Literature DB >> 20588064

Spectroscopic noninvasive measurement of hemoglobin compared with capillary and venous values in neonates.

H Rabe1, R Fernandez Alvarez, T Whitfield, F Lawson, H Jungmann.   

Abstract

BACKGROUND: Venepuncture-related blood loss is a common cause of neonatal anemia. Currently, this is the only way to obtain hemoglobin levels. This causes distress for the infant but can also lead to the need for blood transfusions. Recently, a new technique for measuring hemoglobin levels non-invasively has been developed to reduce iatrogenic blood loss and pain.
OBJECTIVE: To compare hemoglobin levels obtained using a transcutaneous spectroscopic device (Mediscan 2000, MBR Optical Systems, Wuppertal, Germany) with venous or capillary blood samples in neonates.
METHODS: Single-center prospective cohort study of term and preterm infants. The white light spectroscopic device was placed on the forearm for 60 s to measure hemoglobin content within 4 h of venous or capillary blood sampling. Pain reactions of the infants were assessed by using a neonatal pain assessment tool. Results were analyzed by Bland-Altman comparison and Wilcoxon signed-rank test.
RESULTS: 80 infants (mean gestational age 29.8 +/- 3.8 weeks, mean birth weight 1,300 +/- 690 g) were enrolled into the study. A total of 313 spectroscopic recordings within 2 h of a clinically indicated blood sample (181 capillary, 142 venous) were taken. The correlation coefficient R(2) was 0.96 for capillary/spectroscopic and 0.99 for venous/spectroscopic pairs. Pain scores were significantly less for the spectroscopic measurements (p < 0.01).
CONCLUSION: The results show good correlation between the hemoglobin blood levels and spectroscopic measurements. The slightly lower correlation coefficient for the capillary samples demonstrates a naturally higher variance in these values due to the laboratory method.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20588064     DOI: 10.1159/000261019

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  6 in total

1.  Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants.

Authors:  C Nicholas; R George; S Sardesai; M Durand; R Ramanathan; R Cayabyab
Journal:  J Perinatol       Date:  2015-03-05       Impact factor: 2.521

2.  Continuous noninvasive hemoglobin monitoring: the standard of care and future impact.

Authors:  Gerald J Kost; Nam K Tran
Journal:  Crit Care Med       Date:  2011-10       Impact factor: 7.598

Review 3.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

4.  Minimizing blood loss and the need for transfusions in very premature infants.

Authors:  Brigitte Lemyre; Megan Sample; Thierry Lacaze-Masmonteil
Journal:  Paediatr Child Health       Date:  2015 Nov-Dec       Impact factor: 2.253

Review 5.  Emerging point-of-care technologies for anemia detection.

Authors:  Ran An; Yuning Huang; Yuncheng Man; Russell W Valentine; Erdem Kucukal; Utku Goreke; Zoe Sekyonda; Connie Piccone; Amma Owusu-Ansah; Sanjay Ahuja; Jane A Little; Umut A Gurkan
Journal:  Lab Chip       Date:  2021-05-18       Impact factor: 6.799

Review 6.  Prevention of iron deficiency anemia in infants and toddlers.

Authors:  Sripriya Sundararajan; Heike Rabe
Journal:  Pediatr Res       Date:  2020-04-24       Impact factor: 3.756

  6 in total