Exposure from the large C-arm versus the mini C-arm using hand/wrist and elbow phantoms.

Gordon Singer1, Brent Herron, David Herron.   

Abstract

PURPOSE: This study tests the conventional wisdom that using fluoroscopy under identical geometrical conditions results in less radiation when using the mini C-arm relative to the large C-arm.
METHODS: We evaluated the radiation dose for both direct exposure and scatter 2.54 cm outside the intensifier. We used 3 mini and 3 large C-arms in a vertical orientation with the image intensifier below the specimen and the source above. We used 2 specimens: a cadaver hand/wrist and a cadaver elbow. Specimens were tested both directly on the intensifier and on a hand table placed on the intensifier.
RESULTS: For the same setup, use of the mini C-arm resulted in direct patient radiation exposure greater than the exposure delivered by the large C-arm. Specifically, exposure using the mini C-arm was 53% to 70% greater than that using the large C-arm. In addition, use of the hand table resulted in exposure 80% to 94% greater compared with placing the specimen directly on the intensifier. In all cases, scatter at 2.54 cm from the intensifier resulted in an average exposure of 1.5% (SD, 0.24%) of the direct beam. Tube current, and therefore machine radiation output, was approximately 13 to 14 times greater for the large C-arm.
CONCLUSIONS: Direct radiation exposure to the patient and scatter to the surgeon are minimized when the C-arm is positioned with the intensifier below and the extremity is placed directly on the intensifier. Under identical geometrical conditions with the intensifier below the specimen, the large C-arm with its greater source to image intensifier distance is associated with less radiation exposure than the mini C-arm.
Copyright © 2011 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21414730     DOI: 10.1016/j.jhsa.2011.01.010

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  6 in total

1.  Mini C-Arm Fluoroscopy: Does Its Configuration Matter for Radiation Exposure to the Surgeon?

Authors:  Talia Chapman; Dennis P Martin; Christopher Williamson; Brian Tinsley; Mark L Wang; Asif M Ilyas
Journal:  Hand (N Y)       Date:  2017-06-23

2.  Hand Surgery and Fluoroscopic Eye Radiation Dosage: A Prospective Pilot Comparison of Large Versus Mini C-Arm Fluoroscopy Use.

Authors:  Mark L Wang; C Edward Hoffler; Asif M Ilyas; William H Kirkpatrick; Pedro K Beredjiklian; Charles F Leinberry
Journal:  Hand (N Y)       Date:  2016-04-06

3.  Radiation Exposure and Hand Dominance Using Mini C-Arm Fluoroscopy in Hand Surgery.

Authors:  Corey B Fuller; Montri D Wongworawat; Barth B Riedel
Journal:  Hand (N Y)       Date:  2016-02-02

4.  Elevated Radiation Exposure Associated With Above Surface Flat Detector Mini C-Arm Use.

Authors:  Dennis P Martin; Talia Chapman; Christopher Williamson; Brian Tinsley; Asif M Ilyas; Mark L Wang
Journal:  Hand (N Y)       Date:  2017-11-22

5.  Are Indian orthopaedic surgeons aware of the health hazards of radiation exposure? A survey and review on awareness and ways to mitigate them.

Authors:  Binoti Sheth; Arzan Jesia; Rohan Parihar; Rohit Somani; Pankaj Pawar
Journal:  J Clin Orthop Trauma       Date:  2022-08-07

6.  Confirming Proper Button Deployment of Suspensory Fixation During ACL Reconstruction.

Authors:  Daniel F O'Brien; Lilah Fones; Victoria Stoj; Cory Edgar; Katherine Coyner; Robert A Arciero
Journal:  Orthop J Sports Med       Date:  2021-01-19
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.