| Literature DB >> 32943840 |
Anastasios John Kanellopoulos1,2.
Abstract
PURPOSE: To evaluate and compare the repeatability and agreement of Scheimpflug vs scanning-slit tomography of the cornea and the anterior chamber in terms of keratometric and tomographic indices in healthy eyes.Entities:
Keywords: Orbscan; Pentacam; Scheimpflug corneal tomography; scanning-slit corneal tomography
Year: 2020 PMID: 32943840 PMCID: PMC7481306 DOI: 10.2147/OPTH.S251998
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1This image illustrates the scans evaluated with both technologies in a random eye of one the patients studied: all 3 scanning-slit tomography individual scans (on the left) as well as all 3 respective Scheimpflug scans (on the right) for that specific eye are included.
Summary of the Comparative Statistical Analysis of the Parameters Studied
| SD Pentacam | SD Orbscan | CV (%) Pentacam | CV (%) Orbscan | ICC Pentacam | ICC Orbscan | P | |
|---|---|---|---|---|---|---|---|
| K2 steep keratometry | 0.11 | 0.17 | 0.25 | 0.39 | 0.99 | 0.98 | 0.034 |
| K1 flat keratometry | 0.10 | 0.16 | 0.23 | 0.37 | 0.99 | 0.98 | 0.027 |
| Kmax Steepest | 0.13 | 0.23 | 0.30 | 0.54 | 0.98 | 0.96 | >0.05 |
| X angle kappa | 0.23 | 0.13 | 193.78 | 31.91 | 0.52 | 0.72 | >0.05 |
| Y Angle kappa | 1.29 | 0.12 | 170.24 | 209.14 | 0.40 | 0.85 | >0.05 |
| K mean power | 0.10 | 0.14 | 0.23 | 0.34 | 0.98 | 0.98 | >0.05 |
| BFS front | 0.01 | 0.02 | 0.08 | 0.25 | 0.99 | 0.98 | 0.021 |
| BFS posterior | 0.02 | 0.03 | 0.30 | 0.45 | 0.99 | 0.98 | 0.039 |
| ACD | 0.01 | 0.01 | 0.43 | 0.29 | 0.99 | 0.99 | >0.05 |
| Astig power | 0.08 | 0.12 | 14.33 | 16.85 | 0.98 | 0.93 | >0.05 |
| Steep Axis | 6.89 | 6.11 | 9.47 | 6.86 | 0.7 | 0.55 | >0.05 |
| Pupil Diameter | 0.09 | 0.10 | 0.57 | 0.97 | 0.8 | 0.98 | >0.05 |
| Thinnest pachymetry | 3.19 | 5.73 | 0.57 | 0.97 | 0.99 | 0.99 | 0.038 |
| Thinnest x from vertex | 0.07 | 0.23 | 17.31 | 91.08 | 0.92 | 0.43 | 0.043 |
| Thinnest y from vertex | 0.07 | 0.23 | 47.59 | 159.59 | 0.87 | 0.65 | 0.031 |
Figure 2This image illustrates the Bland–Altman plots for the keratometric indices analyzed statistically.
Figure 3This image illustrates the Bland–Altman plots for the pachymetric indices analyzed statistically.
Figure 4This image illustrates the Bland–Altman plots for the other indices studied, analyzed statistically.
Statistical Analysis Correlation, Comparing the Parameters Studied Between the 2 Devices with the Correlation Coefficient r, and the 2-Tailed t-Test for the Parameters Studied
| r | P | |
|---|---|---|
| K2 steep keratometry | 0.94 | 0.022 |
| K1 flat keratometry | 0.92 | 0.04 |
| Kmax (steepest) | 0.8 | >0.05 |
| X angle kappa | 0.26 | >0.05 |
| Y angle kappa | 0.17 | >0.05 |
| Kmean power | 0.88 | >0.05 |
| BFS front | 0.96 | 0.032 |
| BFS posterior | 0.93 | >0.05 |
| ACD | 0.99 | <0.001 |
| Astigmatic power | 0.21 | <0.001 |
| Steep axis angle | 0.31 | >0.05 |
| Pupil diameter | 0.83 | <0.001 |
| Thinnest Pachymetry | 0.99 | <0.001 |
| Thinnest x from vertex | 0.66 | 0.003 |
| Thinnest y from vertex | 0.23 | >0.05 |