Literature DB >> 21545905

Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station.

Marlin Wayne Causey1, Seth Miller, Andrew Foster, Alec Beekley, David Zenger, Matthew Martin.   

Abstract

BACKGROUND: Hemoglobin levels must be obtained through blood draws, which are invasive, time-consuming, and provide only 1 data point at a time rather than continuous measurements. The Masimo Radical-7 SpHb Station (Masimo Corporation, Irvine, CA) has been shown by its manufacturers to provide accurate noninvasive hemoglobin measurements in physiologically normal patients. The objective of this study was to validate noninvasive hemoglobin measurements using the Masimo Radical-7 device.
METHODS: Data were prospectively collected in 2 cohorts of patients: major operations requiring hemodynamic monitoring (operating room [OR]) and critically ill patients (intensive care unit [ICU]). Noninvasive hemoglobin measurements (SpHb) were recorded and were then compared with laboratory hemoglobin measurements.
RESULTS: Data were collected on 60 patients (OR = 25 and ICU = 45). The overall correlation of the Masimo SpHb and the laboratory Hb was .77 (P < .001) in the OR group with a mean difference of .29 g/dL (95% confidence interval [CI], .08-.49). The overall correlation in the ICU group was .67 (P < .001) with a mean difference of .05 g/dL (95% CI, -.22 to -.31).
CONCLUSIONS: Noninvasive hemoglobin monitoring is a new technology that correlated with laboratory values and supports the continued study of noninvasive hemoglobin monitoring.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21545905     DOI: 10.1016/j.amjsurg.2011.01.020

Source DB:  PubMed          Journal:  Am J Surg        ISSN: 0002-9610            Impact factor:   2.565


  20 in total

1.  Reliability of Continuous Non-Invasive Assessment of Hemoglobin and Fluid Responsiveness: Impact of Obesity and Abdominal Insufflation Pressures.

Authors:  Mia DeBarros; Marlin W Causey; Patrick Chesley; Matthew Martin
Journal:  Obes Surg       Date:  2015-07       Impact factor: 4.129

2.  Comparison of the accuracy of noninvasive hemoglobin monitoring for preoperative evaluation between adult and pediatric patients: a retrospective study.

Authors:  Yong-Hee Park; Seho Lim; Hyun Kang; Hwa-Yong Shin; Chong Wha Baek; Young Cheol Woo
Journal:  J Clin Monit Comput       Date:  2018-01-04       Impact factor: 2.502

3.  Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill?

Authors:  Benjamin A von Schweinitz; Robert A De Lorenzo; Peter J Cuenca; Richard L Anschutz; Paul B Allen
Journal:  Intern Emerg Med       Date:  2014-10-17       Impact factor: 3.397

4.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

5.  Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin.

Authors:  Douglas A Colquhoun; Katherine T Forkin; Marcel E Durieux; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2012-01-17       Impact factor: 2.502

6.  Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants.

Authors:  C Nicholas; R George; S Sardesai; M Durand; R Ramanathan; R Cayabyab
Journal:  J Perinatol       Date:  2015-03-05       Impact factor: 2.521

7.  Continuous noninvasive hemoglobin monitoring: the standard of care and future impact.

Authors:  Gerald J Kost; Nam K Tran
Journal:  Crit Care Med       Date:  2011-10       Impact factor: 7.598

8.  Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients.

Authors:  Werner Baulig; Burkhardt Seifert; Donat R Spahn; Oliver M Theusinger
Journal:  J Clin Monit Comput       Date:  2015-12-19       Impact factor: 2.502

9.  Validation of noninvasive hemoglobin measurements using co-oximetry in anesthetized dogs.

Authors:  Matt R Read; Jenna Rondeau; Grace P S Kwong
Journal:  Can Vet J       Date:  2016-11       Impact factor: 1.008

10.  The accuracy of non-invasively continuous total hemoglobin measurement by pulse CO-Oximetry undergoing acute normovolemic hemodilution and reinfusion of autologous blood.

Authors:  Junichi Saito; Masato Kitayama; Masafumi Oishi; Tomoyuki Kudo; Masahiro Sawada; Hiroshi Hashimoto; Kazuyoshi Hirota
Journal:  J Anesth       Date:  2014-06-28       Impact factor: 2.078

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.