Motion-robust cardiac B1+ mapping at 3T using interleaved bloch-siegert shifts.

Sebastian Weingärtner1,2,3, Fabian Zimmer3, Gregory J Metzger2, Kâmil Uğurbil2, Pierre-Francois Van de Moortele2, Mehmet Akçakaya1,2.   

Abstract

PURPOSE: To develop and evaluate a robust motion-insensitive Bloch-Siegert shift based B1+ mapping method in the heart.
METHODS: Cardiac Bloch-Siegert B1+ mapping was performed with interleaved positive and negative off-resonance shifts and diastolic spoiled gradient echo imaging in 12 heartbeats. Numerical simulations were performed to study the impact of respiratory motion. The method was compared with three-dimensional (3D) actual flip angle imaging (AFI) and two-dimensional (2D) saturated double angle method (SDAM) in phantom scans. Cardiac B1+ maps of three different views were acquired in six healthy volunteers using Bloch-Siegert and SDAM during breath-hold and free breathing. In vivo maps were evaluated for inter-view consistency using the correlation coefficients of the B1+ profiles along the lines of intersection between the views.
RESULTS: For the Bloch-Siegert sequence, numerical simulations indicated high similarity between breath-hold and free breathing scans, and phantom results indicated low deviation from the 3D AFI reference (normalized root mean square error [NRMSE] = 2.0%). Increased deviation was observed with 2D SDAM (NRMSE = 5.0%) due to underestimation caused by imperfect excitation slice profiles. Breath-hold and free breathing Bloch-Siegert in vivo B1+ maps were visually comparable with no significant difference in the inter-view consistency (P > 0.36). SDAM showed strongly impaired B1+ map quality during free breathing. Inter-view consistency was significantly lower than with the Bloch-Siegert method (breath-hold: P = 0.014, free breathing: P < 0.0001).
CONCLUSION: The proposed interleaved Bloch-Siegert sequence enables cardiac B1+ mapping with improved inter-view consistency and high resilience to respiratory motion. Magn Reson Med 78:670-677, 2017.
© 2016 International Society for Magnetic Resonance in Medicine. © 2016 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  zzm321990B1+ mapping; Bloch-Siegert shift; motion robustness; abdominal imaging; cardiac imaging

Mesh:

Year:  2016        PMID: 27599782      PMCID: PMC5340643          DOI: 10.1002/mrm.26395

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  68 in total

1.  Cardiac chamber quantification using magnetic resonance imaging at 7 Tesla--a pilot study.

Authors:  Florian von Knobelsdorff-Brenkenhoff; Tobias Frauenrath; Marcel Prothmann; Matthias A Dieringer; Fabian Hezel; Wolfgang Renz; Kerstin Kretschel; Thoralf Niendorf; Jeanette Schulz-Menger
Journal:  Eur Radiol       Date:  2010-07-17       Impact factor: 5.315

Review 2.  T1 Mapping: Basic Techniques and Clinical Applications.

Authors:  Andrew J Taylor; Michael Salerno; Rohan Dharmakumar; Michael Jerosch-Herold
Journal:  JACC Cardiovasc Imaging       Date:  2016-01

Review 3.  Spiral imaging: a critical appraisal.

Authors:  Kai Tobias Block; Jens Frahm
Journal:  J Magn Reson Imaging       Date:  2005-06       Impact factor: 4.813

4.  Enhancing the quantification of tissue sodium content by MRI: time-efficient sodium B1 mapping at clinical field strengths.

Authors:  Jonathan Lommen; Simon Konstandin; Philipp Krämer; Lothar R Schad
Journal:  NMR Biomed       Date:  2015-04-21       Impact factor: 4.044

5.  Transmit B1+ field inhomogeneity and T1 estimation errors in breast DCE-MRI at 3 tesla.

Authors:  Kyunghyun Sung; Bruce L Daniel; Brian A Hargreaves
Journal:  J Magn Reson Imaging       Date:  2013-01-04       Impact factor: 4.813

6.  Three-dimensional T1 mapping of the mouse heart using variable flip angle steady-state MR imaging.

Authors:  Bram F Coolen; Tessa Geelen; Leonie E M Paulis; Arno Nauerth; Klaas Nicolay; Gustav J Strijkers
Journal:  NMR Biomed       Date:  2010-10-19       Impact factor: 4.044

7.  Adiabatic RF pulse design for Bloch-Siegert B1+ mapping.

Authors:  Mohammad Mehdi Khalighi; Brian K Rutt; Adam B Kerr
Journal:  Magn Reson Med       Date:  2012-10-05       Impact factor: 4.668

8.  Simultaneous B(0)- and B(1)+-map acquisition for fast localized shim, frequency, and RF power determination in the heart at 3 T.

Authors:  Michael Schär; Evert-Jan Vonken; Matthias Stuber
Journal:  Magn Reson Med       Date:  2010-02       Impact factor: 4.668

Review 9.  Advances in parametric mapping with CMR imaging.

Authors:  Michael Salerno; Christopher M Kramer
Journal:  JACC Cardiovasc Imaging       Date:  2013-07

10.  A simple correction for B1 field errors in magnetization transfer ratio measurements.

Authors:  Rebecca S Samson; Claudia A M Wheeler-Kingshott; Mark R Symms; Daniel J Tozer; Paul S Tofts
Journal:  Magn Reson Imaging       Date:  2005-12-27       Impact factor: 2.546

View more
  5 in total

1.  Whole-heart, ungated, free-breathing, cardiac-phase-resolved myocardial perfusion MRI by using Continuous Radial Interleaved simultaneous Multi-slice acquisitions at sPoiled steady-state (CRIMP).

Authors:  Ye Tian; Jason Mendes; Brent Wilson; Alexander Ross; Ravi Ranjan; Edward DiBella; Ganesh Adluru
Journal:  Magn Reson Med       Date:  2020-06-03       Impact factor: 4.668

2.  Temporally resolved parametric assessment of Z-magnetization recovery (TOPAZ): Dynamic myocardial T1 mapping using a cine steady-state look-locker approach.

Authors:  Sebastian Weingärtner; Chetan Shenoy; Benedikt Rieger; Lothar R Schad; Jeanette Schulz-Menger; Mehmet Akçakaya
Journal:  Magn Reson Med       Date:  2017-08-30       Impact factor: 4.668

3.  Free-breathing abdominal T1 mapping using an optimized MR fingerprinting sequence.

Authors:  Max H C van Riel; Zidan Yu; Shota Hodono; Ding Xia; Hersh Chandarana; Koji Fujimoto; Martijn A Cloos
Journal:  NMR Biomed       Date:  2021-04-26       Impact factor: 4.044

4.  Ultrafast 3D Bloch-Siegert B <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:msubsup><mml:mrow/> <mml:mn>1</mml:mn> <mml:mo>+</mml:mo></mml:msubsup> </mml:math> -mapping using variational modeling.

Authors:  Andreas Lesch; Matthias Schlöegl; Martin Holler; Kristian Bredies; Rudolf Stollberger
Journal:  Magn Reson Med       Date:  2018-10-12       Impact factor: 4.668

5.  Robust 3D Bloch-Siegert based B 1 + mapping using multi-echo general linear modeling.

Authors:  Nadège Corbin; Julio Acosta-Cabronero; Shaihan J Malik; Martina F Callaghan
Journal:  Magn Reson Med       Date:  2019-07-18       Impact factor: 4.668

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.