| Literature DB >> 27403339 |
Jinhai Huang1, Weicong Lu1, Giacomo Savini2, Hao Chen3, Chengfang Wang3, Xinxin Yu3, Fangjun Bao1, Qinmei Wang1.
Abstract
Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59 ± 27.37 μm, 3.70 ± 0.30 mm, and 3.16 ± 0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14 ± 26.61 μm for CCT, 3.71 ± 0.29 mm for ACD, and 3.17 ± 0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45 ± 6.07 μm, P < 0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between -9.44 and 14.35 μm, -0.15 and 0.12 mm, and -0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.Entities:
Year: 2016 PMID: 27403339 PMCID: PMC4923600 DOI: 10.1155/2016/6347236
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Comparison of central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD) measured by the AL-Scan partial coherence interferometry and Visante optical coherence tomography.
| Device pairings | Mean difference ± SD |
| 95% LoA |
|---|---|---|---|
| CCT ( | 2.45 ± 6.07 | 0.002 | −9.44 to 14.35 |
| ACD (mm) | −0.01 ± 0.07 | 0.119 | −0.15 to 0.12 |
| AD (mm) | −0.02 ± 0.07 | 0.077 | −0.15 to 0.12 |
SD: standard deviation.
Figure 1Difference in central corneal thickness measurements between AL-Scan optical biometer and Pentacam rotating Scheimpflug imaging device against their mean values. The solid line indicates the mean difference, and 95% limits of agreement are indicated by solid and dotted lines, respectively.
Figure 2Difference in anterior chamber depth measurements between AL-Scan optical biometer and Pentacam rotating Scheimpflug imaging device against their mean values. The solid line indicates the mean difference, and 95% limits of agreement are indicated by solid and dotted lines, respectively.
Figure 3Difference in aqueous depth measurements between AL-Scan optical biometer and Pentacam rotating Scheimpflug imaging device against their mean values. The solid line indicates the mean difference, and 95% limits of agreement are indicated by solid and dotted lines, respectively.