| Literature DB >> 26289071 |
Arturo Navarro-Martin1, Jon Cacicedo2, Olwen Leaman3, Ismael Sancho4, Elvira García5, Valentin Navarro6, Ferran Guedea7.
Abstract
BACKGROUND: To compare thermoplastic masks (TMP) and vacuum cushion system (VCS) to assess differences in interfraction set up accuracy in patients treated with stereotactic radiotherapy (SBRT) for oligometastatic lung cancer. Secondarily, to survey radiotherapy technologists to assess their satisfaction with the two systems.Entities:
Mesh:
Year: 2015 PMID: 26289071 PMCID: PMC4554348 DOI: 10.1186/s13014-015-0484-7
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Patients Characteristics and treatments
| Variable | |||||
|---|---|---|---|---|---|
| Age | Mean, 68 yrs (r, 39-89) | n = 73 | % | ||
| Target Location | Right Upper Lobe | 28 | 38.35 | ||
| Left Upper Lobe | 22 | 30.13 | |||
| Right Lower Lobe | 10 | 13.7 | |||
| Left Lower Lobe | 8 | 11 | |||
| T12 vertebrae | 1 | 1.37 | |||
| Median Lobe | 3 | 4.10 | |||
| Sacrum | 1 | 1.37 | |||
| Primary | Lung | 62 | 85 | ||
| Colorectal | 8 | 11 | |||
| Breast | 2 | 2.7 | |||
| Neuroendocrine | 1 | 1.37 | |||
| Treatment | Dose | TMP | VCS | % | |
| 34Gy x 1fr | 9 | 0 | 12.3 | 0 | |
| 18Gy x 3fr | 9 | 35 | 12.3 | 48 | |
| 12.5Gy x 4fr | 9 | 3 | 12.3 | 4 | |
| 16Gy x 1fr | 2 | 0 | 2.7 | 0 | |
| 7.5Gy x 8fr | 4 | 1 | 5.5 | 1.3 | |
| 10Gy x 5fr | 0 | 1 | 0 | 1.3 | |
| Total | 33 | 40 | 45 | 54 | |
TMP indicates thermoplastic masks; VCS, vacuum cushion system; fr, fraction
Fig. 1Thermoplastic mask and vacuum cushion system components. Fig. 1a depicts the vacuum cushion system with its components: a: Vacuum Cushion, b: Common Couch, c: Abdominal Compressor. d: Marker Block, e: Arm Support. Fig 1B shows the thermoplastic mask system and its components: a: Thermoplastic Mask, b: Common Couch, c: Abdominal Compressor. d: Marker Block e: Arm support
Fig. 2Plane-height graphics
Results of survey of technologists. Scores range from 1 (lowest) to 4 (highest)
| TMP | VC | |
|---|---|---|
| Indexing | Mean Score | |
| Simple | 2.86 | 1.61 |
| Fast | 2.43 | 1.61 |
| Manageable | 2.71 | 1.8 |
| Mask Utility | 3.86 | NA |
| Additional Vacuum Cushion | 1.43 | NA |
| Arc Adjustment | 3.14 | 2 |
| Versatility | 2.14 | 1.4 |
| Positioning | ||
| Simple | 3.43 | 1.8 |
| Fast | 2.71 | 1.4 |
| Reproducible | 3.71 | 1 |
| Comfortable | 2.43 | 2.4 |
| Learning Curve | ||
| Simple | 3.71 | 2.4 |
| Fast | 3.57 | 2.6 |
Abbreviation: TMP indicates thermoplastic ask; VCS vacuum cushion system; NA not applicable
Fig. 3Displacements in Module
Mean Displacements and Variance
| MEAN | T- test Welch | Variance ANOVA test | ||||
|---|---|---|---|---|---|---|
| TMP | VCS | Value | p-value | ratio (TMP/VCS) | p-value | |
| Session 1 | 0.64 | 1.05 | −3.91 | 0.0002 | 0.4007 | 0.0081 |
| Session 2 | 0.49 | 1.02 | −4.49 | <0.0001 | 0.5921 | 0.1508 |
| Session 3 | 0.56 | 0.97 | −3.41 | 0.0011 | 0.3631 | 0.0075 |
Abbreviation: TMP Thermoplastic Mask, VCS Vacuum cushion system