Predicting long-term wear performance of hard-on-hard bearing couples: effect of cup orientation.

S Shankar1, K Gowthaman2, M S Uddin3.   

Abstract

Wear is the major cause of implant failure, resulting in expensive revision surgeries of total hip arthroplasty. Therefore, understanding of wear mechanism and its progression is crucial to improve the physiological performance of implants. This paper presents a three-dimensional (3D) finite element (FE) wear modeling approach to estimate evolution of wear in hard-on-hard bearing components with the effect of cup abduction angle. Three bearing couples were considered, and they were PCD-on-PCD, Al2O3-on-Al2O3 and Si3N4-on-Si3N4, while the cup abduction angle varied from 35° to 70° with an increment of 5°. By adopting actual physiological hip gait loading and rotational movement for normal walking cycle in FE modeling, the contact pressure and the sliding distance were calculated to predict wear. A femoral head of 32 mm in diameter was considered, while a constant frictional contact at the inference between head and cup was used. During simulation, the geometry of cup surface was updated at a reasonable interval of gait cycles to consider the effect of wear. Wear was simulated for up to 20 million cycles which is an equivalent of 20 years of implant's life in human body. Simulation results showed that compared to other two bearing couples, the predicted linear and volumetric wear in PCD-on-PCD couple exhibited the least wear evolution for all cup angles considered. The increase in abduction angle from 35° to 70° decreases the volumetric wear by 28 % for all three bearing couples, due to the reduction in sliding distance. Steep cup angle, e.g., 70° for Al2O3 and Si3N4 bearing couples, encountered edge contact, which leads to more wear. Further, wear results were discussed and analyzed with respect to in vitro and/or clinical studies available in the literature to justify the efficacy of wear modeling.

Entities:  

Keywords:  Abduction angle; Contact pressure; Finite element method; Hip joint prostheses; Long-term wear

Mesh:

Year:  2015        PMID: 26718551     DOI: 10.1007/s11517-015-1440-2

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  25 in total

Review 1.  Biological reactions to wear debris in total joint replacement.

Authors:  E Ingham; J Fisher
Journal:  Proc Inst Mech Eng H       Date:  2000       Impact factor: 1.617

2.  Alumina-alumina artificial hip joints. Part II: characterisation of the wear debris from in vitro hip joint simulations.

Authors:  J L Tipper; A Hatton; J E Nevelos; E Ingham; C Doyle; R Streicher; A B Nevelos; J Fisher
Journal:  Biomaterials       Date:  2002-08       Impact factor: 12.479

3.  The influence of acetabular cup angle on the wear of "BIOLOX Forte" alumina ceramic bearing couples in a hip joint simulator.

Authors:  J E Nevelos; E Ingham; C Doyle; A B Nevelos; J Fisher
Journal:  J Mater Sci Mater Med       Date:  2001-02       Impact factor: 3.896

4.  The influence of the pelvic bone on the computational results of the acetabular component of a total hip prosthesis.

Authors:  Sara Barreto; João Folgado; Paulo R Fernandes; Jacinto Monteiro
Journal:  J Biomech Eng       Date:  2010-05       Impact factor: 2.097

5.  Results of cemented metal-backed acetabular components: a 10-year-average follow-up study.

Authors:  F S Chen; P E Di Cesare; A A Kale; J F Lee; V H Frankel; S A Stuchin; J D Zuckerman
Journal:  J Arthroplasty       Date:  1998-12       Impact factor: 4.757

6.  Do large femoral heads reduce the risks of impingement in total hip arthroplasty with optimal and non-optimal cup positioning?

Authors:  Gianluca Cinotti; Niccolò Lucioli; Andrea Malagoli; Carlo Calderoli; Ferdinando Cassese
Journal:  Int Orthop       Date:  2010-02-17       Impact factor: 3.075

7.  Contact of dual mobility implants: effects of cup wear and inclination.

Authors:  M S Uddin
Journal:  Comput Methods Biomech Biomed Engin       Date:  2014-07-14       Impact factor: 1.763

8.  Hip contact forces and gait patterns from routine activities.

Authors:  G Bergmann; G Deuretzbacher; M Heller; F Graichen; A Rohlmann; J Strauss; G N Duda
Journal:  J Biomech       Date:  2001-07       Impact factor: 2.712

9.  Polyethylene wear and acetabular component orientation.

Authors:  Shantanu Patil; Arnie Bergula; Peter C Chen; Clifford W Colwell; Darryl D D'Lima
Journal:  J Bone Joint Surg Am       Date:  2003       Impact factor: 5.284

10.  Fabrication and testing of silicon nitride bearings in total hip arthroplasty: winner of the 2007 "HAP" PAUL Award.

Authors:  B Sonny Bal; Ashok Khandkar; R Lakshminarayanan; Ian Clarke; Aaron A Hoffman; Mohamed N Rahaman
Journal:  J Arthroplasty       Date:  2008-04-28       Impact factor: 4.757

View more
  1 in total

1.  Computational wear assessment of hard on hard hip implants subject to physically demanding tasks.

Authors:  R Nithyaprakash; S Shankar; M S Uddin
Journal:  Med Biol Eng Comput       Date:  2017-11-02       Impact factor: 2.602