Mohammad Mehdi Khalighi1, Brian K Rutt, Adam B Kerr.
Abstract
The Bloch-Siegert (B-S) B1+ mapping method has been shown to be fast and accurate, yet it suffers from high Specific Absorption Rate (SAR) and moderately long echo time. An adiabatic RF pulse design is introduced here for optimizing the off-resonant B-S RF pulse to achieve more B-S B1+ measurement sensitivity for a given pulse width. The extra sensitivity can be used for higher angle-to-noise ratio B1+ maps or traded off for faster scans. Using numerical simulations and phantom experiments, it is shown that a numerically optimized 2-ms adiabatic B-S pulse is 2.5 times more efficient than a conventional 6-ms Fermi-shaped B-S pulse. The adiabatic B-S pulse performance is validated in a phantom, and in vivo brain B1+ mapping at 3T and 7T are shown.Entities:
Keywords: Bloch‐Siegert shift; adiabatic RF pulse design; flip angle; off resonance; parallel transmit
Mesh:
Year: 2012 PMID: 23041985 PMCID: PMC3796009 DOI: 10.1002/mrm.24507
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668