Performance evaluation of a noninvasive hemoglobin monitoring device.

Etienne Gayat1, Antoine Bodin, Caroline Sportiello, Mireille Boisson, Jean-François Dreyfus, Emmanuel Mathieu, Marc Fischler.   

Abstract

STUDY
OBJECTIVE: Hemoglobin measurement is a routine procedure, and a noninvasive point-of-care device may increase the quality of care. The aim of the present study is to compare hemoglobin concentration values obtained with a portable totally noninvasive device, the Masimo Labs Radical-7 Pulse CO-Oximeter, with the results obtained by the ADVIA 2120 in the laboratory.
METHODS: This was a prospective monocentric open trial enrolling patients consulting in the emergency department of a university hospital from June 16 to December 17, 2009. The main outcome measure was the agreement between both methods and evaluation of the percentage of potential decision error for transfusion.
RESULTS: Samples from 300 consecutive patients were assessed. Hemoglobin concentration could not be obtained with the new device for 24 patients. In others, the mean bias, the lower and the upper limits of agreement between the 2 methods, was 1.8 g/dL (95% estimated confidence interval [CI] 1.5 to 2.1 g/dL), -3.3 g/dL (95% CI -3.8 to -2.8 g/dL), and 6.9 g/dL (95% CI 6.4 to 7.4 g/dL), respectively. The intraclass correlation coefficient was 0.53 (estimated 95% CI 0.10 to 0.74). The number of potential errors about transfusion decision was 38 (13% of patients). The peripheral oxygen saturation and the true value of hemoglobin concentration were independently associated with the bias.
CONCLUSION: Results from this widely available noninvasive point-of-care hemoglobin monitoring device were systematically biased and too unreliable to guide transfusion decisions.
Copyright © 2011 American College of Emergency Physicians. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21251729     DOI: 10.1016/j.annemergmed.2010.11.032

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  19 in total

1.  Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room.

Authors:  Jan Zatloukal; Jiri Pouska; Jakub Kletecka; Richard Pradl; Jan Benes
Journal:  J Clin Monit Comput       Date:  2015-10-27       Impact factor: 2.502

2.  Comparison of the accuracy of noninvasive hemoglobin monitoring for preoperative evaluation between adult and pediatric patients: a retrospective study.

Authors:  Yong-Hee Park; Seho Lim; Hyun Kang; Hwa-Yong Shin; Chong Wha Baek; Young Cheol Woo
Journal:  J Clin Monit Comput       Date:  2018-01-04       Impact factor: 2.502

3.  Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill?

Authors:  Benjamin A von Schweinitz; Robert A De Lorenzo; Peter J Cuenca; Richard L Anschutz; Paul B Allen
Journal:  Intern Emerg Med       Date:  2014-10-17       Impact factor: 3.397

4.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

5.  Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin.

Authors:  Douglas A Colquhoun; Katherine T Forkin; Marcel E Durieux; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2012-01-17       Impact factor: 2.502

6.  Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Authors:  Matthias L Riess; Paul S Pagel
Journal:  J Cardiothorac Vasc Anesth       Date:  2016-03-24       Impact factor: 2.628

7.  Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants.

Authors:  C Nicholas; R George; S Sardesai; M Durand; R Ramanathan; R Cayabyab
Journal:  J Perinatol       Date:  2015-03-05       Impact factor: 2.521

8.  Continuous noninvasive hemoglobin monitoring: the standard of care and future impact.

Authors:  Gerald J Kost; Nam K Tran
Journal:  Crit Care Med       Date:  2011-10       Impact factor: 7.598

9.  Continuous noninvasive hemoglobin monitor from pulse ox: ready for prime time?

Authors:  Bellal Joseph; Pantelis Hadjizacharia; Hassan Aziz; Kara Snyder; Julie Wynne; Narong Kulvatunyou; Andrew Tang; Terence O'Keeffe; Rifat Latifi; Randall Friese; Peter Rhee
Journal:  World J Surg       Date:  2013-03       Impact factor: 3.352

10.  Does a digital regional nerve block improve the accuracy of noninvasive hemoglobin monitoring?

Authors:  Ronald D Miller; Theresa A Ward; Charles E McCulloch; Neal H Cohen
Journal:  J Anesth       Date:  2012-08-01       Impact factor: 2.078

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.