| Literature DB >> 26913857 |
Marrigje F Meijer1, Alexander L Boerboom2, Sjoerd K Bulstra2, Inge H F Reininga3, Martin Stevens2.
Abstract
PURPOSE: Objective of this study was to compare intraoperative computer-assisted surgery (CAS) alignment measurements during total knee arthroplasty (TKA) with pre- and postoperative coronal alignment measurements using EOS 3D reconstructions.Entities:
Keywords: CAS; Computer navigation; Computer-assisted surgery; Coronal alignment; Long-leg radiograph; TKA; TKR; Total knee arthroplasty; Total knee replacement
Mesh:
Year: 2016 PMID: 26913857 PMCID: PMC5570772 DOI: 10.1007/s00167-016-4031-3
Source DB: PubMed Journal: Knee Surg Sports Traumatol Arthrosc ISSN: 0942-2056 Impact factor: 4.342
Comparison of CAS and EOS measurements
| Mean | SD | Mean difference (95 % CI) | SDΔ | Range of difference CAS-EOS |
| Spearman’s |
| |
|---|---|---|---|---|---|---|---|---|
|
| ||||||||
| VVCAS | 0 | 8.3 | ||||||
| VV3D | −3 | 10.3 | 3 (1.5– 4.6) | 5.4 | −7 to 24 | ≤0.001* | 0.87 | 0.34 |
| mLDFA CAS | 2 | 3.9 | ||||||
| mLDFA EOS | 1 | 2.8 | 1 (−0.2–1.3) | 2.6 | −7 to 6 | 0.12 | 0.76 | 0.58 |
| mMPTA CAS | −2 | 6.3 | ||||||
| mMPTA EOS | −4 | 5.5 | 2 (0.4–3.3) | 4.8 | −9 to 15 | 0.01* | 0.67 | 0.44 |
|
| ||||||||
| VVCAS | 0 | 3.7 | ||||||
| VV3D | −2 | 3.3 | 2 (1.2–3.3) | 3.6 | −2 to 21 | ≤0.001* | 0.68 | 0.19 |
For calculating the mean difference, the angle measured by the EOS system was subtracted from the CAS angle
SD = standard deviation; 95 % CI = 95 % confidence interval; CAS = computer-assisted surgery; VVCAS = varus/valgus angle measured using CAS; VV3D = varus/valgus angle measured in 3D using EOS; mLDFA = mechanical lateral distal femoral angle; mMPTA = mechanical medial proximal tibial angle
* Statistical significance (P < 0.05)
Fig. 1Bland & Altman plot of the primary CAS measurement and preoperative EOS measurement of the varus/valgus angle, showing heteroscedasticity
Fig. 2For varus legs, EOS measures more varus, and for valgus legs it measures more valgus than CAS
Fig. 3Bland & Altman plot of the primary CAS measurement and preoperative EOS measurement of the mechanical lateral distal femoral angle, showing no systematic bias
Fig. 4Bland & Altman plot of the primary CAS measurement and preoperative EOS measurement of the mechanical medial proximal tibial angle, showing a systematic bias
Fig. 5Bland & Altman plot of the second CAS measurement and postoperative EOS measurement of the varus/valgus angle, showing a systematic bias