Literature DB >> 36147258

Distribution Pattern of Total Corneal Thickness in Keratoconus versus Normal Eyes Using an Optical Coherence Tomography.

Hassan Hashemi1, Zahra Heidari2,3, Mehrdad Mohammadpour1,4, Hamed Momeni-Moghaddam5,6, Mehdi Khabazkhoob7.   

Abstract

Purpose: To evaluate the total corneal thickness distribution pattern using a high-resolution spectral-domain optical coherence tomography (HR SD-OCT) for distinguishing normal eyes from keratoconus (KCN).
Methods: One hundred and forty-four patients were enrolled in three groups (55 normal, 45 mild KCN, and 44 moderate-to-severe KCN eyes) in this prospective diagnostic test study. Total corneal thickness was measured in 8 semi-meridians using HR SD-OCT (Heidelberg Engineering, Heidelberg, Germany) in 5 and 7 mm zones. The central corneal thickness (CCT), corneal focal thinning (minimum thickness [Min], min minus median and maximum [Min-Med, Min-Max]), and asymmetry indices (inferior minus superior [I-S] and supranasal minus infratemporal [SN-IT]) were calculated. One-way analysis of variance and the area under the receiver operating characteristic curve (AUC) were used for the analysis.
Results: Thinner CCT, lower Min thickness, more negative Min-Max, Min-Med, and greater I-S and SN-IT were found in KCN eyes compared to the control group (P < 0.001). The inferior and IT semi-meridians were the thinnest locations in KCN cases in the 5 mm central zone (P < 0.001). CCT followed by Min-Med had the highest discriminative ability for differentiating mild KCN (AUC, sensitivity and specificity: 0.822, 87.0%, 60.37% and 0.805, 82.93%, 66.0%, respectively) and moderate-to-severe KCN (0.902, 87.82%, 73.08% and 0.892, 85.37%, and 78.85%, respectively) from normal corneas.
Conclusion: The inferior and IT sectors of the cornea with the largest thickness changes in the 5 mm zone are the most common thinning sites in keratoconic corneas, and CCT and Min-Med are the most sensitive indices for the diagnosis of KCN. Copyright:
© 2022 Journal of Current Ophthalmology.

Entities:  

Keywords:  Corneal thickness; Keratoconus; Spectral-domain optical coherence tomography

Year:  2022        PMID: 36147258      PMCID: PMC9487006          DOI: 10.4103/joco.joco_198_21

Source DB:  PubMed          Journal:  J Curr Ophthalmol        ISSN: 2452-2325


INTRODUCTION

Keratoconus (KCN) is a bilateral noninflammatory ectatic corneal disorder characterized by asymmetry in corneal pachymetry and corneal thinning, increased curvature, and the development of irregular astigmatism.12 Because KCN is a common implication for corneal transplantation,3 a comprehensive corneal assessment is important in the timely diagnosis and management of the disease. Corneal thickness is considered an important parameter in KCN diagnosis,2 and accurate corneal thickness measurement at the central (CCT) and thinnest points is important for refractive surgery screening.4 This parameter associated with peripheral corneal thickness is important for corneal cross-linking and intrastromal corneal ring implantation in KCN management.5 There are different methods to measure corneal thickness, including ultrasound pachymetry,6 confocal microscopy,7 and slit scanning, Scheimpflug tomography.89 The first two are contact methods that measure CCT at certain points and are associated with a possible risk of infection transmission.10 In addition, their precision is operator dependent and measurements may not be accurate if the probe is not positioned accurately.11 Corneal thickness measurement using tomography imaging systems is preferred to other methods because they can provide a thickness map of the cornea to assess the normality of the thickness map pattern; however, these devices still have some disadvantages such as resolution limitation, which may lead to underestimation of the corneal thickness, especially in the postrefractive assessments.12 In addition, in advanced KCN cases, these devices may fail to provide acceptable maps due to the extensive irregularity of the corneal surface and difficulty in the fixation of KCN eyes.13 Furthermore, they can hardly differentiate corneal irregularity caused by tear film problems or corneal scars from true KCN.14 Today, anterior segment optical coherence tomography (AS-OCT) can assess the anterior and posterior corneal surfaces with a high resolution (HR) based on the reflectivity difference in tissues using time-domain and spectral-domain (SD) methods. The SD method, also known as Fourier-domain, provides higher resolution images at a higher scanning speed and has a deeper penetration into the tissue15 so that corneal thickness measurement using this method may provide information that is not available by tomography technique alone. Previous studies measured the corneal thickness using different AS-OCT devices;111617 however, the exact distribution of corneal thickness along different meridians and the diagnostic ability of OCT parameters in differentiating on KCN from normal eyes have been rarely reported. The purpose of this study was to evaluate the distribution pattern of total corneal thickness and determine the diagnostic ability of different corneal pachymetric parameters using a HR SD-OCT for distinguishing KCN from normal eyes.

METHODS

This diagnostic test study was a part of a prospective study conducted between February 20 and December 21, 2018, at Noor Eye Hospital. The research project was approved by the Ethics Committee of Tehran University of Medical Sciences (code: IR.TUMS.VCR.REC.1396.4815). All stages of the current study were conducted according to the principles of the Declaration of Helsinki. The objectives of this study were explained to the participants, and informed consent was obtained before the study. Participants were recruited from patients referred to the KCN clinic and those referred to the refractive surgery unit. If both eyes were eligible for inclusion, one eye was randomly selected. All patients underwent complete ocular examinations including the measurement of corrected distance visual acuity using a Snellen chart, refractive assessment using retinoscopy by a Heine Streak Retinoscope, slit-lamp biomicroscopy, and fundoscopy. The anterior and posterior corneal surfaces were assessed using the Scheimpflug tomography by Pentacam HR (Oculus, Wetzlar, Germany) and AS-OCT technique using HR SD-OCT (Spectralis, Heidelberg Engineering, Heidelberg, Germany; software version 3.1.0.55). Diagnosis of KCN was based on the presence of scissoring reflex on retinoscopy, and at least one of the following slit-lamp signs including the Fleischer ring, Vogt striae, apical thinning, Munson's sign, Rizutti's sign, and the presence of abnormal topographic criteria using the Pentacam including a skewed asymmetric bow tie or inferior steepening or a claw pattern, SRAX >21°, anterior Kmax >49.0 diopter (D) or inferior-superior (I-S) value >1.9 D.118 Patients were assigned by an experienced corneal specialist with more than 10 years of experience. The severity of KCN was classified based on Pentacam Topographic KCN Classification (TKC). TKC provides four grades (I–IV) for KCN classification. In some intermediate grades (e.g., III–VI), lower values were documented. The KCN Group 1 (G1) corresponded to Grade I, and the KCN Group 2 (G2) corresponded to Grades II–VI. The normal group was selected from refractive surgery candidates who had a normal clinical examination based on retinoscopy and slit-lamp biomicroscopy, and no abnormal topographic criteria.118 The exclusion criteria were any history of ocular surgery, corneal cross-linking, ring implantation, corneal hydrops, corneal scar, any systemic diseases affecting the eyes, and signs and symptoms of dry eye. The patients with a history of contact lens wear in the past 4 weeks were also excluded. Spectralis anterior segment module provides HR images (axial and transverse resolutions 3.9 and 14 μm, respectively) at a high scanning speed per second (40,000 A-scans/s) and can display the corneal thickness in detail.1920 Moreover, this module provides angle-to-angle images for precise assessment of the corneal angles and scleral structure. During the imaging, patients were asked to fixate on a blue light target and blink at intervals between measurements to avoid dry eye. After proper fixation, an average of three measurements was taken in vertical, horizontal, and diagonal meridians were recorded. After imaging, the total corneal thickness was measured in 8 semi-meridians at 45° intervals in two zones of 5 and 7 mm centered on the corneal geometric center. Horizontal scans were used to measure along the nasal (0°) and temporal (180°) semi-meridians, vertical scans were applied to measure along with the superior (90°) and inferior (270°) semi-meridians, and diagonal scans were used to measure along the supranasal (45°), infratemporal (225°), supratemporal (135°), and infranasal (315°) semi-meridians [Figures 1 and 2].
Figure 1

Cross-sectional optical coherence tomography images in all eight meridians in the keratoconus eye

Figure 2

Sector map shows mean pachymetry in each zone in normal and keratoconus eyes (I: Inferior, S: Superior, T: Temporal, N: Nasal, IN: Infranasal, IT: Infratemporal, SN: Supranasal, ST: Supratemporal)

Cross-sectional optical coherence tomography images in all eight meridians in the keratoconus eye Sector map shows mean pachymetry in each zone in normal and keratoconus eyes (I: Inferior, S: Superior, T: Temporal, N: Nasal, IN: Infranasal, IT: Infratemporal, SN: Supranasal, ST: Supratemporal) Moreover, the mean absolute CCT was calculated in semi-meridians in the geometric center of the cornea and recorded in the center of the annular rings. Then, the indices of focal thinning including minimum (Min) thickness, the difference between the minimum and median thickness (Min-Med), and the difference between the minimum and maximum thickness (Min-Max) were calculated for all measurement points. Corneal asymmetry indices including the difference between the mean thickness of the inferior and superior octants of the cornea (I-S), the difference between the mean thickness of supranasal and infratemporal octants of the cornea (supranasal-infratemporal [SN-IT]) were calculated for both 5 mm and 7 mm zones. For evaluation of the intraobserver repeatability test, all measurements were repeated 2 weeks apart by the same observer who was blind to the grouping of the cases. Because the corneal epithelium was not distinguishable from the tear film in some cases, for better interpretation and comparison of the results, the tear film (the first hyperreflective layer) was included in the thickness measurements (the tear film thickness is about 3–6 μ).21 All imaging techniques were performed in a dark room between 9 am and 5 pm by an experienced operator to avoid any bias. SPSS 23 software (IBM Inc., New York, NY, USA) was used for statistical analyses. The normality of data was assessed using the Kolmogorov–Smirnov test. Descriptive statistics including mean, standard deviation, and range of corneal thickness were calculated for each semi-meridian in two 5 and 7 mm zones in both groups. The mean values of the variables were compared between the two groups using the independent t-test. A P < 0.05 was considered statistically significant. A receiver operating characteristic (ROC) curve analysis was performed for all variables to determine the area under the ROC curve (AUC) to evaluate the accuracy of different variables. The Youden index was applied to calculate the optimum cut-off value. To compare the diagnostic ability of each parameters, sensitivity (true positive/[true positive + false negative]) and specificity (true negative/[true negative + false positive]) were reported for all variables. The intraclass correlation coefficient (ICC) was performed to assess the reliability of the repeated measurements. An ICC > 0.80 was considered excellent repeatability.19

RESULTS

A total of 144 eyes (55 normal, 45 mild KCN, and 44 moderate-to-severe KCN eyes) were included in the study. KCN groups included 45 eyes (50.56%) with Grade I (G1), 27 eyes (30.33%) with Grade II, and 17 eyes (19.11%) with Grade III (G2) KCN. Females accounted for 58.2% of normals, whereas 62.9% of the patients with KCN (G1 and G2) were male. The demographic characteristics of the participants are summarized in Table 1, and the mean values of pachymetric parameters are presented in Table 2.
Table 1

Demographic characteristics of study groups

VariablesMean±SD (range)P (ANOVA)

Normal (n=55)KCN G1 (n=45)KCN G2 (n=44)Total (n=144)
Age31.07±7.06 (19-49)32.66±4.70 (25-47)31.68±5.97 (24-46)31.81±6.02 (19-49)0.493
Sphere (D)−2.44±2.27 (−6.00 - +8.00)−1.92±1.95 (−6.75 - +1.00)−2.81±3.53 (−13.50 - +3.75)−2.41±2.62 (−13.5 - +8.00)0.271
Cylinder (D)−1.73±1.68 (0.00 - −7.25)−2.42±1.55 (0.00 - − 8.50)−3.69±1.99 (−0.75 - − 10.75)−2.48±1.88 (0.00 - −10.75) <0.001
CDVA, logMAR0.01±0.05 (0.00 - +0.30)0.13±0.19 (0.10 - + 0.70)0.37±0.19 (0.10 - +0.80)0.16±0.21 (0.00 - +0.80) <0.001

**The significant variables highlighted (ANOVA). ANOVA: Analysis of variance, SD: Standard deviation, KCN: Keratoconus, KCN G1: Mild KCN group, KCN G2: Moderate and severe KCN groups, D: Diopter, CDVA: Corrected distance visual acuity

Table 2

Mean and standard deviation of corneal thickness parameters separately in normal and keratoconus eyes

Variable (µm)Mean±SD (range)P (ANOVA)P (normal versus G1)P (normal versus G2)

Normal (n=55)KCN G1 (n=45)KCN G2 (n=44)
Central corneal thickness558.49±37.59 (473-689)512.24±32.93 (439-583)486.02±43.38 (413-641) <0.001 <0.001 <0.001
Minimum thickness531.88±40.29 (443-662)485.73±30.98 (413-556)461.87±42.78 (383-617) <0.001 <0.001 <0.001
Minimum-median thickness−39.03±9.97 (−75 - −19)−50.23±13.73 (−97 - −24)−66.18±19.83 (−102 - −22) <0.001 0.001 <0.001
Minimum-maximum thickness−98.50±19.22 (−138 - −62)−113±26.78 (−201 - −69)−124.26±26.38 (−189 - −77) <0.001 0.007 <0.001
Total corneal thickness (5 mm)
 S585.03±40.08 (500-659)559.19±39.57 (451-648)555.95±45.27 (450-681) <0.001 0.010 0.003
 I564.81±39.08 (488-669)518.73±35.68 (430-585)501.92±48.81 (414-661) <0.001 <0.001 <0.001
 N569.00±40.25 (499-645)537.04±32.93 (444-605)538.43±44.22 (451-675) <0.001 <0.001 0.001
 T552.56±36.49 (459-639)519.46±35.28 (423-592)513.07±44.89 (431-660) <0.001 <0.001 <0.001
 SN580.33±37.27 (511-657)551.36±35.45 (456-625)553.07±46.55 (469-694) <0.001 0.002 0.004
 IT552.50±38.81 (472-629)509.0±34.42 (433-577)494.0±44.81 (406-619) <0.001 <0.001 <0.001
 ST571.62±36.75 (480-642)545.85±34.54 (438-618)541.21±44.84 (453-677) <0.001 0.005 0.001
 IN566.81±38.71 (566-659)529.56±33.09 (450-606)525.19±44.63 (432-657) <0.001 <0.001 <0.001
 I-S−19.62±20.74 (−75-29)−37.17±35.82 (−131-91)−52.90±33.85 (−124-11) <0.001 0.017 <0.001
 SN-IT28.00±24.22 (−53-79)46.41±26.65 (−2-138)62.27±32.42 (−8-138) <0.001 0.005 <0.001
Total corneal thickness (7 mm)
 S638.09±42.20 (564-724)613.02±43.07 (489-687)603.26±47.08 (507-721) 0.001 0.021 0.001
 I608.28±43.91 (528-717)557.53±36.66 (458-629)558.07±46.90 (477-706) <0.001 <0.001 <0.001
 N610.05±47.71 (511-713)578.17±34.20 (490-665)577.48±42.59 (481-709) <0.001 0.001 0.001
 T586.84±40.80 (490-684)567.21±37.78 (465-636)553.14±42.55 (475-672) <0.001 0.002 <0.001
 SN624.71±41.72 (520-713)597.09±40.04 (489-687)590.60±49.21 (487-731) <0.001 0.009 <0.001
 IT588.52±41.46 (520-679)548.56±35.17 (465-625)542.30±42.18 (462-636) <0.001 <0.001 <0.001
 ST607.43±45.68 (519-694)580.46±39.82 (482-672)576.68±42.97 (492-706) 0.001 0.010 0.002
 IN607.20±43.47 (517-710)573.41±33.34 (481-660)576.07±43.33 (474-697) <0.001 <0.001 0.001
 I-S−29.81±27.89 (−99-35)−44.29±31.34 (−126-24)−43.92±32.10 (−111-24) 0.030 0.0690.080
 SN-IT35.25±30.68 (−41-100)49.36±25.15 (−13-112)50.87±38.77 (−48-137) 0.034 0.1070.064

**The significant variables highlighted (ANOVA, Bonferroni). ANOVA: Analysis of variance, SD: Standard deviation, KCN: Keratoconus, KCN G1: Mild KCN group, KCN G2: Moderate and sever KCN groups, S: Superior, I: Inferior, N: Nasal, T: Temporal, SN: Supranasal, IT: Infratemporal, ST: Supratemporal, IN: Infranasal, I-S: Inferior-superior, SN-IT: Supranasal-infratemporal

Demographic characteristics of study groups **The significant variables highlighted (ANOVA). ANOVA: Analysis of variance, SD: Standard deviation, KCN: Keratoconus, KCN G1: Mild KCN group, KCN G2: Moderate and severe KCN groups, D: Diopter, CDVA: Corrected distance visual acuity Mean and standard deviation of corneal thickness parameters separately in normal and keratoconus eyes **The significant variables highlighted (ANOVA, Bonferroni). ANOVA: Analysis of variance, SD: Standard deviation, KCN: Keratoconus, KCN G1: Mild KCN group, KCN G2: Moderate and sever KCN groups, S: Superior, I: Inferior, N: Nasal, T: Temporal, SN: Supranasal, IT: Infratemporal, ST: Supratemporal, IN: Infranasal, I-S: Inferior-superior, SN-IT: Supranasal-infratemporal The cornea was thinner in the center and became thicker toward the periphery; corneal pachymetric indices were thinner in 5 mm zone compared to 7 mm zone (P < 0.001). The mean CCT was 558.49 ± 37.59, 512.24 ± 32.93, and 486.02 ± 43.38 μm in normal, G1 and G2 KCN eyes, respectively (P < 0.001). Evaluation of the total corneal thickness indices in both 5 and 7 mm zones showed that all parameters were significantly thinner in KCN eyes (P < 0.001). The mean thickness was thinner in inferior and temporal semi-meridians compared to superior and nasal semi-meridians at 5 mm and 7 mm zones in all groups. In KCN groups, the thinnest pachymetric values were found in the inferior and IT semi-meridians, and the thickest corneal area was seen in the superior semi-meridian in both diameter zones. Evaluation of the corneal focal thinning and asymmetric parameters showed lower Min thickness, greater I-S and SN-IT at 5 and 7 mm, and more negative Min-Max and Min-Med in the KCN eyes compared to the control group (P < 0.001); however, corneal asymmetric parameters were higher in 5 mm zone than the 7 mm zone in KCN group (P < 0.001). The total and central corneal thickness (CCT) measurements showed excellent reliability in all groups (ICC >0.80). The ROC curve analysis is shown in Table 3. The highest discriminative ability was found for CCT in G1 (AUC 0.822, sensitivity 87.0%, and specificity 60.37%) and G2 (AUC 0.902, sensitivity 87.82%, and specificity 73.08%), followed by the Min-Med index in G1 (AUC 0.805, sensitivity 82.93%, and specificity 66.0%) and G2 (AUC 0.892, sensitivity 85.37%, and specificity 78.85%). The inferior and IT semi-meridians had the highest AUCs compared to other sectors in both 5 mm and 7 mm zones [Table 3]. ROC curves of thickness parameters with the highest diagnostic ability are demonstrated in Figure 3.
Table 3

The result of receiver operating characteristics analysis to differentiate eyes with keratoconus from normal eyes

Variable (µm)Normal versus KCN G1Normal versus KCN G2


AUC (95% CI)Cut-offSensitivitySpecificityAUC (95% CI)Cut-offSensitivitySpecificity
Central corneal thickness0.822 (0.730-0.893)≤52087.8060.370.902 (0.826-0.956)≤50887.8273.08
Minimum thickness0.759 (0.660-0.842)≤51651.2288.680.865 (0.782-0.929)≤49756.3496.50
Minimum–median thickness0.805 (0.711-0.880)≤−4082.9366.00.892 (0.813-0.948)≤−4585.6778.85
Minimum–maximum thickness0.679 (0.575-0.772)≤−12341.4688.680.779 (0.678-0.856)≤−10680.561.5
Total corneal thickness (5 mm)
 S0.685 (0.581-0.777)≤56563.4077.700.701 (0.605-0.798)≤56056.176.92
 I0.793 (0.697-0.870)≤54280.4964.150.862 (0.779-0.928)≤52975.686.5
 N0.721 (0.619-0.808)≤55275.6066.040.704 (0.607-0.800)≤54768.373.1
 T0.742 (0.641-0.826)≤72458.579.20.768 (0.675-0.854)≤53065.975.0
 SN0.708 (0.606-0.798)≤57782.956.60.697 (0.593-0.789)≤56567.567.3
 IT0.789 (0.692-0.866)≤54080.4160.380.855 (0.766-0.920)≤51877.584.62
 ST0.699 (0.597-0.791)≤54856.1077.360.773 (0.626-0.815)≤53958.570.1
 IN0.764 (0.666-0.846)≤54775.6167.90.776 (0.678-0.856)≤52763.4188.46
 I-S0.694 (0.593-0.787)≤−4048.7886.800.807 (0.701-0.873)≤−4168.5086.54
 SN-IT0.694 (0.590-0.785)>3763.467.30.80 (0.706-0.879)>4277.5074.50
Total corneal thickness (7 mm)
 S0.647 (0.538-0.740)≤63273.252.80.717 (0.614-0.805)≤64280.053.66
 I0.758 (0.658-0.840)≤59280.056.980.799 (0.709-0.879)≤59585.259.62
 N0.693 (0.589-0.784)≤60880.047.170.704 (0.600-0.794)≤60982.9346.2
 T0.707 (0.604-0.796)≤54846.388.80.738 (0.637-0.824)≤54146.392.3
 SN0.686 (0.582-0.778)≤62579.952.80.724 (0.622-0.812)≤62485.453.8
 IT0.746 (0.646-0.830)≤57880.454.720.772 (0.680-0.859)≤58585.353.8
 ST0.671 (0.566-0.764)≤60878560.708 (0.605-0.798)≤60885.457.7
 IN0.731 (0.566-0.764)≤57763.475.50.716 (0.607-0.800)≤57558.578.8
 I-S0.640 (0.536-0.738)≤−53.441.4680.00.636 (0.519-0.724)≤−2278.045.0
 SN-IT0.633 (0.526-0.730)>2880.1144.240.640 (0.533-0.738)>4365.060.78

**The highest AUCs highlighted. AUC: Area under the receiver operating characteristic curve, CI: Confidence interval, KCN: Keratoconus, KCN G1: Mild KCN group, KCN G2: Moderate and sever KCN groups, S: Superior, I: Inferior, N: Nasal, T: Temporal, SN: Supranasal, IT: Infratemporal, ST: Supratemporal, IN: Infranasal, I-S: Inferior-superior, SN-IT: Supranasal-infratemporal

Figure 3

Comparison of parameters with the highest area under the receiver operating characteristic curve to differentiate mild (a), and moderate-to-severe (b) keratoconus from normal eyes (CCT: Central corneal thickness, Min-Med: Minimum-median thickness, IT: Infratemporal)

The result of receiver operating characteristics analysis to differentiate eyes with keratoconus from normal eyes **The highest AUCs highlighted. AUC: Area under the receiver operating characteristic curve, CI: Confidence interval, KCN: Keratoconus, KCN G1: Mild KCN group, KCN G2: Moderate and sever KCN groups, S: Superior, I: Inferior, N: Nasal, T: Temporal, SN: Supranasal, IT: Infratemporal, ST: Supratemporal, IN: Infranasal, I-S: Inferior-superior, SN-IT: Supranasal-infratemporal Comparison of parameters with the highest area under the receiver operating characteristic curve to differentiate mild (a), and moderate-to-severe (b) keratoconus from normal eyes (CCT: Central corneal thickness, Min-Med: Minimum-median thickness, IT: Infratemporal)

DISCUSSION

Regarding thickness parameters in normal eyes, the mean CCT in the current study (558.49 ± 37.59 μm) was close to the value reported by López de la Fuente et al.19(555.5 ± 29.64 μm) using the same technique but higher than the value reported by Catalan et al.22(537.6 ± 30.66 μm). Moreover, the values of focal thinning and asymmetric parameters in the normal group were larger than the results of a study by Kanellopoulos and Asimellis23 using the RTVue OCT in normal corneas. It seems that differences in the results are due to different measurement techniques. In the present study, the measured thickness parameters in KCN eyes for CCT, Min thickness, Min-Med, and Min-Max were larger than the results of a study reported by Catalan et al.22 and Kanellopoulos et al.23 However, evaluation of corneal asymmetry indices showed lower values of I-S and SN-IT at 5 mm zone in the KCN group compared to Temstet et al.'s24 study reports, which could be due to the differences in the severity of KCN in different populations. However, it is noteworthy that in recent studies, these parameters had significantly larger mean differences in KCN versus normal eyes. In general, higher mean thickness values in the present study compared to some previous studies that used other methods of corneal thickness measurement may probably be due to including the tear film in the thickness measurement. In the presence of corneal ectasia, especially in severe KCN cases, it is not possible to differentiate the precorneal tear film (PTF) of the epithelium layer. Therefore, in the current study, tear film was included in corneal thickness components to enhance the accuracy of the measurement. Including PTF in measurements has a low coefficient of variation and high intraobserver repeatability and interobserver reproducibility.19 The present study also showed high repeatability in measuring the total corneal thickness. Few studies have reported corneal pachymetric parameters in different meridians separately. Comparison of thickness in the opposite semi-meridians showed a thinner thickness in the inferior versus superior, temporal versus nasal, and IT versus SN semi-meridians in KCN and normal eyes which was similar to the findings reported by Dutta et al.10 in KCN eyes and Li et al.25 in normal eyes in different meridians using RTVue OCT. However, the present study showed that the inferior and IT semi-meridians in the 5 mm zone had the thinnest thickness in KCN cases. Thinner corneas in the inferior and IT semi-meridians could be due to the inferior displacement of the corneal cone in KCN eyes, and more exposure of these corneal parts to trauma from eye rubbing.2627 Abnormal eye rubbing may lead to the progression of KCN,28 increased oxidative stress, loss of keratocytes, and stromal scarring.29 Although KCN is a disease with a “pancorneal” pathology30 and the paracentral and peripheral corneal thickness can provide valuable information about the progression of corneal thinning, the largest mean difference in pachymetric indices between normal and keratoconic corneas was in the 5 mm zone. ROC analysis showed that CCT and focal thinning parameters, especially Min-Med thickness, had the best diagnostic ability for differentiating KCN from normal corneas. Among total corneal thickness parameters in different semi-meridians, inferior, and IT semi-meridians in the 5 mm zone showed the highest accuracy for KCN detection. Reviewing the literature showed that few studies have investigated the accuracy of focal and total corneal thickness in different semi-meridians, and most of these studies have focused on the epithelial thickness map, which is not available in all clinical centers. Catalan et al.22 studied 22 KCN and 104 normal eyes using the RTVue OCT and found that the Min-Med parameter had the highest diagnostic accuracy (AUC = 0.974). In addition, they found that S-I, with an AUC similar to Min-Med, was the best distinguishing index in both groups; however, this parameter showed less accuracy in the present study. In another study, Li et al.31 compared 37 KCN and 36 normal eyes using the Visante OCT system and found that Min thickness had the highest accuracy (AUC = 0.954). These studies used different measurement techniques from our study and evaluated the cornea from the center to the 5 mm zone, while we also studied the 7 mm corneal zone. In contrast to the present study, they did not investigate KCN corneas in different grades. One of the limitations of the current study was the manual measurement of parameters; however, the results of some parameters were similar to studies using automated measurements. In addition, this method represents a simple measurement for quick evaluation of patients in clinical practice. In summary, the results highlighted an increase in both corneal thickness asymmetry and focal thinning indices and a decrease in corneal thickness in different semi-meridians, especially the inferior and IT sectors of the cornea, in ectatic versus normal eyes. Evaluation of the total corneal thickness showed that the largest thickness changes probably occur in the 5 mm zone and, therefore, this zone is accurate enough for KCN diagnosis. Among all studied indices, the CCT and Min-Med thickness index for assessment of total corneal thickness, and inferior and the IT semi-meridians thickness, especially in the 5 mm zone may be helpful for screening and early detection of KCN eyes with high accuracy. These parameters can be used as complementary diagnostic indices in Scheimpflug tomography devices that provide the corneal thickness map.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  30 in total

1.  Corneal-thickness spatial profile and corneal-volume distribution: tomographic indices to detect keratoconus.

Authors:  Renato Ambrósio; Ruiz Simonato Alonso; Allan Luz; Luis Guillermo Coca Velarde
Journal:  J Cataract Refract Surg       Date:  2006-11       Impact factor: 3.351

2.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

3.  OCT-derived comparison of corneal thickness distribution and asymmetry differences between normal and keratoconic eyes.

Authors:  Anastasios J Kanellopoulos; George Asimellis
Journal:  Cornea       Date:  2014-12       Impact factor: 2.651

4.  Corneal epithelial thickness mapping using Fourier-domain optical coherence tomography for detection of form fruste keratoconus.

Authors:  Cyril Temstet; Otman Sandali; Nacim Bouheraoua; Taous Hamiche; Alice Galan; Mohamed El Sanharawi; Elena Basli; Laurent Laroche; Vincent Borderie
Journal:  J Cataract Refract Surg       Date:  2015-04       Impact factor: 3.351

5.  Pseudokeratoconus caused by rosacea.

Authors:  D Dursun; A M Piniella; S C Pflugfelder
Journal:  Cornea       Date:  2001-08       Impact factor: 2.651

6.  Central corneal pachymetry in patients undergoing laser in situ keratomileusis.

Authors:  F W Price; D L Koller; M O Price
Journal:  Ophthalmology       Date:  1999-11       Impact factor: 12.079

7.  A Case-Control Study of Keratoconus Risk Factors.

Authors:  Sarah Moran; Ludovic Gomez; Kevin Zuber; Damien Gatinel
Journal:  Cornea       Date:  2020-06       Impact factor: 2.651

8.  Corneal thickness in keratoconus: comparing optical, ultrasound, and optical coherence tomography pachymetry.

Authors:  Debarun Dutta; Harsha L Rao; Uday K Addepalli; Pravin K Vaddavalli
Journal:  Ophthalmology       Date:  2012-11-21       Impact factor: 12.079

9.  Evaluation of Corneal Epithelial Healing Under Contact Lens with Spectral-Domain Anterior Segment Optical Coherence Tomography (SD-OCT).

Authors:  Claudine E Pang; Vanathi M; Donald T H Tan; Jodhbir S Mehta
Journal:  Open Ophthalmol J       Date:  2011-06-01

10.  Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Measurement of Thinnest Corneal Thickness in Keratoconus.

Authors:  Chunxiao Wang; Xueying Xia; Bishan Tian; Shiyou Zhou
Journal:  J Ophthalmol       Date:  2015-06-25       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.