Contact of dual mobility implants: effects of cup wear and inclination.

M S Uddin1.   

Abstract

Cup wear and inclination on the pelvic bone are significant factors, which change the contact of the articulating surfaces, thus, impacting the long-term performance of hip implants. This paper presents a finite element (FE) analysis of the contact of the dual mobility implants under the influence of cup wear and inclination. A 3D FE model of the implant was developed with the application of equivalent physiological loading and boundary conditions. Effects of cup inclination angle ranging from 45° to 60° and the wear depth ranging from 0 to 2.46 mm equivalent to up to 30 years of the implant's life on the contact pressure and von Mises stress were investigated. Simulation results show that the contact pressure and von Mises stress decrease significantly with a modest wear depth and remains quite in-sensitive to the cup inclination angle and wear depth up to 1.64 mm. With wear depth further up to 2.46 mm, the cup thickness (i.e. cup thinning on worn region) may be more predominant than increasing of contact area between the cup and the head. The wear on the inner surface of the cup is found to rule out the overall contact pressure and stress in the implant. Furthermore, individual and combined effects of both important parameters are analysed and discussed with respect to available clinical/laboratory studies.

Entities:  

Keywords:  contact pressure; cup inclination angle; dual mobility hip implants; von Mises stress; wear

Mesh:

Year:  2014        PMID: 25023708     DOI: 10.1080/10255842.2014.936856

Source DB:  PubMed          Journal:  Comput Methods Biomech Biomed Engin        ISSN: 1025-5842            Impact factor:   1.763


  3 in total

1.  Predicting long-term wear performance of hard-on-hard bearing couples: effect of cup orientation.

Authors:  S Shankar; K Gowthaman; M S Uddin
Journal:  Med Biol Eng Comput       Date:  2015-12-30       Impact factor: 2.602

2.  CORR Insights®: Spinal Fusion Is Associated With Changes in Acetabular Orientation and Reductions in Pelvic Mobility.

Authors:  Edward Ebramzadeh
Journal:  Clin Orthop Relat Res       Date:  2019-02       Impact factor: 4.176

3.  Computational wear assessment of hard on hard hip implants subject to physically demanding tasks.

Authors:  R Nithyaprakash; S Shankar; M S Uddin
Journal:  Med Biol Eng Comput       Date:  2017-11-02       Impact factor: 2.602

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.