Craig J Della Valle1, Nathan W Mesko, Laura Quigley, Aaron G Rosenberg, Joshua J Jacobs, Jorge O Galante.
Abstract
We previously reported the seven and fifteen-year results of the use of a porous-coated acetabular metal shell inserted without cement in a consecutive series of 204 primary total hip arthroplasties. In the present study, we evaluated the longer-term outcomes of these arthroplasties at a minimum follow-up time of twenty years. One hundred and fourteen (92%) of the 124 hips available for study had retained the original acetabular metal shell. A total of five acetabular components had been revised for aseptic loosening or had radiographic evidence of definite loosening. Fourteen hips with well-fixed acetabular shells required a change of the modular acetabular liner because of excessive wear and/or for the treatment of osteolysis, and liner changes have been recommended for another eight hips. The twenty-year rate of survival of the metal shell, with failure defined as revision because of loosening or radiographic evidence of loosening, was 96% (95% confidence interval, 94% to 98%). Cementless acetabular reconstruction continues to provide durable fixation at twenty years postoperatively. Wear-related complications continue to be the major mode of failure.Entities:
Mesh:
Year: 2009 PMID: 19411461 DOI: 10.2106/JBJS.H.00168