Riedel
Journal:  Hand (N Y)       Date:  2016-02-02

7.  Elevated Radiation Exposure Associated With Above Surface Flat Detector Mini C-Arm Use.

Authors:  Dennis P Martin; Talia Chapman; Christopher Williamson; Brian Tinsley; Asif M Ilyas; Mark L Wang
Journal:  Hand (N Y)       Date:  2017-11-22

8.  Factors influencing radiation exposure during internal fixation of hip fractures.

Authors:  G A Kelly; F E Rowan; C Hurson
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-04-10

9.  Effect of introduction of mini C-arm image intensifier in orthopaedic theatre.

Authors:  S P White
Journal:  Ann R Coll Surg Engl       Date:  2007-04       Impact factor: 1.891

10.  A Radiation Safety Training Program Results in Reduced Radiation Exposure for Orthopaedic Residents Using the Mini C-arm.

Authors:  David Gendelberg; William Hennrikus; Jennifer Slough; Douglas Armstrong; Steven King
Journal:  Clin Orthop Relat Res       Date:  2015-11-13       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.