Results of cemented metal-backed acetabular components: a 10-year-average follow-up study.

F S Chen1, P E Di Cesare, A A Kale, J F Lee, V H Frankel, S A Stuchin, J D Zuckerman.   

Abstract

The clinical and radiographic results of 86 primary total hip arthroplasties performed in 74 patients from 1983 to 1987 with a cemented metal-backed acetabular component and a cemented collared straight femoral stem with a 32-mm head were reviewed at a mean follow-up of 10.1 years. Seven patients (9.2%) underwent acetabular component revision at a mean of 9.0 years after implantation; an additional 24 components (31.6%) demonstrated evidence of radiographic loosening, resulting in a total failure rate of 40.8%. Periacetabular radiolucencies were noted in Charnley zones at the following rates: 34.2% in zone I, 18.4% in zone II, and 27.6% in zone III. In addition, 18.4% and 38.2% of implants demonstrated evidence of migration and excessive polyethylene wear. Excessively vertical cup placement (>49 degrees inclination) at the time of initial arthroplasty was statistically correlated with polyethylene wear, implant migration, and fixation failure. A trend of increasing implant failure was also noted with decreasing polyethylene liner thickness. Periacetabular cement mantle thickness was not statistically correlated with subsequent component loosening or failure. Results of Kaplan-Meier survivorship analysis using revision as an endpoint showed 93.6% survivorship at 10 years and 88.4% at 12 years. The mean modified Harris hip scores were 46.9 preoperatively and 81.8 at final follow-up. The significant overall rates of radiographic loosening, migration, polyethylene wear, and implant revision confirm the suspected trend of increasing failure rates of cemented metal-backed acetabular components over time.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9880177     DOI: 10.1016/s0883-5403(98)90191-7

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  4 in total

1.  Predicting long-term wear performance of hard-on-hard bearing couples: effect of cup orientation.

Authors:  S Shankar; K Gowthaman; M S Uddin
Journal:  Med Biol Eng Comput       Date:  2015-12-30       Impact factor: 2.602

2.  In vitro evaluation of stiffness graded artificial hip joint femur head in terms of joint stresses distributions and dimensions: finite element study.

Authors:  H Fouad
Journal:  J Mater Sci Mater Med       Date:  2011-04-20       Impact factor: 3.896

3.  The benefit of the systematic revision of the acetabular implant in favor of a dual mobility articulation during the treatment of periprosthetic fractures of the femur: a 49 cases prospective comparative study.

Authors:  A Perrin; M Saab; S Putman; K Benad; E Drumez; C Chantelot
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-09-12

Review 4.  Early subsidence of shape-closed hip arthroplasty stems is associated with late revision. A systematic review and meta-analysis of 24 RSA studies and 56 survival studies.

Authors:  Paul van der Voort; Bart G Pijls; Marc J Nieuwenhuijse; Jorrit Jasper; Marta Fiocco; Josepha W M Plevier; Saskia Middeldorp; Edward R Valstar; Rob G H H Nelissen
Journal:  Acta Orthop       Date:  2015       Impact factor: 3.717

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.