Evaluation of the effectiveness of the stereotactic body frame in reducing respiratory intrafractional organ motion using the real-time tumor-tracking radiotherapy system.

Gerard Bengua1, Masayori Ishikawa, Kenneth Sutherland, Kenji Horita, Rie Yamazaki, Katsuhisa Fujita, Rikiya Onimaru, Noriwo Katoh, Tetsuya Inoue, Shunsuke Onodera, Hiroki Shirato.   

Abstract

PURPOSE: To evaluate the effectiveness of the stereotactic body frame (SBF), with or without a diaphragm press or a breathing cycle monitoring device (Abches), in controlling the range of lung tumor motion, by tracking the real-time position of fiducial markers. METHODS AND MATERIALS: The trajectories of gold markers in the lung were tracked with the real-time tumor-tracking radiotherapy system. The SBF was used for patient immobilization and the diaphragm press and Abches were used to actively control breathing and for self-controlled respiration, respectively. Tracking was performed in five setups, with and without immobilization and respiration control. The results were evaluated using the effective range, which was defined as the range that includes 95% of all the recorded marker positions in each setup.
RESULTS: The SBF, with or without a diaphragm press or Abches, did not yield effective ranges of marker motion which were significantly different from setups that did not use these materials. The differences in the effective marker ranges in the upper lobes for all the patient setups were less than 1mm. Larger effective ranges were obtained for the markers in the middle or lower lobes.
CONCLUSION: The effectiveness of controlling respiratory-induced organ motion by using the SBF+diaphragm press or SBF + Abches patient setups were highly dependent on the individual patient reaction to the use of these materials and the location of the markers. They may be considered for lung tumors in the lower lobes, but are not necessary for tumors in the upper lobes. Copyright 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20171800     DOI: 10.1016/j.ijrobp.2009.08.060

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  Influence of respiration on dose calculation in stereotactic body radiotherapy of the lung.

Authors:  Rie Yamazaki; Rikiya Onimaru; Norio Katoh; Tetsuya Inoue; Takeshi Nishioka; Hiroki Shirato; Hiroyuki Date
Journal:  Radiol Phys Technol       Date:  2014-03-19

2.  Accuracy of Breath-hold CT in Treatment Planning for Lung Stereotactic Ablative Radiotherapy.

Authors:  Dominique Mathieu; Charles Martel; Marie-Pierre Campeau; Édith Filion; David Roberge; Houda Bahig; Toni Vu; Louise Lambert; Karim Boudam; Jean-Francois Carrier
Journal:  Cureus       Date:  2014-12-22

3.  Comparative analysis of thermoplastic masks versus vacuum cushions in stereotactic body radiotherapy.

Authors:  Arturo Navarro-Martin; Jon Cacicedo; Olwen Leaman; Ismael Sancho; Elvira García; Valentin Navarro; Ferran Guedea
Journal:  Radiat Oncol       Date:  2015-08-20       Impact factor: 3.481

4.  AAPM Task Group Report 290: Respiratory motion management for particle therapy.

Authors:  Heng Li; Lei Dong; Christoph Bert; Joe Chang; Stella Flampouri; Kyung-Wook Jee; Liyong Lin; Michael Moyers; Shinichiro Mori; Joerg Rottmann; Erik Tryggestad; Sastry Vedam
Journal:  Med Phys       Date:  2022-01-31       Impact factor: 4.506

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.