Literature DB >> 25878306

Radiostereometric analysis study of tantalum compared with titanium acetabular cups and highly cross-linked compared with conventional liners in young patients undergoing total hip replacement.

David C Ayers1, Meridith Greene2, Benjamin Snyder1, Michelle Aubin1, Jacob Drew1, Charles Bragdon2.   

Abstract

BACKGROUND: Radiostereometric analysis provides highly precise measurements of component micromotion relative to the bone that is otherwise undetectable by routine radiographs. This study compared, at a minimum of five years following surgery, the micromotion of tantalum and titanium acetabular cups and femoral head penetration in highly cross-linked polyethylene liners and conventional (ultra-high molecular weight polyethylene) liners in active patients who had undergone total hip replacement.
METHODS: This institutional review board-approved prospective, randomized, blinded study involved forty-six patients. Patients were randomized into one of four cohorts according to both acetabular cup and polyethylene liner. Patients received either a cementless cup with a titanium mesh surface or a tantalum trabecular surface and either a highly cross-linked polyethylene liner or an ultra-high molecular weight polyethylene liner. Radiostereometric analysis examinations and Short Form-36 Physical Component Summary, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), University of California Los Angeles (UCLA) activity, and Harris hip scores were obtained preoperatively, postoperatively, at six months, and annually thereafter.
RESULTS: All patients had significant improvement (p < 0.05) in Short Form-36 Physical Component Summary, WOMAC, UCLA activity, and Harris hip scores postoperatively. On radiostereometric analysis examination, highly cross-linked polyethylene liners showed significantly less median femoral head penetration at five years (p < 0.05). Steady-state wear rates from one year to five years were 0.04 mm per year for ultra-high molecular weight polyethylene liners and 0.004 mm per year for highly cross-linked polyethylene liners. At the five-year follow-up, the median migration (and standard error) was 0.05 ± 0.20 mm proximally for titanium cups and 0.21 ± 0.05 mm for tantalum cups.
CONCLUSIONS: In this young population who had undergone total hip replacement, radiostereometric analysis showed significantly less femoral head penetration in the highly cross-linked polyethylene liners compared with that in the conventional ultra-high molecular weight polyethylene liners. Penetration rates were one order of magnitude less in highly cross-linked polyethylene liners compared with ultra-high molecular weight polyethylene liners. There was no significant difference in proximal migration between the tantalum and titanium acetabular cups through the five-year follow-up (p > 0.19).
Copyright © 2015 by The Journal of Bone and Joint Surgery, Incorporated.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25878306      PMCID: PMC4391964          DOI: 10.2106/JBJS.N.00605

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  33 in total

Review 1.  Cross-linked versus conventional polyethylene for total hip replacement: a meta-analysis of randomised controlled trials.

Authors:  P R T Kuzyk; M Saccone; S Sprague; N Simunovic; M Bhandari; E H Schemitsch
Journal:  J Bone Joint Surg Br       Date:  2011-05

2.  Primary total hip arthroplasty with a porous-coated acetabular component. A concise follow-up, at a minimum of twenty years, of previous reports.

Authors:  Craig J Della Valle; Nathan W Mesko; Laura Quigley; Aaron G Rosenberg; Joshua J Jacobs; Jorge O Galante
Journal:  J Bone Joint Surg Am       Date:  2009-05       Impact factor: 5.284

3.  The epidemiology of revision total hip arthroplasty in the United States.

Authors:  Kevin J Bozic; Steven M Kurtz; Edmund Lau; Kevin Ong; Thomas P Vail; Daniel J Berry
Journal:  J Bone Joint Surg Am       Date:  2009-01       Impact factor: 5.284

4.  Multicentre use of a porous tantalum monoblock acetabular component.

Authors:  Th A Xenakis; G A Macheras; K S Stafilas; A T Kostakos; K Bargiotas; K N Malizos
Journal:  Int Orthop       Date:  2008-05-29       Impact factor: 3.075

5.  The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis.

Authors:  Geraint E R Thomas; David J Simpson; Shahid Mehmood; Adrian Taylor; Peter McLardy-Smith; Harinderjit Singh Gill; David W Murray; Siôn Glyn-Jones
Journal:  J Bone Joint Surg Am       Date:  2011-04-20       Impact factor: 5.284

6.  A monoblock porous tantalum acetabular cup has no osteolysis on CT at 10 years.

Authors:  Todd C Moen; Raju Ghate; Noel Salaz; Jason Ghodasra; S David Stulberg
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

7.  Two-year radiostereometric analysis evaluation of femoral head penetration in a challenging population of young total hip arthroplasty patients.

Authors:  David C Ayers; Peyton L Hays; Jacob M Drew; Mark S Eskander; Daniel Osuch; Charles R Bragdon
Journal:  J Arthroplasty       Date:  2009-09       Impact factor: 4.757

8.  Polyethylene wear and osteolysis with a new modular titanium acetabular component: results at 7 to 13 years.

Authors:  Paul F Lachiewicz; John B Hubbard; Elizabeth S Soileau
Journal:  J Arthroplasty       Date:  2008-03-04       Impact factor: 4.757

9.  Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial.

Authors:  Richard W McCalden; Steven J MacDonald; Cecil H Rorabeck; Robert B Bourne; David G Chess; Kory D Charron
Journal:  J Bone Joint Surg Am       Date:  2009-04       Impact factor: 5.284

10.  Comparison of trabecular metal cups and titanium fiber-mesh cups in primary hip arthroplasty: a randomized RSA and bone mineral densitometry study of 50 hips.

Authors:  Thomas Baad-Hansen; Søren Kold; Poul Torben Nielsen; Mogens Berg Laursen; Poul Hedevang Christensen; Kjeld Soballe
Journal:  Acta Orthop       Date:  2011-03-25       Impact factor: 3.717

View more
  5 in total

Review 1.  [Endoprosthetic wear analysis using virtual CAD-based radiographs].

Authors:  M Haversath; S Klebingat; M Jäger
Journal:  Orthopade       Date:  2018-10       Impact factor: 1.087

Review 2.  Tantalum Versus Titanium Acetabular Cups in Primary Total Hip Arthroplasty: Current Concept and a Review of the Current Literature.

Authors:  Rohit Rambani; Mayur Nayak; Miss Sheweidin Aziz; Krishan Almeida
Journal:  Arch Bone Jt Surg       Date:  2022-05

Review 3.  Role of Hip Arthroscopy in the Treatment of Avascular Necrosis of the Hip: A Systematic Review.

Authors:  Omid Shahpari; Javad Mortazavi; Mohammad H Ebrahimzadeh; Farshid Bagheri; Alireza Mousavian
Journal:  Arch Bone Jt Surg       Date:  2022-06

Review 4.  Efficacy and Safety of Acetabular Cup without Screw Fixation in Total Hip Arthroplasty: A Systematic Review and Meta-Analysis.

Authors:  Takanori Miura; Hiroaki Kijima; Ryota Kimura; Jun Watanabe; Yuji Okazaki; Naohisa Miyakoshi
Journal:  Medicina (Kaunas)       Date:  2022-08-05       Impact factor: 2.948

5.  National Trends in Total Hip Arthroplasty Bearing Surface Usage in Extremely Young Patients Between 2006 and 2016.

Authors:  Christopher M Hart; Clark Chen; Peter P Hsiue; Reza Farshchi; Mauricio Silva; Erik Zeegen; Rachel Thompson; Alexandra Stavrakis
Journal:  Arthroplast Today       Date:  2021-07-09
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.