The influence of acetabular cup angle on the wear of "BIOLOX Forte" alumina ceramic bearing couples in a hip joint simulator.

J E Nevelos1, E Ingham, C Doyle, A B Nevelos, J Fisher.   

Abstract

The wear of "Biolox Forte" alumina ceramic bearing couples has been investigated at two different acetabular cup angles in a physiological hip joint simulator. All cups were set in the anatomical position of 45 degrees inclination in the M/L plane for the first two million cycles and then four of the six cups were re-aligned to 60 degrees for a further three million cycles. A "running-in" wear of 0.14 mm3 per million cycles was observed for the first million cycles, after which a steady state wear rate of 0.05 mm3 per million cycles was observed. Increasing the acetabular cup angle to 60 degrees did not significantly affect the wear rate.

Entities:  

Year:  2001        PMID: 15348320     DOI: 10.1023/a:1008970027306

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  4 in total

1.  Sixteen-years' experience with ceramic hip prostheses.

Authors:  H Mittelmeier; J Heisel
Journal:  Clin Orthop Relat Res       Date:  1992-09       Impact factor: 4.176

2.  Long-term effects of alumina components in total hip prostheses.

Authors:  J M Dorlot
Journal:  Clin Orthop Relat Res       Date:  1992-09       Impact factor: 4.176

3.  [Total arthroplasty of the hip by fritted aluminum prosthesis. Experimental study and 1st clinical applications].

Authors:  P Boutin
Journal:  Rev Chir Orthop Reparatrice Appar Mot       Date:  1972 Apr-May

4.  Examination of alumina ceramic components from total hip arthroplasties.

Authors:  A B Nevelös; P A Evans; P Harrison; M Rainforth
Journal:  Proc Inst Mech Eng H       Date:  1993       Impact factor: 1.617

  4 in total
  17 in total

1.  Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions.

Authors:  T Stewart; J Tipper; R Streicher; E Ingham; J Fisher
Journal:  J Mater Sci Mater Med       Date:  2001 Oct-Dec       Impact factor: 3.896

2.  Differences between the wear couples metal-on-polyethylene and ceramic-on-ceramic in the stability against dislocation of total hip replacement.

Authors:  Rainer Bader; Erwin Steinhauser; Susanne Zimmermann; Wolfram Mittelmeier; Roger Scholz; Raymonde Busch
Journal:  J Mater Sci Mater Med       Date:  2004-06       Impact factor: 3.896

3.  Biotribological study of large diameter ceramic-on-CFR-PEEK hip joint including fluid uptake, wear and frictional heating.

Authors:  Qian Qian Wang; Jun Jie Wu; Anthony Unsworth; Adam Briscoe; Marcus Jarman-Smith; Conor Lowry; David Simpson; Simon Collins
Journal:  J Mater Sci Mater Med       Date:  2012-03-28       Impact factor: 3.896

4.  Predicting long-term wear performance of hard-on-hard bearing couples: effect of cup orientation.

Authors:  S Shankar; K Gowthaman; M S Uddin
Journal:  Med Biol Eng Comput       Date:  2015-12-30       Impact factor: 2.602

5.  Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty.

Authors:  Jiri Gallo; Vitezslav Havranek; Jana Zapletalova
Journal:  Int Orthop       Date:  2009-02-13       Impact factor: 3.075

6.  High cup angle and microseparation increase the wear of hip surface replacements.

Authors:  Ian J Leslie; Sophie Williams; Graham Isaac; Eileen Ingham; John Fisher
Journal:  Clin Orthop Relat Res       Date:  2009-04-11       Impact factor: 4.176

7.  Does acetabular inclination angle affect survivorship of alumina-ceramic articulations?

Authors:  Henning R Johansson; Aaron J Johnson; Michael G Zywiel; Marybeth Naughton; Michael A Mont; Peter M Bonutti
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

8.  A rolling-gliding wear simulator for the investigation of tribological material pairings for application in total knee arthroplasty.

Authors:  Berna I Richter; Sven Ostermeier; Anke Turger; Berend Denkena; Christof Hurschler
Journal:  Biomed Eng Online       Date:  2010-06-15       Impact factor: 2.819

9.  The John Charnley Award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital.

Authors:  Mark C Callanan; Bryan Jarrett; Charles R Bragdon; David Zurakowski; Harry E Rubash; Andrew A Freiberg; Henrik Malchau
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

Review 10.  Review on squeaking hips.

Authors:  Yadin David Levy; Selin Munir; Shane Donohoo; William Lindsay Walter
Journal:  World J Orthop       Date:  2015-11-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.