Alumina-alumina artificial hip joints. Part II: characterisation of the wear debris from in vitro hip joint simulations.

J L Tipper1, A Hatton, J E Nevelos, E Ingham, C Doyle, R Streicher, A B Nevelos, J Fisher.   

Abstract

Until recently it was not possible to reproduce clinically relevant wear rates and wear patterns in in vitro hip joint simulators for alumina ceramic-on-ceramic hip prostheses. The introduction of microseparation of the prosthesis components into in vitro wear simulations produced clinically relevant wear rates and wear patterns for the first time. The aim of this study was to characterise the wear particles generated from standard simulator testing and microseparation simulator testing of hot isostatically pressed (HIPed) and non-HIPed alumina ceramic-on-ceramic hip prostheses, and compare these particles to those generated in vivo. Standard simulation conditions produced wear rates of approximately 0.1 mm3 per million cycles for both material types. No change in surface roughness was detected and very few wear features were observed. In contrast, when microseparation was introduced into the wear simulation, wear rates of between 1.24 (HIPed) and 1.74 mm3 per million cycles (non-HIPed) were produced. Surface roughness increased and a wear stripe often observed clinically on retrieved femoral heads was also reproduced. Under standard simulation conditions only nanometre-sized wear particles (2-27.5 nm) were observed by TEM, and it was thought likely that these particles resulted from relief polishing of the alumina ceramic. However, when microseparation of the prosthesis components was introduced into the simulation, a bi-modal distribution of particle sizes was observed. The nanometre-sized particles produced by relief polishing were present (1-35nm). however, larger micrometre-sized particles were also observed by both transmission electron microscopy (TEM) (0.021 microm) and scanning electron microscopy (SEM) (0.05-->10 microm). These larger particles were thought to originate from the wear stripe and were produced by trans-granular fracture of the alumina ceramic. In Part I of this study, alumina ceramic wear particles were isolated from the periprosthetic tissues from around Mittelmeier ceramic-on-ceramic hip prostheses. Characterisation of the particles by TEM and SEM revealed a bi-modal size distribution. SEM analysis revealed particles in the 0.05-3.2 microm size range. and TEM revealed particles in the 5-90 nm size range, indicating that microseparation of the prosthesis components may be a common event in vivo. This study (Part II) has revealed that the introduction of microseparation of the prosthesis components during the swing phase of the wear simulation reproduced clinically relevant wear rates, wear patterns and wear particles in in vitro hip joint simulators.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12099287     DOI: 10.1016/s0142-9612(02)00048-0

Source DB:  PubMed          Journal:  Biomaterials        ISSN: 0142-9612            Impact factor:   12.479


  17 in total

1.  Assessment of wear on the cones of modular stainless steel Exeter hip stems.

Authors:  R P S Chaplin; A J C Lee; R M Hooper
Journal:  J Mater Sci Mater Med       Date:  2004-09       Impact factor: 3.896

2.  Predicting long-term wear performance of hard-on-hard bearing couples: effect of cup orientation.

Authors:  S Shankar; K Gowthaman; M S Uddin
Journal:  Med Biol Eng Comput       Date:  2015-12-30       Impact factor: 2.602

Review 3.  Human health risk assessment for aluminium, aluminium oxide, and aluminium hydroxide.

Authors:  Daniel Krewski; Robert A Yokel; Evert Nieboer; David Borchelt; Joshua Cohen; Jean Harry; Sam Kacew; Joan Lindsay; Amal M Mahfouz; Virginie Rondeau
Journal:  J Toxicol Environ Health B Crit Rev       Date:  2007       Impact factor: 6.393

4.  High cup angle and microseparation increase the wear of hip surface replacements.

Authors:  Ian J Leslie; Sophie Williams; Graham Isaac; Eileen Ingham; John Fisher
Journal:  Clin Orthop Relat Res       Date:  2009-04-11       Impact factor: 4.176

Review 5.  Mediators of the inflammatory response to joint replacement devices.

Authors:  Neil Cobelli; Brian Scharf; Giovanna M Crisi; John Hardin; Laura Santambrogio
Journal:  Nat Rev Rheumatol       Date:  2011-09-06       Impact factor: 20.543

6.  Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulation.

Authors:  Christopher Brown; Sophie Williams; Joanne L Tipper; John Fisher; Eileen Ingham
Journal:  J Mater Sci Mater Med       Date:  2006-12-14       Impact factor: 4.727

7.  A review of the biologic effects of spine implant debris: Fact from fiction.

Authors:  Nadim James Hallab
Journal:  SAS J       Date:  2009-12-01

8.  In vitro assessment of serum-saline ratios for fluid simulator testing of highly modular spinal implants with articulating surfaces.

Authors:  Nadim Hallab; Ashutosh Khandha; George Malcolmson; J P Timm
Journal:  SAS J       Date:  2008-12-01

9.  Effect of femoral head size on the wear of metal on metal bearings in total hip replacements under adverse edge-loading conditions.

Authors:  Mazen Al-Hajjar; John Fisher; Sophie Williams; Joanne L Tipper; Louise M Jennings
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2012-12-20       Impact factor: 3.368

10.  Manufacturing conditioned roughness and wear of biomedical oxide ceramics for all-ceramic knee implants.

Authors:  Anke Turger; Jens Köhler; Berend Denkena; Tomas A Correa; Christoph Becher; Christof Hurschler
Journal:  Biomed Eng Online       Date:  2013-08-29       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.